Abstract 6025: Optimal Stent Deployment. Five Seconds Is Not Enough
View through CrossRef
Background
. The risk of thrombosis which had been well described since the advent of stenting, is currently being re-addressed in the era of drug-eluting stents (DES). Among the predictors of thrombosis risk, there are some factors which may be favourably influenced by the operator. Indeed, the quality of stent deployment is crucial, which has been well demonstrated with IVUS. Inadequate deployment may be associated with numerous parameters. The time factor has never been explored accurately so far.
Objectives.
The purpose of this study was to analyse the relationship between the duration of stent balloon inflation and the quality of stent deployment as measured by StentBoost
™
in patients with de novo lesions.
Methods.
StentBoost
™
is a software developed by Philips on the flat panel detector, which averages frames, correcting for motion by using the balloon markers. This creates an enhanced image of the stent with an improved signal-to-noise ratio and allows precise measurements. Pressure of stent deployment was selected according to the reference diameter of the lesion and the diameter/pressure chart of the stent delivery system. A constant pressure was applied for 25 seconds and stent size measured at 5, 15, 25 sec., after deflation and post dilatation when performed. Primary endpoint was optimal stent expansion defined as minimal stent diameter/reference diameter at the lesion site > 80%.
Results.
Data were collected prospectively in 77 patients aged 67.8± 12.4 yrs, diabetic (38%), acute coronary syndrome (37 %), in whom 109 stents (length 18.7±6.4 mm; diameter 2.96±0.49 mm) were deployed at a mean pressure of 13.5±2.48 atm. Lesion reference diameter was 3.06±0.58 mm.
Conclusion.
Better stent implantation may be achieved by increasing the duration of stent deployment. This may influence the risk of stent thrombosis.
Ovid Technologies (Wolters Kluwer Health)
Title: Abstract 6025: Optimal Stent Deployment. Five Seconds Is Not Enough
Description:
Background
.
The risk of thrombosis which had been well described since the advent of stenting, is currently being re-addressed in the era of drug-eluting stents (DES).
Among the predictors of thrombosis risk, there are some factors which may be favourably influenced by the operator.
Indeed, the quality of stent deployment is crucial, which has been well demonstrated with IVUS.
Inadequate deployment may be associated with numerous parameters.
The time factor has never been explored accurately so far.
Objectives.
The purpose of this study was to analyse the relationship between the duration of stent balloon inflation and the quality of stent deployment as measured by StentBoost
™
in patients with de novo lesions.
Methods.
StentBoost
™
is a software developed by Philips on the flat panel detector, which averages frames, correcting for motion by using the balloon markers.
This creates an enhanced image of the stent with an improved signal-to-noise ratio and allows precise measurements.
Pressure of stent deployment was selected according to the reference diameter of the lesion and the diameter/pressure chart of the stent delivery system.
A constant pressure was applied for 25 seconds and stent size measured at 5, 15, 25 sec.
, after deflation and post dilatation when performed.
Primary endpoint was optimal stent expansion defined as minimal stent diameter/reference diameter at the lesion site > 80%.
Results.
Data were collected prospectively in 77 patients aged 67.
8± 12.
4 yrs, diabetic (38%), acute coronary syndrome (37 %), in whom 109 stents (length 18.
7±6.
4 mm; diameter 2.
96±0.
49 mm) were deployed at a mean pressure of 13.
5±2.
48 atm.
Lesion reference diameter was 3.
06±0.
58 mm.
Conclusion.
Better stent implantation may be achieved by increasing the duration of stent deployment.
This may influence the risk of stent thrombosis.
Related Results
CLINICAL APPLICATION OF DOMESTIC LONG BIODEGRADABLE POLYMER COATED SIROLIMUS ELUTING STENTS FOR TREATMENT OF LONG CORONARY ARTERY LESIONS
CLINICAL APPLICATION OF DOMESTIC LONG BIODEGRADABLE POLYMER COATED SIROLIMUS ELUTING STENTS FOR TREATMENT OF LONG CORONARY ARTERY LESIONS
Objectives
To evaluate the efficacy and safety of domestic long biodegradable polymer coated sirolimus eluting stents (Excel stent, Shandong JW) for treatment of ...
Minimum stent area, stent expansion and ultrasonic flow ratio post pci: insights from multivessel talent trial
Minimum stent area, stent expansion and ultrasonic flow ratio post pci: insights from multivessel talent trial
Abstract
Background
Ultrasonic flow ratio (UFR) is a novel method for fast computation of fractional flow reserve (FFR) from int...
Value of stent boost imaging in decision making after coronary stenting
Value of stent boost imaging in decision making after coronary stenting
AbstractBackground: Several studies reported the comparability of digital stent enhancement techniques (including stent boost imaging) in detecting suboptimal results of coronary s...
Comparative outcome analysis of enterprise and neuroform stent-assisted coiling of cerebral aneurysms: A review of the literature
Comparative outcome analysis of enterprise and neuroform stent-assisted coiling of cerebral aneurysms: A review of the literature
Introduction: One of the popular treatment strategies for complex cerebral aneurysms with wide necks or low dome-to-neck ratios is stent-assisted coiling. The most widely used intr...
P5637A novel bioresorbable magnesium-based stent: initial clinical results and 6-months follow-up by coronary computed tomography angiography scan
P5637A novel bioresorbable magnesium-based stent: initial clinical results and 6-months follow-up by coronary computed tomography angiography scan
Abstract
Background
Drug-eluting stents (DES) are the gold standard in percutaneous coronary interventions (PCI), but leave a pe...
Long-term efficacy of stent placement for treating inferior vena cava stenosis following liver transplantation
Long-term efficacy of stent placement for treating inferior vena cava stenosis following liver transplantation
The aims of this study were to evaluate both the efficacy of stent placement for treating inferior vena cava (IVC) stenosis and the patency of hepatic veins (HVs) following IVC ste...
Predictors of insistent restenosis in patients undergoing percutaneous intervention
Predictors of insistent restenosis in patients undergoing percutaneous intervention
Coronary artery disease is one of the leading causes of morbidity and mortality worldwide. The first percutaneous coronary intervention was performed by Andreas Gruntzig on Septemb...
Broken stent of endoscopic retrograde cholangiopancreatography mimicking a common bile duct stone: a case report
Broken stent of endoscopic retrograde cholangiopancreatography mimicking a common bile duct stone: a case report
Long standing biliary stent for biliary stricture may have complications like cholangitis, cholecystitis, stent fracture and stent migration. Treatment includes re-do endoscopic re...