Dosimetric characteristics of radiophotoluminescent glass dosimeters for proton beams

View through CrossRef
Abstract Background: The study of radiophotoluminescent glass dosimeters (RGDs) in the clinical usage of proton beams is limited. The aim of this study was to investigate the dosimetric characteristics of RGDs for pencil beam scanning proton therapy. The feasibility of using an RGD in end-to-end testing of intensity-modulated proton therapy (IMPT) plans at various treatment sites was also evaluated. Materials and methods: The dosimetric characteristics of the GD-302M type glass dosimeter were studied in terms of uniformity, short-term and long-term reproducibility, stability of the magazine position readout, dose linearity in the range from 0.2-20 Gy, energy response in 70-220 MeV, MU/spot, dose rate response, and fading effect. The reference conditions of the spot scanning beam from the Varian ProBeam Compact system were operation at 160 MeV, a 2 cm water equivalent depth in a solid water phantom, a 10×10 cm field size at the isocenter, and 2 Gy dose delivery. End-to-end testing of IMPT plans for the head, abdomen, and pelvis was verified by using the Alderson Rando phantom. The overall uncertainty analysis was confirmed in this study. Results: The relative response of RGDs for the uniformity test was within 0.95-1.05. The %CVs of the short-term and long-term reproducibility were 1.16% and 1.50%, respectively. The FGD-1000 automatic reader showed stable magazine position readout. The dose linearity was found to have an obviously good linear relationship, with R2 = 0.9988. The energy response relative to 160 MeV was approximately within 4.0%. The MU/spot and dose rate had less effect on the RGD readout. The fading effect was relatively stable for 10 weeks of storage, within 2.4%. For the end-to-end test, the maximum difference between the treatment plan and RGD measurement showed a very good result that was within 1.0%. The overall uncertainty of the RGD measurement for the proton beam was 4.6%. Conclusion: RGDs have confirmed the potential for proton dosimetry, including in end-to-end testing. The appropriate correction factor for the energy response can be applied for dose verification of scanning proton beams.
Title: Dosimetric characteristics of radiophotoluminescent glass dosimeters for proton beams
Description:
Abstract Background: The study of radiophotoluminescent glass dosimeters (RGDs) in the clinical usage of proton beams is limited.
The aim of this study was to investigate the dosimetric characteristics of RGDs for pencil beam scanning proton therapy.
The feasibility of using an RGD in end-to-end testing of intensity-modulated proton therapy (IMPT) plans at various treatment sites was also evaluated.
Materials and methods: The dosimetric characteristics of the GD-302M type glass dosimeter were studied in terms of uniformity, short-term and long-term reproducibility, stability of the magazine position readout, dose linearity in the range from 0.
2-20 Gy, energy response in 70-220 MeV, MU/spot, dose rate response, and fading effect.
The reference conditions of the spot scanning beam from the Varian ProBeam Compact system were operation at 160 MeV, a 2 cm water equivalent depth in a solid water phantom, a 10×10 cm field size at the isocenter, and 2 Gy dose delivery.
End-to-end testing of IMPT plans for the head, abdomen, and pelvis was verified by using the Alderson Rando phantom.
The overall uncertainty analysis was confirmed in this study.
Results: The relative response of RGDs for the uniformity test was within 0.
95-1.
05.
The %CVs of the short-term and long-term reproducibility were 1.
16% and 1.
50%, respectively.
The FGD-1000 automatic reader showed stable magazine position readout.
The dose linearity was found to have an obviously good linear relationship, with R2 = 0.
9988.
The energy response relative to 160 MeV was approximately within 4.
0%.
The MU/spot and dose rate had less effect on the RGD readout.
The fading effect was relatively stable for 10 weeks of storage, within 2.
4%.
For the end-to-end test, the maximum difference between the treatment plan and RGD measurement showed a very good result that was within 1.
0%.
The overall uncertainty of the RGD measurement for the proton beam was 4.
6%.
Conclusion: RGDs have confirmed the potential for proton dosimetry, including in end-to-end testing.
The appropriate correction factor for the energy response can be applied for dose verification of scanning proton beams.

Related Results

Odd version Mathieu-Gaussian beam based on Green function
Odd version Mathieu-Gaussian beam based on Green function
Like the theoretical pattern of non-diffracting Bessel beams, ideal non-diffracting Mathieu beams also carry infinite energy, but cannot be generated as a physically realizable ent...
Construction of Bi-Pearcey beams and their mathematical mechanism
Construction of Bi-Pearcey beams and their mathematical mechanism
We present a theoretical expression in the form of the Pearcey function by deducing the Fresnel diffraction distribution of an elliptic line. Then, we numerically simulate and expe...
Dynamic Characteristics Analysis of Three-Layer Steel–Concrete Composite Beams
Dynamic Characteristics Analysis of Three-Layer Steel–Concrete Composite Beams
The dynamic behavior of three-layer composite beams, consisting of concrete slabs and steel beams, is influenced by the structural configuration of each layer as well as the shear ...
Current Developments in Radiotherapy Quality Control Methods
Current Developments in Radiotherapy Quality Control Methods
Quality control is of critical importance to ensure the accurate and reliable implementation of treatment in radiotherapy. Radiotherapy is a primary method of treating diseases suc...
Experimental study and computational analysis of structural performance of reinforced geopolymer concrete beams
Experimental study and computational analysis of structural performance of reinforced geopolymer concrete beams
PurposeIn this study, the aim is to explore the effects of geopolymer concrete (GPC) strength and reinforcement ratio on the flexural performance of reinforced GPC beams. Furthermo...
Investigation of alanine dosimeter for output measurement in radiotherapy: Results from multicenter in Thailand
Investigation of alanine dosimeter for output measurement in radiotherapy: Results from multicenter in Thailand
Background: Alanine dosimeters are generally used in high-dose industrial applications (kGy). Later, research into employing alanine as a dosimeter in radiotherapy (1-20 Gy) has in...
Unraveling the mechanism of proton translocation in the extracellular half-channel of bacteriorhodopsin
Unraveling the mechanism of proton translocation in the extracellular half-channel of bacteriorhodopsin
AbstractBacteriorhodopsin, a light activated protein that creates a proton gradient in halobacteria, has long served as a simple model of proton pumps. Within bacteriorhodopsin, se...
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
<p>This dissertation investigates the dosimetric influence of mixed photon beams in volumetric modulated arc therapy (VMAT) for prostate cancer and presents new algorithmic f...

Back to Top