Assessment of visual, manual, and automatic coronary calcium scoring methods - analysis of dedicated CT scans and low dose CT scans acquired during cardiac 13N-ammonia-PET/CT
View through CrossRef
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
Coronary artery calcium (CAC) is a well-known predictor of major adverse cardiac events and is scored manually from dedicated, ECG-triggered CT scans. In the present study, we investigated the accuracy of risk categorisation based on visual and automatic AI calcium scoring from low dose CT (LDCT) scans and dedicated Calcium Score CT (CSCT) scans.
Purpose
To assess the agreement of risk prediction based on visual and automatic AI CAC scoring from CSCT scans and LDCT scans as compared to a gold standard, manual CSCT scoring.
Methods
We retrospectively enrolled 222 patients. Each patient received a 13N-ammonia PET with LDCT and CSCT scan. The time interval between LDCT and CSCT was less than 6 months.
Each LDCT and CSCT scan was scored visually, manually, and automatically with AI. For visual scoring we used a previously described 6–points scale (0; 1-10; 11-100; 101-400; 401-100; >1000 Agatston score). For manual scoring we used a generally available software package (Syngo.via,Siemens). The automatic AI scoring was performed with commercially available software based on a deep learning algorithm (included in Syngo.via,Siemens). Each manually and automatically measured Agatston score was converted into the 6-points scale.
We performed a per patient analysis; the risk group categorization was based on the total Agatston score. Spearman correlation coefficient was used to analyse the association between manual and automatic AI scoring methods. Agreement between visual, manual, and automatic AI scoring methods was determined using weighted kappa test with 95% confidence intervals (95%CI).
Results
The correlation between manual scoring from LDCT and CSCT scans was 0.96 (p < 0.001).The agreement between manual scoring from two scans, however, was low with weighted kappa equal 0.57 (95% CI 0.51 – 0.63). 91,9% of calcium scores measured by AI software on CSCT were in the same risk group as manual CSCT scores.The agreement between AI scoring and manual scoring using CSCT was excellent, the weighted kappa was equal 0.95 (95% CI 0.92 - 0.97).Based on visual scoring on LDCT scans, 74,3% of the scores were in the same category as manual scoring on CSCT scans. The agreement between the visual scoring on LDCT scans and a gold standard was strong, weighted kappa equal was 0.82 (95% CI 0.77 – 0.86). The agreement between manual and automatic scoring on LDCT using manual CSCT as the gold standard was low (0.57, 95 % CI 0.51 – 0.63; 0.49, 95 % CI 0.43 – 0.56, respectively). Based on visual LDCT scoring, 7 patients were incorrectly classified as calcium score 0, which underestimated the overall patients’ risk.The AI method scoring CSCT scans, classified 2 patients incorrectly as non-calcium risk group.
Conclusions
CAC can be automatically assessed from CSCT scans with commercially available AI software.Of manual, automatic, and visual CAC scoring on LDCT scans the visual scoring showed the highest agreement with the gold standard manual CSCT CAC scoring.
Oxford University Press (OUP)
Title: Assessment of visual, manual, and automatic coronary calcium scoring methods - analysis of dedicated CT scans and low dose CT scans acquired during cardiac 13N-ammonia-PET/CT
Description:
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
Coronary artery calcium (CAC) is a well-known predictor of major adverse cardiac events and is scored manually from dedicated, ECG-triggered CT scans.
In the present study, we investigated the accuracy of risk categorisation based on visual and automatic AI calcium scoring from low dose CT (LDCT) scans and dedicated Calcium Score CT (CSCT) scans.
Purpose
To assess the agreement of risk prediction based on visual and automatic AI CAC scoring from CSCT scans and LDCT scans as compared to a gold standard, manual CSCT scoring.
Methods
We retrospectively enrolled 222 patients.
Each patient received a 13N-ammonia PET with LDCT and CSCT scan.
The time interval between LDCT and CSCT was less than 6 months.
Each LDCT and CSCT scan was scored visually, manually, and automatically with AI.
For visual scoring we used a previously described 6–points scale (0; 1-10; 11-100; 101-400; 401-100; >1000 Agatston score).
For manual scoring we used a generally available software package (Syngo.
via,Siemens).
The automatic AI scoring was performed with commercially available software based on a deep learning algorithm (included in Syngo.
via,Siemens).
Each manually and automatically measured Agatston score was converted into the 6-points scale.
We performed a per patient analysis; the risk group categorization was based on the total Agatston score.
Spearman correlation coefficient was used to analyse the association between manual and automatic AI scoring methods.
Agreement between visual, manual, and automatic AI scoring methods was determined using weighted kappa test with 95% confidence intervals (95%CI).
Results
The correlation between manual scoring from LDCT and CSCT scans was 0.
96 (p < 0.
001).
The agreement between manual scoring from two scans, however, was low with weighted kappa equal 0.
57 (95% CI 0.
51 – 0.
63).
91,9% of calcium scores measured by AI software on CSCT were in the same risk group as manual CSCT scores.
The agreement between AI scoring and manual scoring using CSCT was excellent, the weighted kappa was equal 0.
95 (95% CI 0.
92 - 0.
97).
Based on visual scoring on LDCT scans, 74,3% of the scores were in the same category as manual scoring on CSCT scans.
The agreement between the visual scoring on LDCT scans and a gold standard was strong, weighted kappa equal was 0.
82 (95% CI 0.
77 – 0.
86).
The agreement between manual and automatic scoring on LDCT using manual CSCT as the gold standard was low (0.
57, 95 % CI 0.
51 – 0.
63; 0.
49, 95 % CI 0.
43 – 0.
56, respectively).
Based on visual LDCT scoring, 7 patients were incorrectly classified as calcium score 0, which underestimated the overall patients’ risk.
The AI method scoring CSCT scans, classified 2 patients incorrectly as non-calcium risk group.
Conclusions
CAC can be automatically assessed from CSCT scans with commercially available AI software.
Of manual, automatic, and visual CAC scoring on LDCT scans the visual scoring showed the highest agreement with the gold standard manual CSCT CAC scoring.
Related Results
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Objective: To determine the frequency of common chromosomal aberrations in local population idiopathic determine the frequency of common chromosomal aberrations in local population...
SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
A enfermagem é o pilar da assistência pois está na linha de frente do cuidado holístico, todavia esta é estigmatizada e desvalorizada, assim como não possui reconhecimento consider...
Experimental and numerical study on ammonia fueled compression ignition engine
Experimental and numerical study on ammonia fueled compression ignition engine
Ammonia is currently receiving more interest as a carbon-free alternative fuel for internal combustion engines. A promising energy carrier, easy to store and transport, being liqui...
Evolution of Antimicrobial Resistance in Community vs. Hospital-Acquired Infections
Evolution of Antimicrobial Resistance in Community vs. Hospital-Acquired Infections
Abstract
Introduction
Hospitals are high-risk environments for infections. Despite the global recognition of these pathogens, few studies compare microorganisms from community-acqu...
Radiation Exposure to the Personnel Performing Myocardial Blood Flow Quantification Study Using 13N-ammonia Positron Emission Tomography/Computed Tomography
Radiation Exposure to the Personnel Performing Myocardial Blood Flow Quantification Study Using 13N-ammonia Positron Emission Tomography/Computed Tomography
Purpose:
The present study aimed to evaluate radiation exposure to staff performing coronary flow reserve (CFR) measurement using 13N-ammonia.
...
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
ObjectiveTo compare the diagnostic differences between total-body PET/CT (positron emission tomography/computed tomography) and PET/MR (positron emission tomography/magnetic resona...
British Food Journal Volume 45 Issue 9 1943
British Food Journal Volume 45 Issue 9 1943
I now pass on to an aspect of calcium metabolism which is more topical, but probably more controversial. I refer to the incidence of calcium deficiency. By what means can we determ...
Sensor Gas Amonia Berbasis Polimer Konduktif Polianilina: Sebuah Review
Sensor Gas Amonia Berbasis Polimer Konduktif Polianilina: Sebuah Review
Artikel review ini memberikan informasi tentang aplikasi polianilina (PANI) dan kompositnya sebagai sensor gas berbahaya khususnya amonia (NH3). Kajian yang dibahas pada artikel in...