Short term biatrial remodeling after pulsed field ablation vs cryoablation: a comparative cardiac magnetic resonance volumetric and strain analysis
View through CrossRef
Abstract
Background
Pulsed field ablation (PFA) is emerging as a non-thermal modality for Pulmonary vein isolation (PVI) in the treatment of atrial fibrillation (AF) demonstrating favorable clinical outcomes when compared to traditional radiofrequency and cryoablation(1). While PVI reduces AF burden and may promote reversal of atrial cardiomyopathy, ablation lesions can adversely affect left atrial compliance and function. (2) (3) The impact of PFA on atrial structural and functional remodeling, particularly in relation to cryoablation, remains underexplored through advanced imaging modalities.
Purpose
To compare the impact of pulsed field ablation to cryoablation for AF on right and left atrial volumes and functions by cardiac magnetic resonance imaging (CMR).
Methods
This comparative cohort study included 30 patients with AF who underwent PVI using either PFA ( n=13) or cryoablation (n=17) at a single center. Patients with prior atrial ablation or significant structural heart disease, apart from AF induced cardiomyopathy, were excluded. Both procedures were done under fluoroscopy guidance with contrast injection to delineate pulmonary veins. PFA was done using pentaspline catheter using standard protocols(4). All patients underwent CMR before and 3–6 months after ablation to quantify biatrial volumes and strain parameters (reservoir, conduit and contractile). Comparisons between pre and post ablation parameters collectively and between both groups were made using appropriate tests. The primary analysis of post-ablation CMR parameters was performed using analysis of covariance (ANCOVA), adjusted for baseline values, atrial fibrillation type, and time to follow-up imaging.
Results
Baseline demographic and clinical characteristics were comparable between groups, except for a higher proportion of persistent AF in the PFA cohort and a shorter interval to follow-up CMR. PFA patients had significantly larger ventricular volumes at initial studies (Figure 1).
At follow-up, both groups exhibited significant reduction in left atrial expansion index, left atrial contractile ejection fraction and left atrial strain parameters. Meanwhile right atrial contractile ejection fraction and contractile strain showed marked improvement. (Figure 2, Panel A )
The analysis of changes of atrial parameters between the two groups demonstrated no statistically significant difference in the adjusted and unadjusted changes of atrial volumetric or functional parameters. (Figure 2 Panel B)
Conclusion
In our CMR based study, both PFA and cryoablation achieved comparable short-term atrial remodeling marked by enhanced right atrial contractility and declined LA functional parameters. Although PFA group had more advanced AF pattern and higher ventricular volumes, no significant differences were detected in right or left atrial parameters between the two modalities. Larger, randomized studies are warranted to confirm these findings.
Oxford University Press (OUP)
Title: Short term biatrial remodeling after pulsed field ablation vs cryoablation: a comparative cardiac magnetic resonance volumetric and strain analysis
Description:
Abstract
Background
Pulsed field ablation (PFA) is emerging as a non-thermal modality for Pulmonary vein isolation (PVI) in the treatment of atrial fibrillation (AF) demonstrating favorable clinical outcomes when compared to traditional radiofrequency and cryoablation(1).
While PVI reduces AF burden and may promote reversal of atrial cardiomyopathy, ablation lesions can adversely affect left atrial compliance and function.
(2) (3) The impact of PFA on atrial structural and functional remodeling, particularly in relation to cryoablation, remains underexplored through advanced imaging modalities.
Purpose
To compare the impact of pulsed field ablation to cryoablation for AF on right and left atrial volumes and functions by cardiac magnetic resonance imaging (CMR).
Methods
This comparative cohort study included 30 patients with AF who underwent PVI using either PFA ( n=13) or cryoablation (n=17) at a single center.
Patients with prior atrial ablation or significant structural heart disease, apart from AF induced cardiomyopathy, were excluded.
Both procedures were done under fluoroscopy guidance with contrast injection to delineate pulmonary veins.
PFA was done using pentaspline catheter using standard protocols(4).
All patients underwent CMR before and 3–6 months after ablation to quantify biatrial volumes and strain parameters (reservoir, conduit and contractile).
Comparisons between pre and post ablation parameters collectively and between both groups were made using appropriate tests.
The primary analysis of post-ablation CMR parameters was performed using analysis of covariance (ANCOVA), adjusted for baseline values, atrial fibrillation type, and time to follow-up imaging.
Results
Baseline demographic and clinical characteristics were comparable between groups, except for a higher proportion of persistent AF in the PFA cohort and a shorter interval to follow-up CMR.
PFA patients had significantly larger ventricular volumes at initial studies (Figure 1).
At follow-up, both groups exhibited significant reduction in left atrial expansion index, left atrial contractile ejection fraction and left atrial strain parameters.
Meanwhile right atrial contractile ejection fraction and contractile strain showed marked improvement.
(Figure 2, Panel A )
The analysis of changes of atrial parameters between the two groups demonstrated no statistically significant difference in the adjusted and unadjusted changes of atrial volumetric or functional parameters.
(Figure 2 Panel B)
Conclusion
In our CMR based study, both PFA and cryoablation achieved comparable short-term atrial remodeling marked by enhanced right atrial contractility and declined LA functional parameters.
Although PFA group had more advanced AF pattern and higher ventricular volumes, no significant differences were detected in right or left atrial parameters between the two modalities.
Larger, randomized studies are warranted to confirm these findings.
Related Results
Data from Cryoablation Does Not Significantly Contribute to Systemic Effector Immune Responses in a Poorly Immunogenic B16F10 Melanoma Model
Data from Cryoablation Does Not Significantly Contribute to Systemic Effector Immune Responses in a Poorly Immunogenic B16F10 Melanoma Model
<div>AbstractPurpose:<p>Cryoablation is a minimally invasive procedure implemented to destroy solid tumors. It also results in the release of tumor antigens into the sy...
Data from Cryoablation Does Not Significantly Contribute to Systemic Effector Immune Responses in a Poorly Immunogenic B16F10 Melanoma Model
Data from Cryoablation Does Not Significantly Contribute to Systemic Effector Immune Responses in a Poorly Immunogenic B16F10 Melanoma Model
<div>AbstractPurpose:<p>Cryoablation is a minimally invasive procedure implemented to destroy solid tumors. It also results in the release of tumor antigens into the sy...
Primerjalna književnost na prelomu tisočletja
Primerjalna književnost na prelomu tisočletja
In a comprehensive and at times critical manner, this volume seeks to shed light on the development of events in Western (i.e., European and North American) comparative literature ...
Microwave Ablation with or Without Chemotherapy in Management of Non-Small Cell Lung Cancer: A Systematic Review
Microwave Ablation with or Without Chemotherapy in Management of Non-Small Cell Lung Cancer: A Systematic Review
Abstract
Introduction
Microwave ablation (MWA) has emerged as a minimally invasive treatment for patients with inoperable non-small cell lung cancer (NSCLC). However, whether it i...
Cardiac Myxoma Post-Transseptal Ablation: Coincidence or Causation?
Cardiac Myxoma Post-Transseptal Ablation: Coincidence or Causation?
Background: Cardiac myxomas are benign cardiac neoplasms usually found solitarily located within a single cardiac chamber, most commonly in the left atrium. With no established cau...
Cryoablation Technology:Mechanism, Medical DeviceSystem, Applications, and Future Directions.
Cryoablation Technology:Mechanism, Medical DeviceSystem, Applications, and Future Directions.
Cryoablation is a minimally invasive procedure that has emerged as an attractive alternative for
treating various diseases by leveraging subzero temperatures rather than heat. Unli...
Analysis of magnetohydrodynamic drag character for hypersonic vehicles
Analysis of magnetohydrodynamic drag character for hypersonic vehicles
In hypersonic flight, a very high temperature area can form ahead of the nose of aerocraft due to the shock aerodynamic heating, which leads to air weakly ionized. Many researchers...
Cryoablation and radiofrequency ablation during mitral valve surgery for rheumatic mitral valve disease: a retrospective cohort study
Cryoablation and radiofrequency ablation during mitral valve surgery for rheumatic mitral valve disease: a retrospective cohort study
Background and aims
Atrial fibrillation (AF) frequently accompanies rheumatic mitral valve disease (MVD) and adversely affects postoperative outcomes. Radiofreq...