Evaluation of qt prolongation in patients following conduction system pacing implantation
View through CrossRef
Abstract
Background
Conduction system pacing is becoming increasingly popular as a pacing technique; however, its impact on repolarization, particularly on QT interval prolongation, remains unclear.
Purpose
This study aimed to evaluate the effect of conduction system pacing on the QT interval.
Methods
We conducted a retrospective review of all patients who received a left bundle branch lead at two centers between January 1, 2022, and June 30, 2024. Patients with left bundle pacing were included if they: identified as having deep septal or left bundle pacing (satisfying 2 of 3 markers—LVAT <80 ms, R’ in V1, PR1R6 >30 ms) and had a prolonged corrected QT interval (QTc >550 ms). Demographic and clinical data were collected through electronic medical records.
Results
From 597 patients who received a 3830 lead, 14 patients with left bundle branch pacing were identified as having QTc >550 ms (mean age 82 ± 6 years, 5 females). Nine patients received their device for complete heart block, 3 for tachy-brady and one after AV [MG1] node ablation. The average ejection fraction was 47 ± 13%. Post-implant ECG analysis showed an average QRS width of 127 ± 14 ms, an average QTc interval of 584 ± 28 ms, and an average heart rate of 74 ± 14 bpm. Notably, one patient experienced Torsades de Pointes three hours post-AV node ablation, with a QT interval of 600 ms that persisted for two weeks post-implant. All patients survived, with a mean follow-up period of 609 ± 470 days.
Conclusion
QT prolongation can occur following the placement of a conduction system pacing lead, which may elevate the risk of Torsades de Pointes post-implantation. In cases of prolonged QT, increasing heart rate may help to reduce the QT interval, though it may not fully normalize it. QT prolongation may persist for more than two weeks post-implant. Clinicians should be mindful of the potential for QT prolongation with conduction system pacing and consider closer follow-up until QT normalizes.
Oxford University Press (OUP)
Title: Evaluation of qt prolongation in patients following conduction system pacing implantation
Description:
Abstract
Background
Conduction system pacing is becoming increasingly popular as a pacing technique; however, its impact on repolarization, particularly on QT interval prolongation, remains unclear.
Purpose
This study aimed to evaluate the effect of conduction system pacing on the QT interval.
Methods
We conducted a retrospective review of all patients who received a left bundle branch lead at two centers between January 1, 2022, and June 30, 2024.
Patients with left bundle pacing were included if they: identified as having deep septal or left bundle pacing (satisfying 2 of 3 markers—LVAT <80 ms, R’ in V1, PR1R6 >30 ms) and had a prolonged corrected QT interval (QTc >550 ms).
Demographic and clinical data were collected through electronic medical records.
Results
From 597 patients who received a 3830 lead, 14 patients with left bundle branch pacing were identified as having QTc >550 ms (mean age 82 ± 6 years, 5 females).
Nine patients received their device for complete heart block, 3 for tachy-brady and one after AV [MG1] node ablation.
The average ejection fraction was 47 ± 13%.
Post-implant ECG analysis showed an average QRS width of 127 ± 14 ms, an average QTc interval of 584 ± 28 ms, and an average heart rate of 74 ± 14 bpm.
Notably, one patient experienced Torsades de Pointes three hours post-AV node ablation, with a QT interval of 600 ms that persisted for two weeks post-implant.
All patients survived, with a mean follow-up period of 609 ± 470 days.
Conclusion
QT prolongation can occur following the placement of a conduction system pacing lead, which may elevate the risk of Torsades de Pointes post-implantation.
In cases of prolonged QT, increasing heart rate may help to reduce the QT interval, though it may not fully normalize it.
QT prolongation may persist for more than two weeks post-implant.
Clinicians should be mindful of the potential for QT prolongation with conduction system pacing and consider closer follow-up until QT normalizes.
Related Results
Acute Changes in Cardiac Synchrony and Output According to RV Pacing Sites in Koreans with Normal Cardiac Function
Acute Changes in Cardiac Synchrony and Output According to RV Pacing Sites in Koreans with Normal Cardiac Function
Introduction: The synchrony of the pacing heart can be affected by the right ventricular (RV) pacing site and is crucial to cardiac function in pacemaker recipients. We evaluated t...
Changes in pacing variation with increasing race duration in ultra-triathlon races
Changes in pacing variation with increasing race duration in ultra-triathlon races
Abstract
Background: Despite the increasing scientific interest in the relationship between pacing and performance in endurance sports, little information is available abou...
Long‐Term Follow‐up Impact of Dual‐Chamber Pacing on Patients with Hypertrophic Obstructive Cardiomyopathy
Long‐Term Follow‐up Impact of Dual‐Chamber Pacing on Patients with Hypertrophic Obstructive Cardiomyopathy
BackgroundPacing has been proposed as a treatment for patients with hypertrophic obstructive cardiomyopathy (HOCM), but there are few studies with long‐term follow‐up. We evaluated...
Retinitis Pigmentosa
Retinitis Pigmentosa
In studying the cases with typical and atypical pigmentary degeneration of the retina we strived to analyse in the clinical material all ophthalmoscopic and ocular changes together...
Right ventricular pacing is not harmful as perceived previously
Right ventricular pacing is not harmful as perceived previously
Abstract
Background: Pacing from the apex of the RV is not optimal. This study aimed to assess and compare RV and LV pacing of LV function and dyssynchrony.
Thirty-six pati...
Prophylactic epicardial pacemaker implantation in tricuspid valve replacement
Prophylactic epicardial pacemaker implantation in tricuspid valve replacement
Abstract
OBJECTIVES
Patients undergoing surgical tricuspid valve replacement (TVR) are at high risk of atriovent...
Treatment strategy for acute myocarditis in pediatric patients requiring emergency intervention
Treatment strategy for acute myocarditis in pediatric patients requiring emergency intervention
Abstract
Background
Patients with acute myocarditis present with a wide range of symptoms. Treatment strategies for pedia...
Treatment Strategy for Acute Myocarditis in Pediatric Patients Requiring Emergency Intervention
Treatment Strategy for Acute Myocarditis in Pediatric Patients Requiring Emergency Intervention
Abstract
Background: Patients with acute myocarditis present with a wide range of symptoms. Treatment strategies for pediatric patients with circulatory failure comprise ex...