Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees
View through CrossRef
The objective of this study is to examine the mechanical behavior of two-piece abutments (Morse taper with 16° internal angulation and Morse taper with 11.5° internal angulation) before and after cyclic fatigue testing, following ISO 14801:2016 guidelines. The specimens were divided into three groups: a modified Morse taper with a taper angle of 16° (GM group), a conventional Morse taper (taper angle of 11.5° deg) with a two-piece (CMt group), and one-piece abutments (CMo group). Each experimental group was formed by ten implants and ten abutments (n = 10) for a total of 30 specimens (n = 30). The abutments were tightened and loosened, and a fatigue test was applied with 15 Hz and 5 × 106 cycles. Subsequently, the abutments were loosened, and a pull-out test was performed on the CMt group. Finite element analysis (FEA) was conducted on stress concentration regions. The statistical analysis of the loosening test was performed using two-way ANOVA and Tukey’s tests (p < 0.05) to compare screw loosening within each group and between the groups with and without mechanical fatigue. Significant differences were found among the three groups in the loosening test when analyzing the values with and without fatigue (p < 0.001) within each group. When the groups were compared with each other, there was also a significant difference between them (p < 0.001), except between groups GM and CMt without fatigue (p = 0.840). In the pull-out test of the CMt group, the sample exhibited frictional locking only after fatigue (mean = 94.2 N). The FEA demonstrated a varied stress distribution in all groups. The stress was found to be more concentrated in the upper third and middle third regions of the implant, as well as in the opposite region of the load application for all three groups. Although the CMo group showed lower rates of loosening, it displayed a poorer stress distribution in comparison to the GM and CMt groups. On the other hand, the CMt group exhibited a satisfactory frictional lock after undergoing the fatigue tests.
Title: Mechanical Behavior of One-Piece and Two-Piece Tapered Prosthetic Abutments with 11.5 and 16 Degrees
Description:
The objective of this study is to examine the mechanical behavior of two-piece abutments (Morse taper with 16° internal angulation and Morse taper with 11.
5° internal angulation) before and after cyclic fatigue testing, following ISO 14801:2016 guidelines.
The specimens were divided into three groups: a modified Morse taper with a taper angle of 16° (GM group), a conventional Morse taper (taper angle of 11.
5° deg) with a two-piece (CMt group), and one-piece abutments (CMo group).
Each experimental group was formed by ten implants and ten abutments (n = 10) for a total of 30 specimens (n = 30).
The abutments were tightened and loosened, and a fatigue test was applied with 15 Hz and 5 × 106 cycles.
Subsequently, the abutments were loosened, and a pull-out test was performed on the CMt group.
Finite element analysis (FEA) was conducted on stress concentration regions.
The statistical analysis of the loosening test was performed using two-way ANOVA and Tukey’s tests (p < 0.
05) to compare screw loosening within each group and between the groups with and without mechanical fatigue.
Significant differences were found among the three groups in the loosening test when analyzing the values with and without fatigue (p < 0.
001) within each group.
When the groups were compared with each other, there was also a significant difference between them (p < 0.
001), except between groups GM and CMt without fatigue (p = 0.
840).
In the pull-out test of the CMt group, the sample exhibited frictional locking only after fatigue (mean = 94.
2 N).
The FEA demonstrated a varied stress distribution in all groups.
The stress was found to be more concentrated in the upper third and middle third regions of the implant, as well as in the opposite region of the load application for all three groups.
Although the CMo group showed lower rates of loosening, it displayed a poorer stress distribution in comparison to the GM and CMt groups.
On the other hand, the CMt group exhibited a satisfactory frictional lock after undergoing the fatigue tests.
Related Results
Fatigue Behavior of Multi/Unit-Supported Dental Restorations: Implant Platform vs. Prosthetic Platform
Fatigue Behavior of Multi/Unit-Supported Dental Restorations: Implant Platform vs. Prosthetic Platform
The increasing popularity of Multi/Unit abutments in dental restorations is attributed to their clinical advantages, yet little is known about their mechanical behavior, particular...
Comparative Desensitization Effect of Vital Abutments Realized by Different Methods
Comparative Desensitization Effect of Vital Abutments Realized by Different Methods
Vital abutments presents different degree of dentinal sensitivity (DS) after their preparation. Protective methods of dentinal wound have the purpose to fill up the dentinal tubule...
The Performance of PEKK vs Zirconia Abutments for Screw-Retained Crowns in Two-Piece Zirconia Implants: An in vitro Study
The Performance of PEKK vs Zirconia Abutments for Screw-Retained Crowns in Two-Piece Zirconia Implants: An in vitro Study
This study aimed to compare the in vitro performance of anterior and posterior crowns with screw-retained polyetherketoneketone (PEKK) or zirconia abutments on two-piece zirconia i...
Study on friction behavior at the interface between prosthetic socket and liner
Study on friction behavior at the interface between prosthetic socket and liner
Purpose: The friction characteristics at the interface between prosthetic socket and liner have an important influence on the walking function and wearing comfort of amputees. The ...
Dynamic characterization of thickness tapered laminated composite plates
Dynamic characterization of thickness tapered laminated composite plates
This study investigates the dynamic properties of the thickness tapered laminated composite plate. The governing differential equations of motion of the various configurations of a...
Pendampingan dan Evaluasi Produk Kaki Prostetik bagi Kelainan Kongenital Kaki di Atas Lutut (Above Knee Prosthesis)
Pendampingan dan Evaluasi Produk Kaki Prostetik bagi Kelainan Kongenital Kaki di Atas Lutut (Above Knee Prosthesis)
A prosthetic leg is a lower extremity movement aid in assisting the mobility of persons with physical disabilities. A prosthetic leg is used to replace a missing leg, either due to...
Diagnostic Accuracy of Cardiac Computed Tomographic Angiography and Transesophageal echocardiography in Evaluation of Patients with Prosthetic paravalvular leakage
Diagnostic Accuracy of Cardiac Computed Tomographic Angiography and Transesophageal echocardiography in Evaluation of Patients with Prosthetic paravalvular leakage
Purpose: Computed Tomography (CT) scan has been well addressed
to provide a diagnostic information for patients with prosthetic heart
valve dysfunction. However, its role in the as...
Bearing capacity of tapered walls: Physical modeling and numerical analysis
Bearing capacity of tapered walls: Physical modeling and numerical analysis
Abstract
This study investigates the load–displacement behavior and failure mechanisms of tapered walls using numerical methods validated against experimental data. ...