X-Ray Protective Aprons Re-Evaluated
View through CrossRef
Background The evaluation of the protective effect of X-ray protective clothing requires new criteria. The current concept assumes more or less uniform covering of the torso with protective material. The frequently worn heavy wrap-around aprons can weigh 7 to 8 kg. As relevant studies show, orthopedic damage can result from long-term activity. It should therefore be investigated whether the apron weight can be reduced by optimizing the material distribution. For a radiobiological evaluation of the protective effect, the “effective dose” should be used.
Methods Numerous laboratory measurements were performed with an Alderson Rando phantom as well as dose measurements on clinical personnel. The measurements were supplemented by Monte Carlo simulation of an interventional workplace in which a female ICRP reference phantom was used for the operator. The measured back doses on the Alderson phantom as well as the measured back doses at interventional workplaces were based on the personal equivalent dose Hp(10). Monte Carlo simulations were used to introduce protection factors for the protective clothing based on the “effective dose” introduced in radiation protection.
Results Back doses in clinical radiology personnel are largely negligible. Therefore, back protection can be much lower than currently used or can even be eliminated. The Monte Carlo simulations show that the protective effect of protective aprons worn on the body is higher than when the flat protective material is radiated through (3 D effect). About 80 % of the effective dose is attributed to the body region from the gonads to the chest. By additional shielding of this area, the effective dose can be lowered or, optionally, aprons with less weight can be produced. Attention must also be paid to the “radiation leaks” (upper arms, neck, skull), which can reduce the whole-body protective effect.
Conclusion In the future, the evaluation of the protective effect of X-ray protective clothing should be based on the effective dose. For this purpose, effective dose-based protection factors could be introduced, while the lead equivalent should be used for measurement purposes only. If the results are implemented, protective aprons with approx. 40 % less weight can be produced with a comparable protective effect.
Key Points:
Citation Format
Title: X-Ray Protective Aprons Re-Evaluated
Description:
Background The evaluation of the protective effect of X-ray protective clothing requires new criteria.
The current concept assumes more or less uniform covering of the torso with protective material.
The frequently worn heavy wrap-around aprons can weigh 7 to 8 kg.
As relevant studies show, orthopedic damage can result from long-term activity.
It should therefore be investigated whether the apron weight can be reduced by optimizing the material distribution.
For a radiobiological evaluation of the protective effect, the “effective dose” should be used.
Methods Numerous laboratory measurements were performed with an Alderson Rando phantom as well as dose measurements on clinical personnel.
The measurements were supplemented by Monte Carlo simulation of an interventional workplace in which a female ICRP reference phantom was used for the operator.
The measured back doses on the Alderson phantom as well as the measured back doses at interventional workplaces were based on the personal equivalent dose Hp(10).
Monte Carlo simulations were used to introduce protection factors for the protective clothing based on the “effective dose” introduced in radiation protection.
Results Back doses in clinical radiology personnel are largely negligible.
Therefore, back protection can be much lower than currently used or can even be eliminated.
The Monte Carlo simulations show that the protective effect of protective aprons worn on the body is higher than when the flat protective material is radiated through (3 D effect).
About 80 % of the effective dose is attributed to the body region from the gonads to the chest.
By additional shielding of this area, the effective dose can be lowered or, optionally, aprons with less weight can be produced.
Attention must also be paid to the “radiation leaks” (upper arms, neck, skull), which can reduce the whole-body protective effect.
Conclusion In the future, the evaluation of the protective effect of X-ray protective clothing should be based on the effective dose.
For this purpose, effective dose-based protection factors could be introduced, while the lead equivalent should be used for measurement purposes only.
If the results are implemented, protective aprons with approx.
40 % less weight can be produced with a comparable protective effect.
Key Points:
Citation Format.
Related Results
A method of calibrating effective area of focusing X-ray detector by using normal spectrum of Crab pulsar
A method of calibrating effective area of focusing X-ray detector by using normal spectrum of Crab pulsar
X-ray detector is a core component for X-ray astronomical observation and pulsar navigation.The on-orbit observation performance of X-ray detector will change gradually,owing to th...
INSPEKSI ALAT PELINDUNG DIRI (APD) RADIASI MENGGUNAKAN METODE UJI VISUAL, PALPASI, DAN SINAR-X
INSPEKSI ALAT PELINDUNG DIRI (APD) RADIASI MENGGUNAKAN METODE UJI VISUAL, PALPASI, DAN SINAR-X
Radiation personal protective equipment (PPE) is highly susceptible to defects and damage, compromising the safety of healthcare workers and patients. The purpose of this study was...
New X-ray supernova remnants in NGC 7793
New X-ray supernova remnants in NGC 7793
Context. This work focuses on the detection of X-ray supernova remnants (SNRs) in the galaxy NGC 7793 and the study of their properties.
Aims. X-ray SNRs in galaxies beyond the Loc...
X-ray focusing optics and its application in X-ray communication system
X-ray focusing optics and its application in X-ray communication system
X-ray communication, which was first introduced by Keith Gendreau in 2007, is potential to compete with conventional communication methods, such as microware and laser communicatio...
Microfocus X-ray Tubes
Microfocus X-ray Tubes
Introduction. X-ray inspection plays a unique role among all nondestructive testing methods for products and materials due to sufficiently high resolution and high penetrability. T...
Utilization of personal protective equipment and its key factors among WA oil factory workers in Debre Markos town, Ethiopia
Utilization of personal protective equipment and its key factors among WA oil factory workers in Debre Markos town, Ethiopia
BackgroundEdible oil manufacturing is a labor-intensive sector with significant technological demands, where employees face various occupational hazards. The use of personal protec...
Joint Rapid Airfield Construction (JRAC) Program 2004 Demonstration Project--Fort Bragg, North Carolina
Joint Rapid Airfield Construction (JRAC) Program 2004 Demonstration Project--Fort Bragg, North Carolina
This report describes the demonstration of technologies and procedures developed during April 2002 and May 2004 under the Joint Rapid Airfield Construction (JRAC) Program. The demo...
Exploring Hard X-Ray Properties of Gamma-Ray Emitting Narrow-line Seyfert I Galaxies through NuSTAR Observations
Exploring Hard X-Ray Properties of Gamma-Ray Emitting Narrow-line Seyfert I Galaxies through NuSTAR Observations
Abstract
We studied the six gamma-ray-detected Narrow Line Seyfert 1 (NLSy1) galaxies using the hard X-ray observations from Nuclear Spectros...