Ultrasound vs. standard radiography to determine peripherally-inserted central catheter tip location

View through CrossRef
Background The use of a peripherally-inserted central catheter (PICC) has increased in preterm neonates to facilitate the administration of total parenteral nutrition. Standard radiography (thoracoabdominal X-ray) is the gold standard for determining the position of the PICC tip. However, radiography is not always accurate, influenced by the position of the extremities and anatomic variations, time-consuming procedural process, involves radiation, and is costly. Ultrasonography (USG) may serve as an easier, safer, less costly, and more real-time alternative in the neonatal intensive care unit (NICU) patients. Objective To assess the accuracy of USG use in determining PICC tip position compared to that of standard radiography. Methods This diagnostic study was conducted in the NICU at Dr. Cipto Mangunkusumo Hospital, Jakarta. The PICCs were placed using standard NICU procedure, then the tip position was evaluated using the USG immediately before standard radiography was performed. A 2x2 table was constructed to compare the diagnostic accuracy of the two modalities. Results A total of 29 neonates were included in our study. Subjects’ mean gestational age and weight were 31.7 weeks and 1,618.9 g respectively. Concordance of PICC tip positioning between standard radiography and USG occurred in 27 neonates (93.1%). USG had 88.89% sensitivity, 95% specificity, and 93.1% diagnostic accuracy. Conclusion USG has excellent diagnostic accuracy for confirmation of the PICC tip position.
Title: Ultrasound vs. standard radiography to determine peripherally-inserted central catheter tip location
Description:
Background The use of a peripherally-inserted central catheter (PICC) has increased in preterm neonates to facilitate the administration of total parenteral nutrition.
Standard radiography (thoracoabdominal X-ray) is the gold standard for determining the position of the PICC tip.
However, radiography is not always accurate, influenced by the position of the extremities and anatomic variations, time-consuming procedural process, involves radiation, and is costly.
Ultrasonography (USG) may serve as an easier, safer, less costly, and more real-time alternative in the neonatal intensive care unit (NICU) patients.
Objective To assess the accuracy of USG use in determining PICC tip position compared to that of standard radiography.
Methods This diagnostic study was conducted in the NICU at Dr.
Cipto Mangunkusumo Hospital, Jakarta.
The PICCs were placed using standard NICU procedure, then the tip position was evaluated using the USG immediately before standard radiography was performed.
A 2x2 table was constructed to compare the diagnostic accuracy of the two modalities.
Results A total of 29 neonates were included in our study.
Subjects’ mean gestational age and weight were 31.
7 weeks and 1,618.
9 g respectively.
Concordance of PICC tip positioning between standard radiography and USG occurred in 27 neonates (93.
1%).
USG had 88.
89% sensitivity, 95% specificity, and 93.
1% diagnostic accuracy.
Conclusion USG has excellent diagnostic accuracy for confirmation of the PICC tip position.

Related Results

Ultrasound-guided central line insertion in children: how much radiation is really needed?
Ultrasound-guided central line insertion in children: how much radiation is really needed?
Background: A recent survey revealed that most pediatric surgeons use intraoperative fluoroscopy and routine postoperative chest radiography for catheter tip location in central li...
Sodium Citrate Versus Heparin Locking Solutions for Tunneled Hemodialysis Catheters
Sodium Citrate Versus Heparin Locking Solutions for Tunneled Hemodialysis Catheters
Introduction: Hemodialysis catheter dysfunction compromises dialysis adequacy and may limit catheter survival. Strategies to prevent hemodialysis catheter dysfunction include cathe...
Epidural Catheter Reconnection
Epidural Catheter Reconnection
Background An in vitro model of epidural catheter contamination was used to determine if disconnected catheters can be safely reconnected. ...

Back to Top