Search engine for discovering works of Art, research articles, and books related to Art and Culture
ShareThis
Javascript must be enabled to continue!

Are new implants for patellar fractures reasonable? A biomechanical comparison of intramedullary locking nails versus tension-band osteosynthesis

View through CrossRef
Abstract Background Patellar fractures are predominantly treated surgically by open reduction and internal fixation using classic tension band osteosynthesis. Biomechanical testing was conducted to examine transverse patellar fracture stabilization using locked intramedullary nail systems in comparison to classic tension band osteosynthesis. Methods Twenty-four Sawbones ® with transverse patellar fractures type AO 34-C1.1 were divided equally into three test groups. Three principles for fracture stabilization were used: classical tension band osteosynthesis and newly developed locked intramedullary double nail- and single nail- prototypes. Knee motion (0°/0°/90°) was simulated in a dynamic test set-up using a “single muscle model” by applying tensile forces up to 300 N. The widening and the symmetry of the fracture gap were investigated over 1000 motion cycles using a video-optical system. The significance level was set at α = 0.05 for all statistical evaluations using the Bonferroni–Holm corrected unpaired unilateral t-test and mixed ANOVA with post-hoc Tukey test. Results The fixation principles showed different maximum fragment displacements of M = 2.04 ± 0.67 mm using classical tension band osteosynthesis, M = 1.34 ± 0.74 mm using intramedullary single nail system and M = 0.55 ± 0.31 mm using intramedullary double nail system. There was a statistically significant change in width of the fracture gap in flexed knee position comparing tension band osteosynthesis and double nail system ( p  = 0.0015) and comparing single nail and double nail system ( p  = 0.02). Furthermore, the tension-band osteosynthesis showed a significantly greater divergence from symmetry in the fracture gap compared to both intramedullary treatment methods ( p  < 0.001). Conclusions Prototypes of locked intramedullary nail systems showed comparable fixation results for transverse patellar fractures compared to tension band wiring. They showed more symmetrical fracture gap openings during biomechanical testing and offered significantly higher stability to the fragments.
Title: Are new implants for patellar fractures reasonable? A biomechanical comparison of intramedullary locking nails versus tension-band osteosynthesis
Description:
Abstract Background Patellar fractures are predominantly treated surgically by open reduction and internal fixation using classic tension band osteosynthesis.
Biomechanical testing was conducted to examine transverse patellar fracture stabilization using locked intramedullary nail systems in comparison to classic tension band osteosynthesis.
Methods Twenty-four Sawbones ® with transverse patellar fractures type AO 34-C1.
1 were divided equally into three test groups.
Three principles for fracture stabilization were used: classical tension band osteosynthesis and newly developed locked intramedullary double nail- and single nail- prototypes.
Knee motion (0°/0°/90°) was simulated in a dynamic test set-up using a “single muscle model” by applying tensile forces up to 300 N.
The widening and the symmetry of the fracture gap were investigated over 1000 motion cycles using a video-optical system.
The significance level was set at α = 0.
05 for all statistical evaluations using the Bonferroni–Holm corrected unpaired unilateral t-test and mixed ANOVA with post-hoc Tukey test.
Results The fixation principles showed different maximum fragment displacements of M = 2.
04 ± 0.
67 mm using classical tension band osteosynthesis, M = 1.
34 ± 0.
74 mm using intramedullary single nail system and M = 0.
55 ± 0.
31 mm using intramedullary double nail system.
There was a statistically significant change in width of the fracture gap in flexed knee position comparing tension band osteosynthesis and double nail system ( p  = 0.
0015) and comparing single nail and double nail system ( p  = 0.
02).
Furthermore, the tension-band osteosynthesis showed a significantly greater divergence from symmetry in the fracture gap compared to both intramedullary treatment methods ( p  < 0.
001).
Conclusions Prototypes of locked intramedullary nail systems showed comparable fixation results for transverse patellar fractures compared to tension band wiring.
They showed more symmetrical fracture gap openings during biomechanical testing and offered significantly higher stability to the fragments.

Related Results

ANALYSIS OF SURGICAL TREATMENT RESULTS IN PATIENTS WITH FRACTURES OF THE TIBIAL BONES
ANALYSIS OF SURGICAL TREATMENT RESULTS IN PATIENTS WITH FRACTURES OF THE TIBIAL BONES
Relevance: By frequency, tibial bone fractures rank second, accounting for 13% to 21.4% of all musculoskeletal injuries or 64.3%-70% of lower limb bone fractures. The nature of the...
Spontaneous Bilateral Patellar Tendon Rupture in Patient with Ehlers–Danlos Syndrome: A Case Report
Spontaneous Bilateral Patellar Tendon Rupture in Patient with Ehlers–Danlos Syndrome: A Case Report
Introduction: Bilateral spontaneous patellar tendon ruptures are rare, though it is hypothesized to be more common in patients with risk factors such as connective tissue disorders...
Evaluation of Patellar Dimension and Bristol Index in Asian Population: An MRI Study
Evaluation of Patellar Dimension and Bristol Index in Asian Population: An MRI Study
Abstract Introduction: Restoration of patellar thickness is pivotal during a total knee arthroplasty (TKA). This study aims to evaluate the patellar dimension and Bristol I...
Pathogenic Organisms Associated with Artificial Fingernails Worn by Healthcare Workers
Pathogenic Organisms Associated with Artificial Fingernails Worn by Healthcare Workers
Objective:To determine differences in the identity and quantity of microbial flora from healthcare workers (HCWs) wearing artificial nails compared with control HCWs with native na...
Intramedullary Nail Versus Dynamic Hip Screw; Intramedullary Nail (Advantages And Disadvantages)
Intramedullary Nail Versus Dynamic Hip Screw; Intramedullary Nail (Advantages And Disadvantages)
since 1950s surgical treatment of extracapsular hip fractures is done using a variety of different implants. Unstable fractures are fractures with lateral wall or posteromedial c...
Effect of angular stability and other locking parameters on the mechanical performance of intramedullary nails
Effect of angular stability and other locking parameters on the mechanical performance of intramedullary nails
AbstractTo extend the indications of intramedullary nails for distal or proximal fractures, nails with angle stable locking options have been developed. Studies on the mechanical e...
A Preliminary Study of Patellar Tendon Torques during Jumping
A Preliminary Study of Patellar Tendon Torques during Jumping
The etiology of patellar tendinopathy (jumper’s knee) has been attributed to a significant increase in patellar tendon torques associated with jumping. While some investigators hav...
Patellofemoral resurfacing and patellar denervation in primary total knee arthroplasty
Patellofemoral resurfacing and patellar denervation in primary total knee arthroplasty
AbstractPurposeTo conduct a systematic review and network meta‐analysis of randomized controlled trials (RCTs) with the aim of comparing relevant clinical outcomes between patellar...

Back to Top