pO2 and pCO2 Increment in Post‐dialyzer Blood: The Role of Dialysate
View through CrossRef
Abstract: Blood returning from a dialyzer during hemodialysis has a higher pO2 and pCO2 content than blood entering the dialyzer, and this has been attributed to the dialysate. The present study investigates this phenomenon. Acid–base and blood‐gas parameters (pH, pO2, pCO2 and HCO3) were measured in three groups of stable chronic hemodialysis patients (A, B, and C) undergoing high‐flux hemodialysis. In group A (n = 15), “arterial” (a) and “venous” (v) samples were withdrawn simultaneously before dialysis (samples A0), 5 min after circulation of the blood with the dialysate in the by‐pass mode (samples A5), and 5 min after high‐flux hemodialysis at a zero ultrafiltration rate (samples A10). In group B (n = 11) (a) and (v) samples were withdrawn simultaneously before dialysis (samples B0), 5 min after isolated‐ultrafiltration with closed dialysate ports (“isolated‐closed” ultrafiltration) (samples B5), and 5 min after high‐flux hemodialysis at a zero ultrafiltration rate (samples B10). In group C (n = 14), after an initial arterial blood sample withdrawal before hemodialysis (sample C0), high‐flux hemodialysis at a zero ultrafiltration rate was initiated. Five minutes later, blood and dialysate samples were withdrawn simultaneously from the hemodialysis lines (samples C5). In all cases blood and dialysate (bicarbonate) flow rates were set at 0.300 and 0.700 L/min, respectively. FLX‐18 hemodialyzers (membrane PEPA 1.8 m2) were used in this study. Analysis of variance revealed significant changes only in venous samples. A comparison of arterial and venous samples revealed no differences between groups A and B before the initiation of dialysis (A0a vs. A0v and B0a vs. B0v, P = NS). The pO2 content was higher in A5v samples than in A5a samples (83.5 ± 11.2 vs. 88.8 ± 14.0 mm Hg, P < 0.02), while the level of HCO3 was higher in A5a samples than in A5v samples (20.8 ± 2.0 vs. 20.4 ± 1.8 mEq/L, P < 0.05). A10a samples possessed a higher pH and lower levels of pO2, pCO2, and HCO3 in comparison to A10v samples (P < 0.001 for all). Mean pO2 and pCO2 values in A5v and A10v samples increased by 6.3% and 12.1% and by 1.29% and 52% in comparison to corresponding values of A5a and A10a samples, respectively. The pO2 level was the only parameter that differed significantly between B5a and B5v samples (B5a = 84.6 ± 10.1 vs. B5v = 98.0 ± 12.6 mm Hg, P < 0.005). B10a samples possessed a higher pH and lower levels of pCO2, pO2, and HCO3 in comparison to B10v samples (P < 0.0005 for all comparisons). Mean pO2 and pCO2 values in B5v and B10v samples increased by 16.2% and 16.3% and by −0.29% and 64.8% in comparison to corresponding values of B5a and B10a samples, respectively. C5a samples possessed a higher pH and lower levels of pCO2, pO2, and HCO3 in comparison to C5v samples (P < 0.001 for all). Mean pO2 and pCO2 values in C5v samples were, respectively, 16.0% and 65.0% higher than corresponding values of C5a samples. These results indicate that blood returning from the dialyzer after 5 min of high‐flux hemodialysis has a higher pO2 and pCO2 than blood entering the dialyzer, and that this difference is due to O2 and CO2 transfer from the dialysate space into the blood.
Title: pO2 and pCO2 Increment in Post‐dialyzer Blood: The Role of Dialysate
Description:
Abstract: Blood returning from a dialyzer during hemodialysis has a higher pO2 and pCO2 content than blood entering the dialyzer, and this has been attributed to the dialysate.
The present study investigates this phenomenon.
Acid–base and blood‐gas parameters (pH, pO2, pCO2 and HCO3) were measured in three groups of stable chronic hemodialysis patients (A, B, and C) undergoing high‐flux hemodialysis.
In group A (n = 15), “arterial” (a) and “venous” (v) samples were withdrawn simultaneously before dialysis (samples A0), 5 min after circulation of the blood with the dialysate in the by‐pass mode (samples A5), and 5 min after high‐flux hemodialysis at a zero ultrafiltration rate (samples A10).
In group B (n = 11) (a) and (v) samples were withdrawn simultaneously before dialysis (samples B0), 5 min after isolated‐ultrafiltration with closed dialysate ports (“isolated‐closed” ultrafiltration) (samples B5), and 5 min after high‐flux hemodialysis at a zero ultrafiltration rate (samples B10).
In group C (n = 14), after an initial arterial blood sample withdrawal before hemodialysis (sample C0), high‐flux hemodialysis at a zero ultrafiltration rate was initiated.
Five minutes later, blood and dialysate samples were withdrawn simultaneously from the hemodialysis lines (samples C5).
In all cases blood and dialysate (bicarbonate) flow rates were set at 0.
300 and 0.
700 L/min, respectively.
FLX‐18 hemodialyzers (membrane PEPA 1.
8 m2) were used in this study.
Analysis of variance revealed significant changes only in venous samples.
A comparison of arterial and venous samples revealed no differences between groups A and B before the initiation of dialysis (A0a vs.
A0v and B0a vs.
B0v, P = NS).
The pO2 content was higher in A5v samples than in A5a samples (83.
5 ± 11.
2 vs.
88.
8 ± 14.
0 mm Hg, P < 0.
02), while the level of HCO3 was higher in A5a samples than in A5v samples (20.
8 ± 2.
0 vs.
20.
4 ± 1.
8 mEq/L, P < 0.
05).
A10a samples possessed a higher pH and lower levels of pO2, pCO2, and HCO3 in comparison to A10v samples (P < 0.
001 for all).
Mean pO2 and pCO2 values in A5v and A10v samples increased by 6.
3% and 12.
1% and by 1.
29% and 52% in comparison to corresponding values of A5a and A10a samples, respectively.
The pO2 level was the only parameter that differed significantly between B5a and B5v samples (B5a = 84.
6 ± 10.
1 vs.
B5v = 98.
0 ± 12.
6 mm Hg, P < 0.
005).
B10a samples possessed a higher pH and lower levels of pCO2, pO2, and HCO3 in comparison to B10v samples (P < 0.
0005 for all comparisons).
Mean pO2 and pCO2 values in B5v and B10v samples increased by 16.
2% and 16.
3% and by −0.
29% and 64.
8% in comparison to corresponding values of B5a and B10a samples, respectively.
C5a samples possessed a higher pH and lower levels of pCO2, pO2, and HCO3 in comparison to C5v samples (P < 0.
001 for all).
Mean pO2 and pCO2 values in C5v samples were, respectively, 16.
0% and 65.
0% higher than corresponding values of C5a samples.
These results indicate that blood returning from the dialyzer after 5 min of high‐flux hemodialysis has a higher pO2 and pCO2 than blood entering the dialyzer, and that this difference is due to O2 and CO2 transfer from the dialysate space into the blood.
Related Results
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Cometary Physics Laboratory: spectrophotometric experiments
Cometary Physics Laboratory: spectrophotometric experiments
<p><strong><span dir="ltr" role="presentation">1. Introduction</span></strong&...
Role of Circulatory Hypoxia in the Development of Retinopathy of Prematurity
Role of Circulatory Hypoxia in the Development of Retinopathy of Prematurity
Purpose: to study the pathogenesis of retinopathy of prematurity.Patients and methods.642 preterm infants born before 30 weeks gestation with the weight less 1500 grams were exami...
Atmospheric CO2 flux and planktonic food web relationships in temperate marsh systems: Insights from in situ water measurements
Atmospheric CO2 flux and planktonic food web relationships in temperate marsh systems: Insights from in situ water measurements
Abstract
While research has extensively investigated the dynamics of CO2 water partial pressure (pCO2) and planktonic food webs (PFWs) separately, there has been limited ex...
Haemodialysis with low dialysate flow rates: A comparison of high performance membranes and conventional membranes
Haemodialysis with low dialysate flow rates: A comparison of high performance membranes and conventional membranes
Summary: We performed experimental studies to examine the effects of low dialysate flow rates on the clearance of small molecular weight substances, a middle molecular weight subs...
A STUDY ON DIALYSATE GLUCOSE CONCENTRATION IN PATIENTS WITH CONTINUOUS AMBULATORY PERITONEAL DIALYSIS
A STUDY ON DIALYSATE GLUCOSE CONCENTRATION IN PATIENTS WITH CONTINUOUS AMBULATORY PERITONEAL DIALYSIS
Background: Glucose-based peritoneal dialysis (PD) is the predominantly used dialysate in PD patients. Glucose absorption in peritoneal dialysis (PD) patients may contribute to adv...
HCO3 increment in arterial line can reveal significant vascular access recirculation in high‐flux hemodialysis: A preliminary report
HCO3 increment in arterial line can reveal significant vascular access recirculation in high‐flux hemodialysis: A preliminary report
AbstractWe report a new and simple way that can reveal the presence of vascular access recirculation (VAR) in patients undergoing hemodialysis (HD). Acid‐base and blood gas paramet...
Effect of ocean acidification on early life stages of Atlantic herring (Clupea harengus L.)
Effect of ocean acidification on early life stages of Atlantic herring (Clupea harengus L.)
Abstract. Due to atmospheric accumulation of anthropogenic CO2 the partial pressure of carbon dioxide (pCO2) in surface seawater increases and the pH decreases. This process known ...