SU‐E‐T‐470: Beam Performance of the Radiance 330 Proton Therapy System

View through CrossRef
Purpose:The ProTom Radiance 330 proton radiotherapy system is a fully functional, compact proton radiotherapy system that provides advanced proton delivery capabilities. It supports three‐dimensional beam scanning with energy and intensity modulation. A series of measurements have been conducted to characterize the beam performance of the first installation of the system at the McLaren Proton Therapy Center in Flint, Michigan. These measurements were part of the technical commissioning of the system. Select measurements and results are presented.Methods:The Radiance 330 proton beam energy range is 70–250 MeV for treatment, and up to 330 MeV for proton tomography and radiography. Its 3‐D scanning capability, together with a small beam emittance and momentum spread, provides a highly efficient beam delivery. During the technical commissioning, treatment plans were created to deliver uniform maps at various energies to perform Gamma Index analysis. EBT3 Gafchromic films were irradiated using the Planned irradiation maps. Bragg Peak chamber was used to test the dynamic range during a scan in one layer for high (250 MeV) and Low (70 MeV) energies. The maximum and minimum range, range adjustment and modulation, distal dose falloff (80%–20%), pencil beam spot size, spot placement accuracy were also measured. The accuracy testing included acquiring images, image registration, receiving correction vectors and applying the corrections to the robotic patient positioner.Results:Gamma Index analysis of the Treatment Planning System (TPS) data vs. Measured data showed more than 90% of points within (3%, 3mm) for the maps created by the TPS. At Isocenter Beam Size (One sigma) < 3mm at highest energy (250 MeV) in air. Beam delivery was within 0.6 mm of the intended target at the entrance and the exit of the beam, through the phantom.Conclusion:The Radiance 330 Beam Performance Measurements have confirmed that the system operates as designed with excellent clinical performance specifications.Hovakim Nazaryan, Vahagn Nazaryan and Fuhua Wang are employees of ProTom International, Inc. who contributed to the development and completed the technical commissioning of the Radiance 330 proton therapy delivery system manufactured by ProTom International.
Title: SU‐E‐T‐470: Beam Performance of the Radiance 330 Proton Therapy System
Description:
Purpose:The ProTom Radiance 330 proton radiotherapy system is a fully functional, compact proton radiotherapy system that provides advanced proton delivery capabilities.
It supports three‐dimensional beam scanning with energy and intensity modulation.
A series of measurements have been conducted to characterize the beam performance of the first installation of the system at the McLaren Proton Therapy Center in Flint, Michigan.
These measurements were part of the technical commissioning of the system.
Select measurements and results are presented.
Methods:The Radiance 330 proton beam energy range is 70–250 MeV for treatment, and up to 330 MeV for proton tomography and radiography.
Its 3‐D scanning capability, together with a small beam emittance and momentum spread, provides a highly efficient beam delivery.
During the technical commissioning, treatment plans were created to deliver uniform maps at various energies to perform Gamma Index analysis.
EBT3 Gafchromic films were irradiated using the Planned irradiation maps.
Bragg Peak chamber was used to test the dynamic range during a scan in one layer for high (250 MeV) and Low (70 MeV) energies.
The maximum and minimum range, range adjustment and modulation, distal dose falloff (80%–20%), pencil beam spot size, spot placement accuracy were also measured.
The accuracy testing included acquiring images, image registration, receiving correction vectors and applying the corrections to the robotic patient positioner.
Results:Gamma Index analysis of the Treatment Planning System (TPS) data vs.
Measured data showed more than 90% of points within (3%, 3mm) for the maps created by the TPS.
At Isocenter Beam Size (One sigma) < 3mm at highest energy (250 MeV) in air.
Beam delivery was within 0.
6 mm of the intended target at the entrance and the exit of the beam, through the phantom.
Conclusion:The Radiance 330 Beam Performance Measurements have confirmed that the system operates as designed with excellent clinical performance specifications.
Hovakim Nazaryan, Vahagn Nazaryan and Fuhua Wang are employees of ProTom International, Inc.
who contributed to the development and completed the technical commissioning of the Radiance 330 proton therapy delivery system manufactured by ProTom International.

Related Results

E-Survey: Current Status of Proton Beam Therapy in USA
E-Survey: Current Status of Proton Beam Therapy in USA
The clinical use of Proton Beam Therapy (PBT) in the management of cancer is increasing worldwide. USA has high number of proton therapy centres and this number is growing with new...
Intense-Proton-Beam Transport through an Insulator Beam Guide
Intense-Proton-Beam Transport through an Insulator Beam Guide
In this paper we study intense-proton-beam transport through an insulator guide. In our previous papers [Jpn. J. Appl. Phys. 34 (1995) L520, Jpn. J. Appl. Phys. 35 (1996) L112...
TH‐E‐ValA‐02: Simultaneous Multi‐Pencil Fan‐Beam‐Based Intensity‐Modulated Proton Therapy.
TH‐E‐ValA‐02: Simultaneous Multi‐Pencil Fan‐Beam‐Based Intensity‐Modulated Proton Therapy.
Purpose: Intensity‐modulated proton therapy (IMPT) will improve the conformality of proton radiotherapy while preserving target homogeneity and low integral dose characteristics. I...
Interaction of proton beam with human tissues in proton therapy
Interaction of proton beam with human tissues in proton therapy
Proton therapy is an effective and safe method to treat tumors in human body. Instead of conventional radiation (X-rays), this technique uses a heavy charged particles (protons) to...
Focusing effect of channel target on ultra-intense laser-accelerated proton beam
Focusing effect of channel target on ultra-intense laser-accelerated proton beam
In laser proton acceleration, the inevitable transverse divergence of proton beam restricts its applications in many fields. In this paper, a structured target with a properly wide...
Refining Atmosphere Profiles for Aerial Target Detection Models
Refining Atmosphere Profiles for Aerial Target Detection Models
Atmospheric path radiance in the infrared is an extremely important quantity in calculating system performance in certain infrared detection systems. For infrared search and track ...
Unraveling the mechanism of proton translocation in the extracellular half-channel of bacteriorhodopsin
Unraveling the mechanism of proton translocation in the extracellular half-channel of bacteriorhodopsin
AbstractBacteriorhodopsin, a light activated protein that creates a proton gradient in halobacteria, has long served as a simple model of proton pumps. Within bacteriorhodopsin, se...
Molecular Mechanism Analysis the Apoptosis of Leukemia Cells Induced By CRM1 Selective Inhibitor KPT-330
Molecular Mechanism Analysis the Apoptosis of Leukemia Cells Induced By CRM1 Selective Inhibitor KPT-330
Abstract Objective: To investigate the effect and mechanisms of KPT-330 on cell apoptosis of human leukemia cells. Methods: CCK-8 assay was used to quantify the grow...

Back to Top