In vitro assessment of the accuracy of three electronic apex locators
View through CrossRef
Aim of this in vitro study was to compare accuracy of three electronic apex locators (EALs) (ES-02, ES-03 and ProPex) in root canals of extracted human teeth utilizing two meter readings and diff erent conditions of root canals (RC). RC lengths of 46 single-rooted teeth were visually determined using a K-fi le # 15. Teeth and labial connector were then embedded in alginate. Two sets of measurements were performed for each device - root canal condition combination (dry root canal, 2.5% NaOCl, 17% EDTA, RC prep, Glyde) according to readings “-0.5” and “0.0”. The diff erences were evaluated using descriptive and nonparametric statistics. The mean diff erence between real and measured values at meter reading “0.0” was 0.10 mm with ProPex, 0.31 mm with ES-02 and 0.71 mm with ES-03. At meter reading “-0.5“ that diff erence was 0.28 mm with ProPex, 0.41 mm with ES-02 and 0.87 mm with ES-03. In vitro testing showed ProPex to be the most accurate EAL in most of combinations, followed by ES-02, and ES-03 (p<0.05). Most accurate measurements were performed in canal conditions as follows: dry canals, Glyde, Rc-Prep, NaOCl and EDTA solutions (p<0.05). ProPex was the most accurate at meter reading “0.0”. At reading “-0.5” ES-02 and ProPex were equal, whereas ES-03 was the least accurate. All tested EALs should be used with meter reading “0.0” as a reference because it is more accurate than using the meter reading “-0.5” (p<0.001).
Croatian Medical Association
Title: In vitro assessment of the accuracy of three electronic apex locators
Description:
Aim of this in vitro study was to compare accuracy of three electronic apex locators (EALs) (ES-02, ES-03 and ProPex) in root canals of extracted human teeth utilizing two meter readings and diff erent conditions of root canals (RC).
RC lengths of 46 single-rooted teeth were visually determined using a K-fi le # 15.
Teeth and labial connector were then embedded in alginate.
Two sets of measurements were performed for each device - root canal condition combination (dry root canal, 2.
5% NaOCl, 17% EDTA, RC prep, Glyde) according to readings “-0.
5” and “0.
0”.
The diff erences were evaluated using descriptive and nonparametric statistics.
The mean diff erence between real and measured values at meter reading “0.
0” was 0.
10 mm with ProPex, 0.
31 mm with ES-02 and 0.
71 mm with ES-03.
At meter reading “-0.
5“ that diff erence was 0.
28 mm with ProPex, 0.
41 mm with ES-02 and 0.
87 mm with ES-03.
In vitro testing showed ProPex to be the most accurate EAL in most of combinations, followed by ES-02, and ES-03 (p<0.
05).
Most accurate measurements were performed in canal conditions as follows: dry canals, Glyde, Rc-Prep, NaOCl and EDTA solutions (p<0.
05).
ProPex was the most accurate at meter reading “0.
0”.
At reading “-0.
5” ES-02 and ProPex were equal, whereas ES-03 was the least accurate.
All tested EALs should be used with meter reading “0.
0” as a reference because it is more accurate than using the meter reading “-0.
5” (p<0.
001).
Related Results
Accuracy evaluation of three electronic apex locators in teeth with immature apices
Accuracy evaluation of three electronic apex locators in teeth with immature apices
The aim of this randomized controlled study was to evaluate the accuracy of Root ZX, iPex and YS-RZ-A apex locators in tooth with incomplete apices. Material and methods: Thirty hu...
Electronic working length determination in primary teeth by Propex and Digital Signal Processing
Electronic working length determination in primary teeth by Propex and Digital Signal Processing
AbstractThe purpose of this study was to evaluate the accuracy of electronic apex locators Digital Signal Processing (DSP) and ProPex, for root canal length determination in primar...
Use of electronic apex locator in the impossibility of radiographic visualization of the root apex: report of two cases
Use of electronic apex locator in the impossibility of radiographic visualization of the root apex: report of two cases
Electronic apex locators have been widely used for foraminal identification and for working length determination. The establishment of a correct apical limit is essential for a goo...
Working length determination
Working length determination
Introduction: Working length (WL) determination is one of the crucial steps in endodontic treatment. Conventional Apex locators (EAL) are consistently employed nowadays for WL dete...
Accuracy of two electronic apex locators: in vivo and in vitro study
Accuracy of two electronic apex locators: in vivo and in vitro study
The present study proposed the comparison of the reading between the in vivo and in vitro models during the evaluation of Root ZX II foramen locators and Propex II. The localizatio...
In vitro assessment of the accuracy of three electronic apex locators
In vitro assessment of the accuracy of three electronic apex locators
Aim of this in vitro study was to compare accuracy of three electronic apex locators (EALs) (ES-02, ES-03 and ProPex) in root canals of extracted human teeth utilizing two meter re...
A systematic review of the surgical anatomy of the orbital apex
A systematic review of the surgical anatomy of the orbital apex
Abstract
Purpose
The orbital apex is the narrowest part of the orbit, housing the link between the intracranial cavity and orbit. Knowledge of orbit...
Anatomical Investigation of the Cardiac Apex
Anatomical Investigation of the Cardiac Apex
Background: Understanding cardiac anatomy is the key to solve unknown issues about its function. The continuous and helical myocardial structure plays a fundamental role in its tor...