Comparison of ultrafast half‐Fourier single‐shot turbo spin‐echo sequence with turbo spin‐echo sequences for T2‐weighted imaging of the female pelvis
View through CrossRef
AbstractSo that we might evaluate the ultrafast half‐Fourier single‐shot turbo spin‐echo (HASTE) sequence in T2‐weighted MRI of the female pelvis and compare it with the turbo spin‐echo (TSE) sequence, we prospectively studied 60 consecutive females with suspected abnormalities of the pelvis. For all MR examinations, we used a 1.5‐T superconductive magnet with a phased array coil. The HASTE sequence was applied with TR/effective TE/echo train = ∞/90/64 and a 128 × 256 matri× (acquisition time: .3 sec/slice), conventional TSE imaging with 3,400 to 5,000/132/15 and a 128 × 256 matri× (mean acquisition time: 2 min 4 sec), and high‐resolution TSE imaging with 3,400 to 5,000/132/15 and a 300 × 512 matri× (6 min 4 sec). Although the lesion conspicuity for the HASTE sequence was less than that for the high‐resolution TSE sequences, artifacts (including ghosting, bowel motion, susceptibility difference, and chemical shift) were negligible on HASTE images of all patients. The lesion conspicuity for the HASTE sequence was significantly better than for the conventional TSE sequence. In spite of the very short acquisition time, the subjective scoring of the overall image quality for the HASTE sequence was significantly higher than for the conventional TSE sequence (P < .01) and were slightly lower than for the high‐resolution TSE sequence. Compared with high‐resolution TSE, HASTE provided clearer visualization of large leiomyomas and ovarian tumors but slightly poorer visualization of uterine cancer. In occlusion, HASTE sequence generates higher contrast and is free from motion and chemical shift artifact with much higher time efficacy. Because of limited image resolution, the HASTE sequence should be used when the high‐resolution TSE imaging is suboptimal.
Title: Comparison of ultrafast half‐Fourier single‐shot turbo spin‐echo sequence with turbo spin‐echo sequences for T2‐weighted imaging of the female pelvis
Description:
AbstractSo that we might evaluate the ultrafast half‐Fourier single‐shot turbo spin‐echo (HASTE) sequence in T2‐weighted MRI of the female pelvis and compare it with the turbo spin‐echo (TSE) sequence, we prospectively studied 60 consecutive females with suspected abnormalities of the pelvis.
For all MR examinations, we used a 1.
5‐T superconductive magnet with a phased array coil.
The HASTE sequence was applied with TR/effective TE/echo train = ∞/90/64 and a 128 × 256 matri× (acquisition time: .
3 sec/slice), conventional TSE imaging with 3,400 to 5,000/132/15 and a 128 × 256 matri× (mean acquisition time: 2 min 4 sec), and high‐resolution TSE imaging with 3,400 to 5,000/132/15 and a 300 × 512 matri× (6 min 4 sec).
Although the lesion conspicuity for the HASTE sequence was less than that for the high‐resolution TSE sequences, artifacts (including ghosting, bowel motion, susceptibility difference, and chemical shift) were negligible on HASTE images of all patients.
The lesion conspicuity for the HASTE sequence was significantly better than for the conventional TSE sequence.
In spite of the very short acquisition time, the subjective scoring of the overall image quality for the HASTE sequence was significantly higher than for the conventional TSE sequence (P < .
01) and were slightly lower than for the high‐resolution TSE sequence.
Compared with high‐resolution TSE, HASTE provided clearer visualization of large leiomyomas and ovarian tumors but slightly poorer visualization of uterine cancer.
In occlusion, HASTE sequence generates higher contrast and is free from motion and chemical shift artifact with much higher time efficacy.
Because of limited image resolution, the HASTE sequence should be used when the high‐resolution TSE imaging is suboptimal.
Related Results
Analisis Penggunaan Variasi Turbo Cyclone Terhadap Performa Kendaraan
Analisis Penggunaan Variasi Turbo Cyclone Terhadap Performa Kendaraan
Penelitian ini dilatar belakangi banyaknya kendaraan dengan usia pakai dan pola perawatan yang tidak rutin yang berakibat turunannya performa dan emisi yang meningkat. Penelitian i...
Dynamics of spinor fermions
Dynamics of spinor fermions
Ultracold atomic gases have established themselves as quantum systems, which are clean and offer a high degree of control over crucial parameters. They are well isolated from their...
Introduction
Introduction
Jean Baptiste Joseph Fourier’s powerful idea of decomposition of a signal into sinusoidal components has found application in almost every engineering and science field. An incompl...
Tailoring spin dynamics in asymmetric FM1/Pt/FM2 trilayers via Pt spacer thickness
Tailoring spin dynamics in asymmetric FM1/Pt/FM2 trilayers via Pt spacer thickness
The study of trilayers with a non-magnetic (NM) spacer layer separating two ferromagnetic layers (FM/NM/FM) has been an active area of spintronics research due to their real-world ...
Spin to charge current interconversion in Rasha interfaces and topological insulators
Spin to charge current interconversion in Rasha interfaces and topological insulators
Conversion entre courant de spin et courant de charge dans des interfaces Rashba et des isolants topologiques
L'interconversion entre courants de spin et de charge ...
Peranan Sekuen T1 SPACE Coronal Pada Pemeriksaan MRI Brain Dengan Klinis Epilepsi Lobus Temporal Di Unit Radiologi RSUP Prof. Dr. I.G.N.G. Ngoerah Denpasar
Peranan Sekuen T1 SPACE Coronal Pada Pemeriksaan MRI Brain Dengan Klinis Epilepsi Lobus Temporal Di Unit Radiologi RSUP Prof. Dr. I.G.N.G. Ngoerah Denpasar
Background: Magnetic resonance imaging (MRI) is a cross-sectional imaging technique based on the principle of magnetic resonance of hydrogen atoms due to radio frequency (RF) signa...
Ultrafast Brain MRI: Clinical Deployment and Comparison to Conventional Brain MRI at 3T
Ultrafast Brain MRI: Clinical Deployment and Comparison to Conventional Brain MRI at 3T
ABSTRACTBACKGROUND AND PURPOSETo compare an ultrafast brain magnetic resonance imaging (MRI) protocol to the conventional protocol in motion‐prone inpatient clinical settings.METHO...
Abstract P4-01-06: Evaluation of 3D T2-weighted Breast MRI
Abstract P4-01-06: Evaluation of 3D T2-weighted Breast MRI
Abstract
Background: Although the dynamic contrast enhanced (DCE) sequence has long been considered the most important sequence to characterize benign and malignant ...