Diagnostic performance of CareStart™ malaria HRP2/pLDH (Pf/pan) combo test versus standard microscopy on falciparum and vivax malaria between China-Myanmar endemic borders
View through CrossRef
Abstract
Background
Rapid diagnostic test (RDT) is becoming an alternative way of establishing quickly the diagnosis of malaria infections, by detecting specific malaria antigens in suspected patients’ blood between the China-Myanmar endemic borders areas, towards achieving the National Malaria Elimination programme by 2020. The objective of this study is to evaluate the performance of CareStart™ Malaria Pf/Pan RDT kit for the diagnosis of malaria infections in suspected patients. Blood examination by microscopy was taken as gold standard to evaluate CareStart™ kit’s sensitivity, specificity and predictive value and corrected with PCR assay.
Results
Overall 126 of 241 (52.28%) malaria cases were detected by microscopy compared to 115 of 241(47.72%) CareStart™ kit and 128 of 241 (53.11%) PCR corrected assay. CareStart™ kit’s sensitivity and specificity for the diagnosis of malaria were 89.68% and 98.26% respectively, compared to standard microscopy, whereas the sensitivity and specificity for falciparum malaria were 88.52% and 98.26%, and for vivax malaria: 90.77% and 100%. The CareStart™ positive predictive values were 98.26% (93.88-99.52%, 95% CI) compared to 100% (96.77-100%, 95% CI) for PCR-corrected, and the negative predictive values of 89.68% (83.15-93.87%, 95% CI) were the same in microscopy as PCR-corrected. The diagnostic accuracy of CareStart™ kit versus microscopy and PCR were 93.78% (89.99-96.19%, 95% CI) and 94.61% (90.99-96.82%, 95% CI) respectively. The likelihood of diagnostic of malaria positive was almost similar between microscopy and CareStart™ kit, with an entropy reduction of 60.0% compared to a weak likelihood of misdiagnosis of 0.10 (0.09-0.12, 95% CI), with an entropy reduction of 36.01%.
Conclusion
The accuracy of CareStart™ kit is comparable to gold standard microscopy in these areas, it is easy to perform and suitable for cross-border diagnosis and monitoring of local or imported malaria patterns by any local health staff in endemic remotes.
Springer Science and Business Media LLC
Title: Diagnostic performance of CareStart™ malaria HRP2/pLDH (Pf/pan) combo test versus standard microscopy on falciparum and vivax malaria between China-Myanmar endemic borders
Description:
Abstract
Background
Rapid diagnostic test (RDT) is becoming an alternative way of establishing quickly the diagnosis of malaria infections, by detecting specific malaria antigens in suspected patients’ blood between the China-Myanmar endemic borders areas, towards achieving the National Malaria Elimination programme by 2020.
The objective of this study is to evaluate the performance of CareStart™ Malaria Pf/Pan RDT kit for the diagnosis of malaria infections in suspected patients.
Blood examination by microscopy was taken as gold standard to evaluate CareStart™ kit’s sensitivity, specificity and predictive value and corrected with PCR assay.
Results
Overall 126 of 241 (52.
28%) malaria cases were detected by microscopy compared to 115 of 241(47.
72%) CareStart™ kit and 128 of 241 (53.
11%) PCR corrected assay.
CareStart™ kit’s sensitivity and specificity for the diagnosis of malaria were 89.
68% and 98.
26% respectively, compared to standard microscopy, whereas the sensitivity and specificity for falciparum malaria were 88.
52% and 98.
26%, and for vivax malaria: 90.
77% and 100%.
The CareStart™ positive predictive values were 98.
26% (93.
88-99.
52%, 95% CI) compared to 100% (96.
77-100%, 95% CI) for PCR-corrected, and the negative predictive values of 89.
68% (83.
15-93.
87%, 95% CI) were the same in microscopy as PCR-corrected.
The diagnostic accuracy of CareStart™ kit versus microscopy and PCR were 93.
78% (89.
99-96.
19%, 95% CI) and 94.
61% (90.
99-96.
82%, 95% CI) respectively.
The likelihood of diagnostic of malaria positive was almost similar between microscopy and CareStart™ kit, with an entropy reduction of 60.
0% compared to a weak likelihood of misdiagnosis of 0.
10 (0.
09-0.
12, 95% CI), with an entropy reduction of 36.
01%.
Conclusion
The accuracy of CareStart™ kit is comparable to gold standard microscopy in these areas, it is easy to perform and suitable for cross-border diagnosis and monitoring of local or imported malaria patterns by any local health staff in endemic remotes.
Related Results
Assessment of the Performance of Lactate Dehydrogenase-Based Rapid Diagnostic Test for Malaria in Djibouti in 2022–2023
Assessment of the Performance of Lactate Dehydrogenase-Based Rapid Diagnostic Test for Malaria in Djibouti in 2022–2023
Until 2020, Djiboutian health authorities relied on histidine-rich protein-2 (HRP2)-based rapid diagnostic tests (RDTs) to establish the diagnosis of Plasmodium falciparum. The rap...
Widespread pfhrp2/3 deletions and HRP2-based false-negative results in southern Ethiopia
Widespread pfhrp2/3 deletions and HRP2-based false-negative results in southern Ethiopia
Abstract
Background
Rapid diagnostic tests (RDTs) have played a significant role in expanding case management in peripheral healthcare systems. Histidine-rich protein-2 (H...
Performance of a Novel P. falciparum Rapid Diagnostic Test in Areas of Widespread hrp2/3 Gene Deletion
Performance of a Novel P. falciparum Rapid Diagnostic Test in Areas of Widespread hrp2/3 Gene Deletion
Abstract
Background
Rapid diagnostic tests (RDTs) have improved malaria diagnosis; however, the emergence of hrp2/3 gene deletio...
Malaria misdiagnosis in the routine health system in Arba Minch area district in southwest Ethiopia: an implication for malaria control and elimination
Malaria misdiagnosis in the routine health system in Arba Minch area district in southwest Ethiopia: an implication for malaria control and elimination
Abstract
Background
Microscopy and Rapid Diagnostic Test (RDT) are the two widely used malaria diagnostic tools in Ethiopian health facilities. Accurate diagnosis is vital...
Assessing the Performance of CareStart™ Malaria Rapid Diagnostic Tests in Northwest Ethiopia: A Cross-Sectional Study
Assessing the Performance of CareStart™ Malaria Rapid Diagnostic Tests in Northwest Ethiopia: A Cross-Sectional Study
Background. While rapid diagnostic tests are an alternative diagnostic tests for microscopy in the diagnosis of malaria in rural settings, their performance has been inconsistent. ...
Prevalence, Demographic Patterns, and Seasonal Distribution of Malaria in District Dera Ismail Khan, Pakistan
Prevalence, Demographic Patterns, and Seasonal Distribution of Malaria in District Dera Ismail Khan, Pakistan
Abstract
Malaria remains the most significant vector-borne disease worldwide, with over 200 million cases reported annually, causing approximatel...
Evaluating performance of multiplex real time PCR for the diagnosis of malaria at elimination targeted low transmission settings of Ethiopia
Evaluating performance of multiplex real time PCR for the diagnosis of malaria at elimination targeted low transmission settings of Ethiopia
Abstract
Background
Malaria incidence has declined in Ethiopia in the past 10 years. Current malaria diagnostic tests, including light microscopy an...
MALDI-TOF mass spectrometry combined with machine learning algorithms to identify protein profiles related to malaria infection in human sera from Côte d’Ivoire
MALDI-TOF mass spectrometry combined with machine learning algorithms to identify protein profiles related to malaria infection in human sera from Côte d’Ivoire
Abstract
Background
In sub-Saharan Africa, Plasmodium falciparum is the most prevalent species of malaria parasites. In endemic areas, malaria is...