Analysis of Dosimetric Parameters of Linear Accelerator

View through CrossRef
With the advancements in the radiation delivery techniques and modern Linac systems, the need for better quality control devices also arises. Different devices manufactured by companies are available at hospitals and some of these devices are found to be more accurate in one category then others. The main objective of this study to analyze the dosimetric parameters of linear accelerator was to use PTW QUICKCHECK device at radiotherapy department of BINO hospital, Bahawalpur and evaluate their performances in checking the beam uniformity and symmetry in daily QC and other required periodic QC tests. For Daily Quality Control PTW QUICKCHECK device was used daily in the morning checks for 50 days to monitor CAX, beam flatness, GT symmetry, LR symmetry, Beam Quality Factor for electron beam of 6, 9, 12, 15, 18, 22 MeV energies and photon beam of 6, 15 MV energies with 100 MU given to the QUICK CHECK device at dose rate of 300 MU/min. To ensure the stability of data monitored through QUICKCHEC repeatability and reproducibility tests were performed. PTW QUICKCHECK device can be easily setup on daily basis for daily checks. According to the results it is clear that PTW QUICKCHECK device is quite accurate with regard to symmetry measurements as all data is within tolerance range (3%). However, accuracy in flatness measurement shows uncertainties i.e for 6 MV 7.3%, for 15 MV 7.31%, for 6 MeV 16.12%, for 9 MeV 6.92%, for 12 MeV 5.92%, for 15 MeV 4.01%, for 18 MeV 4.01% and for 22 MeV 4.13% of data are within tolerance range.
Title: Analysis of Dosimetric Parameters of Linear Accelerator
Description:
With the advancements in the radiation delivery techniques and modern Linac systems, the need for better quality control devices also arises.
Different devices manufactured by companies are available at hospitals and some of these devices are found to be more accurate in one category then others.
The main objective of this study to analyze the dosimetric parameters of linear accelerator was to use PTW QUICKCHECK device at radiotherapy department of BINO hospital, Bahawalpur and evaluate their performances in checking the beam uniformity and symmetry in daily QC and other required periodic QC tests.
For Daily Quality Control PTW QUICKCHECK device was used daily in the morning checks for 50 days to monitor CAX, beam flatness, GT symmetry, LR symmetry, Beam Quality Factor for electron beam of 6, 9, 12, 15, 18, 22 MeV energies and photon beam of 6, 15 MV energies with 100 MU given to the QUICK CHECK device at dose rate of 300 MU/min.
To ensure the stability of data monitored through QUICKCHEC repeatability and reproducibility tests were performed.
PTW QUICKCHECK device can be easily setup on daily basis for daily checks.
According to the results it is clear that PTW QUICKCHECK device is quite accurate with regard to symmetry measurements as all data is within tolerance range (3%).
However, accuracy in flatness measurement shows uncertainties i.
e for 6 MV 7.
3%, for 15 MV 7.
31%, for 6 MeV 16.
12%, for 9 MeV 6.
92%, for 12 MeV 5.
92%, for 15 MeV 4.
01%, for 18 MeV 4.
01% and for 22 MeV 4.
13% of data are within tolerance range.

Related Results

Electrostatic Accelerators
Electrostatic Accelerators
Abstract The article contains sections titled: Introduction Types of Electrostatic Accelerators ...
SU‐GG‐T‐03: Dosimetric Study of HDR 192Ir and LDR 125I Brachytherapy Sources Using the Penelope Monte Carlo Code
SU‐GG‐T‐03: Dosimetric Study of HDR 192Ir and LDR 125I Brachytherapy Sources Using the Penelope Monte Carlo Code
Purpose: To determine the dosimetric parameters of two HDR models 192Ir (microSelectron, and M‐19), and LDR 125I brachytherapy sources. Method and Materials: In this study we used ...
Discussion on the Citation of Accelerator Standards and Testing Methods
Discussion on the Citation of Accelerator Standards and Testing Methods
Abstract As several standards are available for the accelerator test, the influence of different standards on accelerator test result was studied. Results showed tha...
Study on Modification of Alkali-Free Liquid Accelerator of Mine-Used Wet-Mix Shotcrete
Study on Modification of Alkali-Free Liquid Accelerator of Mine-Used Wet-Mix Shotcrete
Compound properties of an alkali-free liquid accelerator are the mainstream of accelerator research and development in the present. Combined with the technical characteristics of w...
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
<p>This dissertation investigates the dosimetric influence of mixed photon beams in volumetric modulated arc therapy (VMAT) for prostate cancer and presents new algorithmic f...
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
<p>This dissertation investigates the dosimetric influence of mixed photon beams in volumetric modulated arc therapy (VMAT) for prostate cancer and presents new algorithmic f...
Four‐dimensional dose evaluation using deformable image registration in radiotherapy for liver cancer
Four‐dimensional dose evaluation using deformable image registration in radiotherapy for liver cancer
Purpose:In order to evaluate the dosimetric impact of respiratory motion on the dose delivered to the target volume and critical organs during free‐breathing radiotherapy, a four‐d...
FPGA-based Feature Extraction and Tracking Accelerator for Real-Time Visual SLAM
FPGA-based Feature Extraction and Tracking Accelerator for Real-Time Visual SLAM
Due to the advantages of low latency, low power consumption and high flexibility of FPGA-based acceleration technology, it has been more and more widely studied and applied in the ...

Back to Top