SU‐E‐T‐175: A Study on Design and Fabrication of 25 * 25 cm2 Beam Monitor for Scanning Carbon Beam

View through CrossRef
Purpose: The KHIMA is developing a superconducting cyclotron of 430 MeV/u for carbon therapy. In order to verify the irradiated beam dose and position, detectors have been developed. The detector has a considered in the active area of 25*25cm2 to cover the entire beam scanning area. To minimize the loss of data during beam irradiation, a fast data acquisition system was required. Methods: A PPIC type was adopted. To measure the beam position and profile, the electrode of PPIC was designed to have a strip pattern. The PCB technique was applied to large area with thin FR‐4 plate of electrodes. The chamber has a strip interval of 1.7mm. He was used as the filling gas in order to reduce the ion collection time. Output signal from the detector connected to the two charge integrators was transferred every 10us. The signal can be stored temporarily in memory of FPGA through one of 2 integrators. Results: The fabricated large electrode plates did not show any deformation on their shape. Depending on the beam position, the amount of dose from the ion chamber will be measured. For fast data acquisition, the DAQ board was fabricated by using 2 charge integrators and the FPGA. The obtained data from the ion chamber was displaced on the PC screen every 10us. The fluence map of a single layer will be shown on the PC screen at the end of beam irradiation on the layer. The output data from the profile monitor will be compared with Gafchromic Film results. Conclusions: PCB method for active area of thin electrodes was used in the strip pattern. To reduce the loss of data during irradiation, 2 charge integrators are used alternately. The response of the readout data was set 1 0us. The beam tests will be performed with the MC‐50 Cyclotron.
Title: SU‐E‐T‐175: A Study on Design and Fabrication of 25 * 25 cm2 Beam Monitor for Scanning Carbon Beam
Description:
Purpose: The KHIMA is developing a superconducting cyclotron of 430 MeV/u for carbon therapy.
In order to verify the irradiated beam dose and position, detectors have been developed.
The detector has a considered in the active area of 25*25cm2 to cover the entire beam scanning area.
To minimize the loss of data during beam irradiation, a fast data acquisition system was required.
Methods: A PPIC type was adopted.
To measure the beam position and profile, the electrode of PPIC was designed to have a strip pattern.
The PCB technique was applied to large area with thin FR‐4 plate of electrodes.
The chamber has a strip interval of 1.
7mm.
He was used as the filling gas in order to reduce the ion collection time.
Output signal from the detector connected to the two charge integrators was transferred every 10us.
The signal can be stored temporarily in memory of FPGA through one of 2 integrators.
Results: The fabricated large electrode plates did not show any deformation on their shape.
Depending on the beam position, the amount of dose from the ion chamber will be measured.
For fast data acquisition, the DAQ board was fabricated by using 2 charge integrators and the FPGA.
The obtained data from the ion chamber was displaced on the PC screen every 10us.
The fluence map of a single layer will be shown on the PC screen at the end of beam irradiation on the layer.
The output data from the profile monitor will be compared with Gafchromic Film results.
Conclusions: PCB method for active area of thin electrodes was used in the strip pattern.
To reduce the loss of data during irradiation, 2 charge integrators are used alternately.
The response of the readout data was set 1 0us.
The beam tests will be performed with the MC‐50 Cyclotron.

Related Results

Laser-Induced Blood Coagulation for Surgical Application: A Scoping Review
Laser-Induced Blood Coagulation for Surgical Application: A Scoping Review
AbstractThere is a lack of evidence-based reviews on the effects of laser irradiation on blood coagulation in the literature, despite a large number of clinical trials. We therefor...
SU‐E‐P‐39: Characterization of Dose Impact On Different Beam Fields Size Attenuation for Elekta IGRT Couch
SU‐E‐P‐39: Characterization of Dose Impact On Different Beam Fields Size Attenuation for Elekta IGRT Couch
Purpose:Quantification of the dosimetric impact of the Elekta IGRT treatment couch in different beam field sizes. Established the relationship of relative dose versus beam angle at...
PENYELIDIKAN STRUKTUR UNTUK MENDAPATKAN DATA BAGI ANALISIS PENINGKATAN KAPASITAS GEDUNG
PENYELIDIKAN STRUKTUR UNTUK MENDAPATKAN DATA BAGI ANALISIS PENINGKATAN KAPASITAS GEDUNG
ABSTRACT The number of students of Civil Engineering Department of Bali State Polytechnic has increased dramatically since the opening of Diploma IV program in 2013. ...
SU‐E‐T‐282: Preliminary Simulation Study for 3 Dimensional Dose Delivery in Carbon Beam Active Scanning System of KHIMA
SU‐E‐T‐282: Preliminary Simulation Study for 3 Dimensional Dose Delivery in Carbon Beam Active Scanning System of KHIMA
Purpose: KHIMA (Korea Heavy Ion Medical Accelerator) project has a plan to develop and construct a research beam line for carbon beam active scanning system. The carbon beam line c...
Design
Design
Conventional definitions of design rarely capture its reach into our everyday lives. The Design Council, for example, estimates that more than 2.5 million people use design-related...
Effects of land-use changes on carbon stocks : a case study in Nam Yao Sub-Watershed, Nan Province, Thailand
Effects of land-use changes on carbon stocks : a case study in Nam Yao Sub-Watershed, Nan Province, Thailand
The study was conducted to assess carbon (C) stock potential in forest, reforestation and agricultural land-use types and reliably estimate the impact of land use on C stocks in Na...

Back to Top