Javascript must be enabled to continue!
Planar dose calculation of electron therapy
View through CrossRef
Abstract
Purpose. The aim of this study is to determine the planar dose distribution of irregularly-shaped electron beams at their maximum dose depth (z
max) using the modied lateral build-up ratio (LBR) and curve-fitting methods. Methods. Circular and irregular cutouts were created using Cerrobend alloy for a 14 × 14 cm2 applicator. Percentage depth dose (PDD) at the standard source-surface-distance (SSD = 100 cm) and point dose at different SSD were measured for each cutout. Orthogonal profiles of the cutouts were measured at z
max. Data were collected for 6, 9, 12, and 15 MeV electron beam energies on a VERSA HDTM LINAC using the IBA Blue Phantom2 3D water phantom system. The planar dose distributions of the cutouts were also measured at z
max in solid water using EDR2 films. Results. The measured PDD curves were normalized to a normalization depth (d
0) of 1 mm. The lateral-buildup-ratio (LBR), lateral spread parameter (σ
R
(z)), and effective SSD (SSD
eff
) for each cutout were calculated using the PDD of the open applicator as the reference field. The modified LBR method was then employed to calculate the planar dose distribution of the irregular cutouts within the field at least 5 mm from the edge. A simple curve-fitting model was developed based on the profile shapes of the circular cutouts around the field edge. This model was used to calculate the planar dose distribution of the irregular cutouts in the region from 3 mm outside to 5 mm inside the field edge. Finally, the calculated planar dose distribution was compared with the film measurement. Conclusions. The planar dose distribution of electron therapy for irregular cutouts at z
max was calculated using the improved LBR method and a simple curve-fitting model. The calculated profiles were within 3% of the measured values. The gamma passing rate with a 3%/3 mm and 10% dose threshold was more than 96%.
Title: Planar dose calculation of electron therapy
Description:
Abstract
Purpose.
The aim of this study is to determine the planar dose distribution of irregularly-shaped electron beams at their maximum dose depth (z
max) using the modied lateral build-up ratio (LBR) and curve-fitting methods.
Methods.
Circular and irregular cutouts were created using Cerrobend alloy for a 14 × 14 cm2 applicator.
Percentage depth dose (PDD) at the standard source-surface-distance (SSD = 100 cm) and point dose at different SSD were measured for each cutout.
Orthogonal profiles of the cutouts were measured at z
max.
Data were collected for 6, 9, 12, and 15 MeV electron beam energies on a VERSA HDTM LINAC using the IBA Blue Phantom2 3D water phantom system.
The planar dose distributions of the cutouts were also measured at z
max in solid water using EDR2 films.
Results.
The measured PDD curves were normalized to a normalization depth (d
0) of 1 mm.
The lateral-buildup-ratio (LBR), lateral spread parameter (σ
R
(z)), and effective SSD (SSD
eff
) for each cutout were calculated using the PDD of the open applicator as the reference field.
The modified LBR method was then employed to calculate the planar dose distribution of the irregular cutouts within the field at least 5 mm from the edge.
A simple curve-fitting model was developed based on the profile shapes of the circular cutouts around the field edge.
This model was used to calculate the planar dose distribution of the irregular cutouts in the region from 3 mm outside to 5 mm inside the field edge.
Finally, the calculated planar dose distribution was compared with the film measurement.
Conclusions.
The planar dose distribution of electron therapy for irregular cutouts at z
max was calculated using the improved LBR method and a simple curve-fitting model.
The calculated profiles were within 3% of the measured values.
The gamma passing rate with a 3%/3 mm and 10% dose threshold was more than 96%.
Related Results
THE EFFECT OF ATROVASTATIN ON THE QRS DURATION IN THE PATIENTS WITH ACUTE ANTERIOR ST ELEVATION MYOCARDIAL INFARCTION UNDERGOING PRIMARY ANGIOPLASTY
THE EFFECT OF ATROVASTATIN ON THE QRS DURATION IN THE PATIENTS WITH ACUTE ANTERIOR ST ELEVATION MYOCARDIAL INFARCTION UNDERGOING PRIMARY ANGIOPLASTY
Objectives
To evaluate the value of intensive atorvastatin treatment on changes of QRS duration and ventricular arrhythmia in the patients with acute anterior ST ...
Different dose series of human papillomavirus vaccine in young females: a pair-wise meta-analysis and network meta-analysis from randomized controlled trials
Different dose series of human papillomavirus vaccine in young females: a pair-wise meta-analysis and network meta-analysis from randomized controlled trials
ObjectiveTo investigate the application value of different dose of HPV vaccine in young females.Data sourcesThe following databases were searched: Cochrane Library, PubMed, Embase,...
Indeterminate solitary vertebral lesions on planar scintigraphy
Indeterminate solitary vertebral lesions on planar scintigraphy
Summary
Objective: This study aims to evaluate the added value of hybrid SPECT-CT in differential diagnosis of indeterminate solitary vertebral lesion (SVL) on planar sci...
Optimizing Hydroxyurea Dosing in Sickle Cell Anemia: The Uganda MTD Study
Optimizing Hydroxyurea Dosing in Sickle Cell Anemia: The Uganda MTD Study
Introduction. Hydroxyurea treatment has proven safety, feasibility, and efficacy for children with sickle cell anemia living in sub-Saharan Africa. Even in malaria endemic regions,...
REGULAR ARTICLES
REGULAR ARTICLES
L. Cowen and
C. J.
Schwarz
657Les Radio‐tags, en raison de leur détectabilitéélevée, ...
TU‐B‐201‐03: Accounting for MV Imaging Dose and the Future of MV Imaging
TU‐B‐201‐03: Accounting for MV Imaging Dose and the Future of MV Imaging
Radiotherapy is in the era of image guided radiation therapy (IGRT) in which imaging procedures are routinely performed for patient positioning, and target localization. The additi...
Advanced HEMT Characteristics of Epitaxial Quality-improved GaN by Using Patterned Sapphire Substract
Advanced HEMT Characteristics of Epitaxial Quality-improved GaN by Using Patterned Sapphire Substract
INTRODUCTION
Accomplishing with the booming market of personal communication services and the fifth generation (5G) mobile systems, the demands for high frequency an...
Dosimetric comparisons of intensity‐modulated radiation therapy and three‐dimensional conformal radiation therapy for left‐sided breast cancer after radical surgery
Dosimetric comparisons of intensity‐modulated radiation therapy and three‐dimensional conformal radiation therapy for left‐sided breast cancer after radical surgery
AbstractObjectivePostoperative radiotherapy for breast cancer is an effective way to control tumor recurrence; however, there are advantages and disadvantages to different radiothe...

