Javascript must be enabled to continue!
Real-world accuracy of robotic-assisted total knee arthroplasty and its impact on expedited recovery
View through CrossRef
AbstractDespite total knee arthroplasty (TKA) being the gold standard for end-stage knee osteoarthritis, 20% of patients remain dissatisfied. Robotic-assisted arthroplasty promises unparalleled control of the accuracy of bone cuts, implant positioning, control of gap balance, and resultant hip–knee–ankle (HKA) axis. Patients underwent clinical and radiological assessments, including knee CT scans and patient-reported outcome measures (PROMs), preoperatively. Follow-up assessments were conducted at 2 weeks, 6 weeks, and 3 months post-operatively, with imaging repeated at 6 weeks. A total of 155 patients underwent robotic-assisted TKA and have completed 3 months of follow-up. Mean pre-operative HKA axis was 7.39 ± 5.52 degrees varus, improving to 1.34 ± 2.22 degrees varus post-operatively. Restoration of HKA axis was 0.76 ± 1.9 degrees from intra-operative planning (p < 0.0005). Implant placement accuracy in the coronal plane was 0.08 ± 1.36 degrees (p = 0.458) for the femoral component and 0.71 ± 1.3 degrees (p < 0.0005) for the tibial component. Rotational alignment mean deviation was 0.39 ± 1.49 degrees (p = 0.001). Most patients (98.1%) had ≤ 2 mm difference in extension–flexion gaps. PROM scores showed improvement and exceeded pre-operative scores by 6 weeks post-surgery. Robotic-assisted knee arthroplasty provides precise control over traditionally subjective factors, demonstrating excellent early post-operative outcomes.Level of evidence Prospective observational study—II.
Springer Science and Business Media LLC
Title: Real-world accuracy of robotic-assisted total knee arthroplasty and its impact on expedited recovery
Description:
AbstractDespite total knee arthroplasty (TKA) being the gold standard for end-stage knee osteoarthritis, 20% of patients remain dissatisfied.
Robotic-assisted arthroplasty promises unparalleled control of the accuracy of bone cuts, implant positioning, control of gap balance, and resultant hip–knee–ankle (HKA) axis.
Patients underwent clinical and radiological assessments, including knee CT scans and patient-reported outcome measures (PROMs), preoperatively.
Follow-up assessments were conducted at 2 weeks, 6 weeks, and 3 months post-operatively, with imaging repeated at 6 weeks.
A total of 155 patients underwent robotic-assisted TKA and have completed 3 months of follow-up.
Mean pre-operative HKA axis was 7.
39 ± 5.
52 degrees varus, improving to 1.
34 ± 2.
22 degrees varus post-operatively.
Restoration of HKA axis was 0.
76 ± 1.
9 degrees from intra-operative planning (p < 0.
0005).
Implant placement accuracy in the coronal plane was 0.
08 ± 1.
36 degrees (p = 0.
458) for the femoral component and 0.
71 ± 1.
3 degrees (p < 0.
0005) for the tibial component.
Rotational alignment mean deviation was 0.
39 ± 1.
49 degrees (p = 0.
001).
Most patients (98.
1%) had ≤ 2 mm difference in extension–flexion gaps.
PROM scores showed improvement and exceeded pre-operative scores by 6 weeks post-surgery.
Robotic-assisted knee arthroplasty provides precise control over traditionally subjective factors, demonstrating excellent early post-operative outcomes.
Level of evidence Prospective observational study—II.
Related Results
PROMs in Total Hip Arthroplasty: A Comparison between Robotic Mako Assisted Versus Manual Technique
PROMs in Total Hip Arthroplasty: A Comparison between Robotic Mako Assisted Versus Manual Technique
Study Purpose: The purpose of this observational study is to evaluate the degree of satisfaction and improvement in quality of life through PROMs (Patient Related Outcome Measures)...
Hip Functions in Patients with Rotating Hinge Knee Arthroplasty
Hip Functions in Patients with Rotating Hinge Knee Arthroplasty
Abstract
Background Rotating hinged knee prosthesis is frequently used for certain indications. The indications and results for rotating hinge knee prosthesis in primary an...
The Influence of Kinematic Alignment on Patellofemoral Joint Biomechanics in Total Knee Arthroplasty
The Influence of Kinematic Alignment on Patellofemoral Joint Biomechanics in Total Knee Arthroplasty
Background: Anterior knee pain is a prevalent issue post total knee arthroplasty, often necessitating revision surgery. Various factors contribute to this complication, including p...
Current therapeutic strategies for erectile function recovery after radical prostatectomy – literature review and meta-analysis
Current therapeutic strategies for erectile function recovery after radical prostatectomy – literature review and meta-analysis
Radical prostatectomy is the most commonly performed treatment option for localised prostate cancer. In the last decades the surgical technique has been improved and modified in or...
Learning curve analysis of robotic-assisted total knee arthroplasty with the HURWA surgical system
Learning curve analysis of robotic-assisted total knee arthroplasty with the HURWA surgical system
Abstract
Purpose The aim of this study was to analyze the learning curve of total operative time, bone-cutting accuracy, and limb alignment in robot-assisted total knee art...
Complications of revision knee arthroplasty
Complications of revision knee arthroplasty
Due to the growing scale of arthroplasty and the widening of indications for knee joint replacement, the durability of joint functioning has become a critical issue. Untimely diagn...
Interesting Cases in Robotic-Arm Assisted Total Knee Arthroplasty
Interesting Cases in Robotic-Arm Assisted Total Knee Arthroplasty
Introduction:Robotic-arm assisted surgery allows for the execution of well-aligned knee arthroplasty regardless of pre-existing deformity. This case series is presented to show the...
Mid-Term Results of Two-stage Revision Total Knee Arthroplasty for Infected Total Knee Arthroplasty
Mid-Term Results of Two-stage Revision Total Knee Arthroplasty for Infected Total Knee Arthroplasty
Introduction. Today, knee arthroplasty is one of the most common surgical procedures. Infection after endoprosthetics is the most dangerous complication in patients who have underg...