Comparative evaluation of the accuracy of the dental arch display using modern intraoral three-dimensional scanners
View through CrossRef
BACKGROUND: At present, modern dentists used dental intraoral three-dimensional (3D) scanners routinely in their daily work. Obtaining an optical 3D image of the teeth and dentition helps avoid errors at the stage of obtaining traditional silicone impressions and significantly reduces the level of discomfort during dental procedures. Intraoral scanner systems are commercially available today. Despite their advantages over traditional silicone impressions, the accuracy of the optical impressions obtained during total and subtotal prosthetics on the upper and lower jaw are still questionable.
AIM: This study aimed to evaluate the accuracy of scanning models of the patients dentition obtained using intraoral scanners and to determine the currently available models of digital devices that are optimal in terms of price and quality.
MATERIAL AND METHODS: The reference scan (master model) was an optical image of the reference model of the upper jaw of a patient with a full dentition made by additive 3D-printing technology obtained using a laboratory scanner. To level out the shrinkage errors of photopolymerization materials, the resulting model was scanned on the same day using intraoral scanners selected for the study. The obtained scan results were compared with the reference scan of the control group. Based on the measurements made at reference points, the average error value relative to the master model and its precision were calculated for each scanner. In addition to the average values, the error parameters of the scanners in the frontal and chewing sections were calculated, as well as the arc error, which was equal to the difference in the values of the discrepancy at the reference points at which the maximum and minimum values were obtained. The cost and availability of scanners on the domestic market, under the sanctions policy of foreign countries, were also considered.
RESULTS: The CEREC Primescan AC scanner showed the best accuracy according to the results of model discrepancies. It has an average error of 13.727.34. The arc error was 18.8 microns, and the discrepancies in the frontal area and chewing area were 18 and 6.8 microns, respectively. 3Shape Trios 3 scanner has the closest accuracy, with an average error of 16.285.94 microns. The error value of Aoralscan 3 was 42.0818.34 microns, its arc accuracy was 65 microns, and the discrepancies in the frontal and chewing areas were 33 and 55.1 microns, respectively. Emerald S Mode C had an average error of 35.8422.29 microns, which was higher than that of Medit i500 and Aoralscan 3; however, Aoralscan 3 showed better precision (18.34 microns versus 22.29 microns). According to the comparison results of the models in the MeshLab program, CEREC Primescan AC showed the smallest median of distances (18 microns). The TRIOS 3 and Emerald S Mode C differed from the standard by an average of 29 microns, and Aoralscan 3 scanners had of 33 microns and Medit i500 had 41 microns.
CONCLUSION: The precision of Aoralscan 3 scanner, which is the best among its analogs in the price category, makes it the most affordable scanner for dental surgical, orthodontic, and orthopedic fields. CEREC Primescan AC is the leader of the premium scanners involved in the study and available in the dental market.
Title: Comparative evaluation of the accuracy of the dental arch display using modern intraoral three-dimensional scanners
Description:
BACKGROUND: At present, modern dentists used dental intraoral three-dimensional (3D) scanners routinely in their daily work.
Obtaining an optical 3D image of the teeth and dentition helps avoid errors at the stage of obtaining traditional silicone impressions and significantly reduces the level of discomfort during dental procedures.
Intraoral scanner systems are commercially available today.
Despite their advantages over traditional silicone impressions, the accuracy of the optical impressions obtained during total and subtotal prosthetics on the upper and lower jaw are still questionable.
AIM: This study aimed to evaluate the accuracy of scanning models of the patients dentition obtained using intraoral scanners and to determine the currently available models of digital devices that are optimal in terms of price and quality.
MATERIAL AND METHODS: The reference scan (master model) was an optical image of the reference model of the upper jaw of a patient with a full dentition made by additive 3D-printing technology obtained using a laboratory scanner.
To level out the shrinkage errors of photopolymerization materials, the resulting model was scanned on the same day using intraoral scanners selected for the study.
The obtained scan results were compared with the reference scan of the control group.
Based on the measurements made at reference points, the average error value relative to the master model and its precision were calculated for each scanner.
In addition to the average values, the error parameters of the scanners in the frontal and chewing sections were calculated, as well as the arc error, which was equal to the difference in the values of the discrepancy at the reference points at which the maximum and minimum values were obtained.
The cost and availability of scanners on the domestic market, under the sanctions policy of foreign countries, were also considered.
RESULTS: The CEREC Primescan AC scanner showed the best accuracy according to the results of model discrepancies.
It has an average error of 13.
727.
34.
The arc error was 18.
8 microns, and the discrepancies in the frontal area and chewing area were 18 and 6.
8 microns, respectively.
3Shape Trios 3 scanner has the closest accuracy, with an average error of 16.
285.
94 microns.
The error value of Aoralscan 3 was 42.
0818.
34 microns, its arc accuracy was 65 microns, and the discrepancies in the frontal and chewing areas were 33 and 55.
1 microns, respectively.
Emerald S Mode C had an average error of 35.
8422.
29 microns, which was higher than that of Medit i500 and Aoralscan 3; however, Aoralscan 3 showed better precision (18.
34 microns versus 22.
29 microns).
According to the comparison results of the models in the MeshLab program, CEREC Primescan AC showed the smallest median of distances (18 microns).
The TRIOS 3 and Emerald S Mode C differed from the standard by an average of 29 microns, and Aoralscan 3 scanners had of 33 microns and Medit i500 had 41 microns.
CONCLUSION: The precision of Aoralscan 3 scanner, which is the best among its analogs in the price category, makes it the most affordable scanner for dental surgical, orthodontic, and orthopedic fields.
CEREC Primescan AC is the leader of the premium scanners involved in the study and available in the dental market.
Related Results
Primerjalna književnost na prelomu tisočletja
Primerjalna književnost na prelomu tisočletja
In a comprehensive and at times critical manner, this volume seeks to shed light on the development of events in Western (i.e., European and North American) comparative literature ...
Management of Oblique Root Fracture Using Mineral Trioxide Aggregate: A Case Report
Management of Oblique Root Fracture Using Mineral Trioxide Aggregate: A Case Report
Root fractures are relatively uncommon compared to other types of dental traumas. It is sometimes extremely difficult for practitioners to decide what should be done and which appr...
Guest Editorial
Guest Editorial
Dental caries is one of the major health problems in Indonesia. Data from Indonesian Basic Health Research in 2013, 2015 and 2018 showed a consistent increase in the prevalence of ...
Effect of Different Preparation Depths for an Inlay‐Retained Fixed Partial Denture on the Accuracy of Different Intraoral Scanners: An In Vitro Study
Effect of Different Preparation Depths for an Inlay‐Retained Fixed Partial Denture on the Accuracy of Different Intraoral Scanners: An In Vitro Study
AbstractPurposeThe aim was to evaluate the effect of different preparation depths for inlay‐retained fixed partial dentures on the accuracy of intraoral scanners.Materials and Meth...
In-plane bearing capacity analysis of concrete catenary arch with rigid skeleton: test and simulation
In-plane bearing capacity analysis of concrete catenary arch with rigid skeleton: test and simulation
Abstract
Although the rigid skeleton concrete arch has been widely used in engineering, the research on the failure mode and bearing capacity of this kind of arch is very...
Determination of Mandibular Arch Form During Mixed Dentition Using Custom Made Template in Sulaimani Governorate
Determination of Mandibular Arch Form During Mixed Dentition Using Custom Made Template in Sulaimani Governorate
Objectives: The dental arch form and dimensions change continuously and systematically during human growth and development, but this decreases adulthood. During mixed dentition, ch...
Accuracy and Reliability of Digital Dental Models Obtained by Intraoral Scans Compared with Plaster Models
Accuracy and Reliability of Digital Dental Models Obtained by Intraoral Scans Compared with Plaster Models
Purposes: In orthodontics, determining the parameters of tooth size, dental arch, and conducting Bolton analysis is crucial for diagnosis, treatment planning, and patient outcomes....
EVALUATION OF MAXILLARY ARCH ANALYSIS IN MODINAGAR ADULTS WITH NORMO-OCCLUSION IN VIVO STUDY
EVALUATION OF MAXILLARY ARCH ANALYSIS IN MODINAGAR ADULTS WITH NORMO-OCCLUSION IN VIVO STUDY
Purpose: The aim of this study is to evaluatemaxillary arch analysis in Modinagar adultswith normo-occlusion. Materials and Methods: A total of 100 patients aged between 20 to 30 ...