Crystal-level timing calibration using cascaded photons of 60Co point source for long axial animal PET system
View through CrossRef
Abstract
Objective. Timing calibration is essential for positron emission tomography (PET) system as it enhances timing resolution to improve image quality. Traditionally, positron sources are employed for timing calibration. However, the photons emitted by these sources travel in opposite directions, necessitating that positrons annihilate at multiple locations to collect coincidence data across a greater number of lines of response (LORs). To overcome this limitation, this study proposes a timing calibration method utilising a 60Co point source. Approach. The 60Co source emits cascaded photons without angular correlation, allowing the collection of coincidence events throughout the field of view (FOV) with a single 60Co point source positioned at the centre of the FOV to determine the timing offsets of the pixels. Leveraging the properties of 60Co, we propose a calibration method and implement it on a long axial animal PET system. Initially, we calibrated the timing offsets of the pixels within two blocks to establish reference detectors, and subsequently employed a 60Co point source to determine the timing offsets of all the pixels in the system relative to these reference detectors. In addition, we evaluated the system's timing resolution before and after the calibration to validate the efficacy of the proposed method. Main results. We measured the timing offsets of the pixels across the entire system, ranging from -5.0 to 2.0 ns. After implementing the timing offset lookup table, the system timing resolution was improved from 6.30 ns before calibration to 1.04 ns. Significance. In this study, the 60Co source is employed for timing calibration, offering the advantages of operational simplicity, broad applicability, and potential application in time-of-flight (TOF) PET.
Title: Crystal-level timing calibration using cascaded photons of 60Co point source for long axial animal PET system
Description:
Abstract
Objective.
Timing calibration is essential for positron emission tomography (PET) system as it enhances timing resolution to improve image quality.
Traditionally, positron sources are employed for timing calibration.
However, the photons emitted by these sources travel in opposite directions, necessitating that positrons annihilate at multiple locations to collect coincidence data across a greater number of lines of response (LORs).
To overcome this limitation, this study proposes a timing calibration method utilising a 60Co point source.
Approach.
The 60Co source emits cascaded photons without angular correlation, allowing the collection of coincidence events throughout the field of view (FOV) with a single 60Co point source positioned at the centre of the FOV to determine the timing offsets of the pixels.
Leveraging the properties of 60Co, we propose a calibration method and implement it on a long axial animal PET system.
Initially, we calibrated the timing offsets of the pixels within two blocks to establish reference detectors, and subsequently employed a 60Co point source to determine the timing offsets of all the pixels in the system relative to these reference detectors.
In addition, we evaluated the system's timing resolution before and after the calibration to validate the efficacy of the proposed method.
Main results.
We measured the timing offsets of the pixels across the entire system, ranging from -5.
0 to 2.
0 ns.
After implementing the timing offset lookup table, the system timing resolution was improved from 6.
30 ns before calibration to 1.
04 ns.
Significance.
In this study, the 60Co source is employed for timing calibration, offering the advantages of operational simplicity, broad applicability, and potential application in time-of-flight (TOF) PET.
Related Results
SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
A enfermagem é o pilar da assistência pois está na linha de frente do cuidado holístico, todavia esta é estigmatizada e desvalorizada, assim como não possui reconhecimento consider...
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
Introduction
Heterogeneous gas sensor arrays coupled with machine learning algorithms have been proposed for a wide range of applications. However, i...
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
ObjectiveTo compare the diagnostic differences between total-body PET/CT (positron emission tomography/computed tomography) and PET/MR (positron emission tomography/magnetic resona...
Electric Animal An Interview with Akira Mizuta Lippit & (untitled photographs)
Electric Animal An Interview with Akira Mizuta Lippit & (untitled photographs)
Dr. Akira Mizuta Lippit, author of Electric Animal: Toward a Rhetoric of Wildlife, explores, in the context of the development of cinema, how the concept of “the animal” has become...
Clinical and 18F-FDG PET/CT Imaging Characteristics of Post-radiotherapy Sacral Insufficiency Fractures in Cervical Cancer Patients
Clinical and 18F-FDG PET/CT Imaging Characteristics of Post-radiotherapy Sacral Insufficiency Fractures in Cervical Cancer Patients
Introduction:
Sacral Insufficiency Fractures (SIFs) are a common yet frequently misdiagnosed late complication following pelvic radiotherapy for cervical
cancer...
Baseline Staging Evaluation in Lymphoma: The Role of FDG PET, CT, and Bone Marrow Biopsy
Baseline Staging Evaluation in Lymphoma: The Role of FDG PET, CT, and Bone Marrow Biopsy
Abstract
Abstract 2640
BACKGROUND:
The revised response criteria for malignant lymphoma (Cheson et al JCO 25:579 ...
Multi-photon quantum interference and entanglement
Multi-photon quantum interference and entanglement
Interférence et intrication quantiques à plusieurs photons
Dans cette thèse de doctorat, nous étudions des caractéristiques clés du traitement optique de l'informat...
PET -CT Adapted Therapy After 3 Cycles of ABVD for All Stages of Hodgkin Lymphoma. Preliminary Results in 193 Patients
PET -CT Adapted Therapy After 3 Cycles of ABVD for All Stages of Hodgkin Lymphoma. Preliminary Results in 193 Patients
Abstract
Abstract 1619
Background:
Positron emission tomography using 18F-fluoro-2-deoxy-d-glucose (FDG-PET-CT) i...