SU‐E‐T‐458: Impact of Parameters for Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer
View through CrossRef
Purpose:Different parameters Result in different plans. As the developing of volumetric modulated arc therapy for rectal cancer, the purpose of this study was to work out appropriate parameters for the optimization of rectal cancer.Methods:For ten cases which were optimized using Oncentra4.3 version, varying systematically the number of arcs, collimator angle and the delivery time to create different plans. Homogeneity in planning target volume, median dose in the organs at risk and number of monitor units were evaluated for statistical analysis. Measurements were performed with Delt4, taking the delivery time, and compared to the calculation by the gamma method.Results:Plans with collimator 45° were little superior to plans with collimator 0°. Single arc achieved lower homogeneous indices in the planning target volume compared with dual arc, but also requiring less monitor units per fraction dose than dual arc. Single arc needs less monitor units for the treatment; the measured delivery time was reduced from 4 min of dual arc to 2.5 min of single arc. All the plans could achieve good pass rate with gamma method.Conclusion:Some of the parameters have been found as the appropriate sets which can be used as standard for volumetric modulated arc therapy of rectal cancer.
Title: SU‐E‐T‐458: Impact of Parameters for Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer
Description:
Purpose:Different parameters Result in different plans.
As the developing of volumetric modulated arc therapy for rectal cancer, the purpose of this study was to work out appropriate parameters for the optimization of rectal cancer.
Methods:For ten cases which were optimized using Oncentra4.
3 version, varying systematically the number of arcs, collimator angle and the delivery time to create different plans.
Homogeneity in planning target volume, median dose in the organs at risk and number of monitor units were evaluated for statistical analysis.
Measurements were performed with Delt4, taking the delivery time, and compared to the calculation by the gamma method.
Results:Plans with collimator 45° were little superior to plans with collimator 0°.
Single arc achieved lower homogeneous indices in the planning target volume compared with dual arc, but also requiring less monitor units per fraction dose than dual arc.
Single arc needs less monitor units for the treatment; the measured delivery time was reduced from 4 min of dual arc to 2.
5 min of single arc.
All the plans could achieve good pass rate with gamma method.
Conclusion:Some of the parameters have been found as the appropriate sets which can be used as standard for volumetric modulated arc therapy of rectal cancer.
Related Results
Modified VMAT Plans for Locally Advanced Centrally Located Non-Small Cell Lung Cancer (NSCLC)
Modified VMAT Plans for Locally Advanced Centrally Located Non-Small Cell Lung Cancer (NSCLC)
Objectives: This study aimed to find the optimal radiotherapy VMAT plans, that achieved high conformity and homogeneity to the planned target volume (PTV), and minimize the dose to...
COMPARISON OF COPLANAR AND NON- COPLANARVMAT FOR BRAIN CANCER BY USING THE DOSIMETRICAL AND RADIOBIOLOGICAL INDICES
COMPARISON OF COPLANAR AND NON- COPLANARVMAT FOR BRAIN CANCER BY USING THE DOSIMETRICAL AND RADIOBIOLOGICAL INDICES
Background: Recent techniques of radiotherapy such as volumetric modulated arc therapy (VMAT) that delivered in Coplanar technique or non-coplanar technique allows to deliver high ...
Sequelae after multimodal treatment of rectal cancer
Sequelae after multimodal treatment of rectal cancer
<p dir="ltr">In recent decades, rectal cancer treatment has shifted from traditional surgical resection to include additional modalities such as radiotherapy and chemotherapy...
Sequelae after multimodal treatment of rectal cancer
Sequelae after multimodal treatment of rectal cancer
<p dir="ltr">In recent decades, rectal cancer treatment has shifted from traditional surgical resection to include additional modalities such as radiotherapy and chemotherapy...
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
<p>This dissertation investigates the dosimetric influence of mixed photon beams in volumetric modulated arc therapy (VMAT) for prostate cancer and presents new algorithmic f...
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
On Optimization of Mixed Photon Energies in Volumetric Modulated Arc Therapy
<p>This dissertation investigates the dosimetric influence of mixed photon beams in volumetric modulated arc therapy (VMAT) for prostate cancer and presents new algorithmic f...
Feasibility of 4D VMAT-CT
Feasibility of 4D VMAT-CT
Abstract
Objective. Feasibility of three-dimensional (3D) tracking of volumetric modulated arc therapy (VMAT) based on VMAT–computed tomography (V...
Dose and normal tissue complication probability analysis of various radiotherapy regimens for thymomas
Dose and normal tissue complication probability analysis of various radiotherapy regimens for thymomas
Objective
This study aimed to compare dosimetric differences and normal tissue complication probability (NTCP) among coplanar intensity-modulated radiation ther...