Aveir Leadless Pacemaker implantation in pediatric population: a case series

View through CrossRef
Abstract Background While the Medtronic Micra pacemaker provided a small device for leadless pacemaker implantation, the Aveir device allows chronic retrievability and mapping prior to fixation. We present the first Aveir leadless pacemaker implantations in a pediatric population. Purpose To describe the first retrievable leadless pacemaker implants in children. Methods Retrospective review of pediatric Aveir implants occurred between November 2022 and December 2022. All patients were implanted via internal jugular vein insertion utilizing the Aveir sheath and catheter deployment system with mid-atrial approach taken prior to insertion of the catheter across the tricuspid valve. All patients had initial mapping of capture threshold, impedance and R-wave noted prior to deployment. Indications for pacing included sinus pauses and intermittent atrioventricular block. Results Three patients underwent Aveir leadless pacemaker implantation with weights of 32.3kg, 44kg, and 51kg. Capture thresholds were all 0.75Volts@0.2 milliseconds with impedance range of 370-1110 ohms, and R-waves of 8-14 millivolts (mV). Follow-up at 3 months ranged from 0.5V@0.2ms to 1V@0.2ms with impedances ranging 400-700 ohms, and R-wave 7-14 mV. At VVI 50-55bpm, with pacing ranged from 5% to 14%, predicted longevities ranged from 20 to 24.5 years at 3 months follow-up. Conclusion(s) Retrievable leadless pacemaker implantation is feasible in children and may offer good longevities for intermittent pacing need.Aveir Leadless Pacemaker in RVAveir Leadless Pacemaker in RV
Title: Aveir Leadless Pacemaker implantation in pediatric population: a case series
Description:
Abstract Background While the Medtronic Micra pacemaker provided a small device for leadless pacemaker implantation, the Aveir device allows chronic retrievability and mapping prior to fixation.
We present the first Aveir leadless pacemaker implantations in a pediatric population.
Purpose To describe the first retrievable leadless pacemaker implants in children.
Methods Retrospective review of pediatric Aveir implants occurred between November 2022 and December 2022.
All patients were implanted via internal jugular vein insertion utilizing the Aveir sheath and catheter deployment system with mid-atrial approach taken prior to insertion of the catheter across the tricuspid valve.
All patients had initial mapping of capture threshold, impedance and R-wave noted prior to deployment.
Indications for pacing included sinus pauses and intermittent atrioventricular block.
Results Three patients underwent Aveir leadless pacemaker implantation with weights of 32.
3kg, 44kg, and 51kg.
Capture thresholds were all 0.
75Volts@0.
2 milliseconds with impedance range of 370-1110 ohms, and R-waves of 8-14 millivolts (mV).
Follow-up at 3 months ranged from 0.
5V@0.
2ms to 1V@0.
2ms with impedances ranging 400-700 ohms, and R-wave 7-14 mV.
At VVI 50-55bpm, with pacing ranged from 5% to 14%, predicted longevities ranged from 20 to 24.
5 years at 3 months follow-up.
Conclusion(s) Retrievable leadless pacemaker implantation is feasible in children and may offer good longevities for intermittent pacing need.
Aveir Leadless Pacemaker in RVAveir Leadless Pacemaker in RV.

Related Results

Leadless Pacemaker tine fracture and dislocation: two case reports
Leadless Pacemaker tine fracture and dislocation: two case reports
Abstract The rate of leadless pacemaker implantation is increasing worldwide. To date, there have been few reports of leadless pacemaker dislocation and extraction. This ar...
P185 FOLLOW–UP OF PATIENTS UNDERGOING POST–EXTRACTION LEADLESS PACEMAKER IMPLANTATION
P185 FOLLOW–UP OF PATIENTS UNDERGOING POST–EXTRACTION LEADLESS PACEMAKER IMPLANTATION
Abstract Introduction The implantation of traditional stimulation devices involves numerous complications, in the short and long...
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct Introduction Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
The Geographic Distribution of Pediatric Anesthesiologists Relative to the US Pediatric Population
The Geographic Distribution of Pediatric Anesthesiologists Relative to the US Pediatric Population
BACKGROUND: The geographic relationship between pediatric anesthesiologists and the pediatric population has potentially important clinical and policy implications. In ...
The Pediatric Anesthesiology Workforce: Projecting Supply and Trends 2015–2035
The Pediatric Anesthesiology Workforce: Projecting Supply and Trends 2015–2035
BACKGROUND: A workforce analysis was conducted to predict whether the projected future supply of pediatric anesthesiologists is balanced with the requirements of the in...
When valve needs electrical wires - estimating pacemaker implantation after TAVR
When valve needs electrical wires - estimating pacemaker implantation after TAVR
Abstract Introduction Widespread availability and expanded indication of transcatheter aortic valve replacement (TAVR) modified ...

Back to Top