X-Ray Protective Aprons Re-Evaluated
View through CrossRef
Background The evaluation of the protective effect of X-ray protective clothing requires new criteria. The current concept assumes more or less uniform covering of the torso with protective material. The frequently worn heavy wrap-around aprons can weigh 7 to 8 kg. As relevant studies show, orthopedic damage can result from long-term activity. It should therefore be investigated whether the apron weight can be reduced by optimizing the material distribution. For a radiobiological evaluation of the protective effect, the “effective dose” should be used.
Methods Numerous laboratory measurements were performed with an Alderson Rando phantom as well as dose measurements on clinical personnel. The measurements were supplemented by Monte Carlo simulation of an interventional workplace in which a female ICRP reference phantom was used for the operator. The measured back doses on the Alderson phantom as well as the measured back doses at interventional workplaces were based on the personal equivalent dose Hp(10). Monte Carlo simulations were used to introduce protection factors for the protective clothing based on the “effective dose” introduced in radiation protection.
Results Back doses in clinical radiology personnel are largely negligible. Therefore, back protection can be much lower than currently used or can even be eliminated. The Monte Carlo simulations show that the protective effect of protective aprons worn on the body is higher than when the flat protective material is radiated through (3 D effect). About 80 % of the effective dose is attributed to the body region from the gonads to the chest. By additional shielding of this area, the effective dose can be lowered or, optionally, aprons with less weight can be produced. Attention must also be paid to the “radiation leaks” (upper arms, neck, skull), which can reduce the whole-body protective effect.
Conclusion In the future, the evaluation of the protective effect of X-ray protective clothing should be based on the effective dose. For this purpose, effective dose-based protection factors could be introduced, while the lead equivalent should be used for measurement purposes only. If the results are implemented, protective aprons with approx. 40 % less weight can be produced with a comparable protective effect.
Key Points:
Citation Format
Title: X-Ray Protective Aprons Re-Evaluated
Description:
Background The evaluation of the protective effect of X-ray protective clothing requires new criteria.
The current concept assumes more or less uniform covering of the torso with protective material.
The frequently worn heavy wrap-around aprons can weigh 7 to 8 kg.
As relevant studies show, orthopedic damage can result from long-term activity.
It should therefore be investigated whether the apron weight can be reduced by optimizing the material distribution.
For a radiobiological evaluation of the protective effect, the “effective dose” should be used.
Methods Numerous laboratory measurements were performed with an Alderson Rando phantom as well as dose measurements on clinical personnel.
The measurements were supplemented by Monte Carlo simulation of an interventional workplace in which a female ICRP reference phantom was used for the operator.
The measured back doses on the Alderson phantom as well as the measured back doses at interventional workplaces were based on the personal equivalent dose Hp(10).
Monte Carlo simulations were used to introduce protection factors for the protective clothing based on the “effective dose” introduced in radiation protection.
Results Back doses in clinical radiology personnel are largely negligible.
Therefore, back protection can be much lower than currently used or can even be eliminated.
The Monte Carlo simulations show that the protective effect of protective aprons worn on the body is higher than when the flat protective material is radiated through (3 D effect).
About 80 % of the effective dose is attributed to the body region from the gonads to the chest.
By additional shielding of this area, the effective dose can be lowered or, optionally, aprons with less weight can be produced.
Attention must also be paid to the “radiation leaks” (upper arms, neck, skull), which can reduce the whole-body protective effect.
Conclusion In the future, the evaluation of the protective effect of X-ray protective clothing should be based on the effective dose.
For this purpose, effective dose-based protection factors could be introduced, while the lead equivalent should be used for measurement purposes only.
If the results are implemented, protective aprons with approx.
40 % less weight can be produced with a comparable protective effect.
Key Points:
Citation Format.
Related Results
INSPEKSI ALAT PELINDUNG DIRI (APD) RADIASI MENGGUNAKAN METODE UJI VISUAL, PALPASI, DAN SINAR-X
INSPEKSI ALAT PELINDUNG DIRI (APD) RADIASI MENGGUNAKAN METODE UJI VISUAL, PALPASI, DAN SINAR-X
Radiation personal protective equipment (PPE) is highly susceptible to defects and damage, compromising the safety of healthcare workers and patients. The purpose of this study was...
Joint Rapid Airfield Construction (JRAC) Program 2004 Demonstration Project--Fort Bragg, North Carolina
Joint Rapid Airfield Construction (JRAC) Program 2004 Demonstration Project--Fort Bragg, North Carolina
This report describes the demonstration of technologies and procedures developed during April 2002 and May 2004 under the Joint Rapid Airfield Construction (JRAC) Program. The demo...
(Invited) CdTe/Si-Roic Staced X-Ray Imaging Sensor
(Invited) CdTe/Si-Roic Staced X-Ray Imaging Sensor
CdTe photon-counting X-ray image sensor with charge-counting readout IC was developed for high quality X-ray imaging. CdTe is one of the good material for X-ray imaging, because it...
Quantum efficiency calibration for low energy detector in hard X-ray modulation telescope satellite
Quantum efficiency calibration for low energy detector in hard X-ray modulation telescope satellite
Low energy X-ray telescope, working over 0.7-15 keV energy band, is one of the main payloads in the hard X-ray modulation telescope satellite. The primary scientific objectives are...
Research on erosion mechanism and life prediction of Underground protective tube in deep heavy oil recovery
Research on erosion mechanism and life prediction of Underground protective tube in deep heavy oil recovery
Abstract
Under the dual-carbon context, traditional heavy oil extraction methods exhibit limited effectiveness in deep heavy oil development. The downhole high-power...
Protective Forests for Ecosystem-based Disaster Risk Reduction (Eco-DRR) in the Alpine Space
Protective Forests for Ecosystem-based Disaster Risk Reduction (Eco-DRR) in the Alpine Space
Mountain forests are an efficient Forest-based Solution (FbS) for Ecosystem-based Disaster Risk Reduction (Eco-DRR) by lowering the frequency, magnitude, and/or intensity of natura...
MANAJEMEN SATGAS PASIEN COVID (PENGGUNAAN APD) DI RUMAH SEHAT APARMA UNIPDU
MANAJEMEN SATGAS PASIEN COVID (PENGGUNAAN APD) DI RUMAH SEHAT APARMA UNIPDU
The condition of the spread of Covid-19 has reached almost the entire world with the number of cases continuing to increase. Health workers act as service providers, therefore hosp...
Selected Protective Clothing for Semiconductor Manufacture
Selected Protective Clothing for Semiconductor Manufacture
State-of-the-art technologies in the semiconductor industry require exotic processing techniques using unusual, hazardous chemicals. The very nature of some of these chamicals make...