Comparison of surgical outcomes between sulcus and anterior chamber implanted glaucoma drainage devices

View through CrossRef
PURPOSE: This retrospective case-control cross-sectional study compared the outcomes of sulcus placement of glaucoma drainage devices (GDD) versus traditional anterior chamber (AC) to test the hypothesis that sulcus placement results in fewer complications whilst maintaining similar efficacy. METHODS: This study included 45 patients in the sulcus group and 60 patients in the anterior chamber (AC) group who had undergone surgery from January 2014 to December 2017. Data were collected on pre-operative demographics, operative details and post-operative intraocular pressure and complications. The IOP, number of medications and complications between the two groups was compared. A P value of <5% was considered statistically significant. RESULTS: The sulcus group had significantly lower overall complications compared to the AC group with a comparable IOP decrease between groups. There were significantly lower rates of hyphaema in the sulcus (3 cases) compared to AC group (17 cases) ( P < 0.05). Severe or late complications (implant exposure, corneal decompensation, endophthalmitis, poor vision, choroidal hemorrhage and cornea edema) were significantly lower in the sulcus group [2 eyes; 4.4%] compared to the AC group [13 eyes; 21.7%] ( P < 0.05). The sulcus group required fewer medications during the follow-up period. CONCLUSION: Sulcus implantation of GDD resulted in less postoperative hyphaema and severe complications compared to AC implantation. Our findings concur with the literature that sulcus implantation is safe and effective for controlling IOP for various types of glaucoma. The long-term effects of endothelial cell loss for sulcus versus AC implantation require further evaluation.
Title: Comparison of surgical outcomes between sulcus and anterior chamber implanted glaucoma drainage devices
Description:
PURPOSE: This retrospective case-control cross-sectional study compared the outcomes of sulcus placement of glaucoma drainage devices (GDD) versus traditional anterior chamber (AC) to test the hypothesis that sulcus placement results in fewer complications whilst maintaining similar efficacy.
METHODS: This study included 45 patients in the sulcus group and 60 patients in the anterior chamber (AC) group who had undergone surgery from January 2014 to December 2017.
Data were collected on pre-operative demographics, operative details and post-operative intraocular pressure and complications.
The IOP, number of medications and complications between the two groups was compared.
A P value of <5% was considered statistically significant.
RESULTS: The sulcus group had significantly lower overall complications compared to the AC group with a comparable IOP decrease between groups.
There were significantly lower rates of hyphaema in the sulcus (3 cases) compared to AC group (17 cases) ( P < 0.
05).
Severe or late complications (implant exposure, corneal decompensation, endophthalmitis, poor vision, choroidal hemorrhage and cornea edema) were significantly lower in the sulcus group [2 eyes; 4.
4%] compared to the AC group [13 eyes; 21.
7%] ( P < 0.
05).
The sulcus group required fewer medications during the follow-up period.
CONCLUSION: Sulcus implantation of GDD resulted in less postoperative hyphaema and severe complications compared to AC implantation.
Our findings concur with the literature that sulcus implantation is safe and effective for controlling IOP for various types of glaucoma.
The long-term effects of endothelial cell loss for sulcus versus AC implantation require further evaluation.

Related Results

Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct Introduction Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
COLLATERAL SULCUS MORPHOLOGY IN NORMAL INDIVIDUALS IN MAGNETIC RESONANCE IMAGING OF BRAIN
COLLATERAL SULCUS MORPHOLOGY IN NORMAL INDIVIDUALS IN MAGNETIC RESONANCE IMAGING OF BRAIN
Collateral sulcus (CS) develops during early stage of the gestation and is well formed at 23rd weeks during pregnancy. The temporal part of the CS continues with limbic ssure and ...
Childhood glaucoma profile in a Southwestern Ethiopia tertiary care center: a retrospective study
Childhood glaucoma profile in a Southwestern Ethiopia tertiary care center: a retrospective study
Abstract Background Childhood glaucoma is a major cause of childhood blindness worldwide. The profile of childhood glaucoma has not been well characterized in sub-Saharan ...
Correlation of occlusal characteristics with mentolabial sulcus in patients at a tertiary care hospital: A cross-sectional study
Correlation of occlusal characteristics with mentolabial sulcus in patients at a tertiary care hospital: A cross-sectional study
Objective: To find correlation of mentolabial sulcus depth and mentolabial sulcus angle with skeletal and dental characteristics. Method: The observational, cross-sectional study w...
Genetics of Glaucoma
Genetics of Glaucoma
Abstract Glaucoma is the major cause of blindness worldwide. Primary open‐angle glaucoma (POAG) and angle closure glaucoma (ACG) are the most pr...
Trend of glaucoma internal filtration surgeries in a tertiary hospital in China
Trend of glaucoma internal filtration surgeries in a tertiary hospital in China
AIM: To evaluate the trend of glaucoma internal filtration surgeries for inpatients between 2015 and 2021 at the Eye Hospital of Wenzhou Medical University. METHODS: A review of th...