Automated Slice Thickness Measurement on the Nessoft CT QA Phantom
View through CrossRef
Purpose: This study proposes a method for automatically measuring slice thickness on image of the Neusoft CT quality assurance (QA) phantom.
Method: The Neusoft CT QA phantom was scanned by a Neuviz 16-slice Neusoft CT Scanner. Automated measurement was implemented using IndoQCT software, while manual measurement was conducted using MicroDicom viewer as comparison. The system was evaluated on images with variations of slice thickness (i.e 1.25, 2.5, 3, 5, and 10 mm), tube voltage (i.e. 80,100,120, and 140 kV), and tube current (i.e. 77, 154, 231, and 233 mA).
Results: The results of automated slice thickness method for slice thicknesses of 1.25, 2.5, 3, 5, and 10 mm were 1.47 + 0.17, 2.67 + 0.08, 3.21 + 0.17, 5.21 + 0.13, and 10.95 + 0.28 mm, respectively. By comparison, the results of manual slice thickness method were 2.91 + 0.17, 3.28 + 0.29, 3.56 + 0.29, 4.72 + 0.27, and 11.35 + 2.03, respectively (p-value 0.009, 0.002, 0.047, 0.008, and 0.714). The results of automated method for tube voltages of 80,100,120, and 140 kV were 5.61 + 0.34, 5.12 + 0.23, 5.08 + 0.23, and 4.98 + 0.28 mm. By comparison, the manual slice thickness method results were 4.71 + 0.39, 4.82 + 0.54, 4.89 + 0.50, and 4.79 + 0.43 mm (p-value 0.005, 0.291, 0.473, 0.452). The results of automated method for tube currents of 77, 154, 231, and 233 mA were 5.19 + 0.26, 4.98 + 0.28, 5.06 + 0.41, and 4.96 + 0.13 mm. By comparison, the results of manual slice thickness method were 4.42 + 0.34, 4.92 + 0.11, 4.72 + 0.37, and 4.80 + 0.46 mm (p-value 0.004, 0.642, 0.22, 0.75).
Conclusions: An automated slice thickness measurement on Neusoft CT the QA phantom image was successfully developed. The measurement results of the automated method are closer to the set thickness than the manual method. The results of automatic slice thickness method are accurate for tube voltage, and tube current variations.
Title: Automated Slice Thickness Measurement on the Nessoft CT QA Phantom
Description:
Purpose: This study proposes a method for automatically measuring slice thickness on image of the Neusoft CT quality assurance (QA) phantom.
Method: The Neusoft CT QA phantom was scanned by a Neuviz 16-slice Neusoft CT Scanner.
Automated measurement was implemented using IndoQCT software, while manual measurement was conducted using MicroDicom viewer as comparison.
The system was evaluated on images with variations of slice thickness (i.
e 1.
25, 2.
5, 3, 5, and 10 mm), tube voltage (i.
e.
80,100,120, and 140 kV), and tube current (i.
e.
77, 154, 231, and 233 mA).
Results: The results of automated slice thickness method for slice thicknesses of 1.
25, 2.
5, 3, 5, and 10 mm were 1.
47 + 0.
17, 2.
67 + 0.
08, 3.
21 + 0.
17, 5.
21 + 0.
13, and 10.
95 + 0.
28 mm, respectively.
By comparison, the results of manual slice thickness method were 2.
91 + 0.
17, 3.
28 + 0.
29, 3.
56 + 0.
29, 4.
72 + 0.
27, and 11.
35 + 2.
03, respectively (p-value 0.
009, 0.
002, 0.
047, 0.
008, and 0.
714).
The results of automated method for tube voltages of 80,100,120, and 140 kV were 5.
61 + 0.
34, 5.
12 + 0.
23, 5.
08 + 0.
23, and 4.
98 + 0.
28 mm.
By comparison, the manual slice thickness method results were 4.
71 + 0.
39, 4.
82 + 0.
54, 4.
89 + 0.
50, and 4.
79 + 0.
43 mm (p-value 0.
005, 0.
291, 0.
473, 0.
452).
The results of automated method for tube currents of 77, 154, 231, and 233 mA were 5.
19 + 0.
26, 4.
98 + 0.
28, 5.
06 + 0.
41, and 4.
96 + 0.
13 mm.
By comparison, the results of manual slice thickness method were 4.
42 + 0.
34, 4.
92 + 0.
11, 4.
72 + 0.
37, and 4.
80 + 0.
46 mm (p-value 0.
004, 0.
642, 0.
22, 0.
75).
Conclusions: An automated slice thickness measurement on Neusoft CT the QA phantom image was successfully developed.
The measurement results of the automated method are closer to the set thickness than the manual method.
The results of automatic slice thickness method are accurate for tube voltage, and tube current variations.
Related Results
Body image and perception among adults with and without phantom limb pain
Body image and perception among adults with and without phantom limb pain
AbstractBackgroundFollowing lower‐limb amputation, phantom limb pain (i.e., pain perceived as coming from the amputated portion of the limb) is common. Phantom limb pain may be ass...
Anatomical Information of Slice Thickness Variation in Ischemic Stroke with CT Scan
Anatomical Information of Slice Thickness Variation in Ischemic Stroke with CT Scan
Ischemic stroke causes partial or complete blockage of blood flow to the brain. Signs of ischemic stroke can be observed on CT scans of the head, namely Hyperdense Artery Sign (HAS...
Distance between Slices Investigation on Siemens CT Phantom Images
Distance between Slices Investigation on Siemens CT Phantom Images
This study aims to develop an algorithm for automatic measurement of distance between slices on a Siemens CT phantom with variations of slice thickness and field of view (FOV). The...
Comparison of MTFs Measured using IndoQCT and ImQuest Software on GE CT Phantom Images
Comparison of MTFs Measured using IndoQCT and ImQuest Software on GE CT Phantom Images
Purpose: This study aims to compare the modulation transfer function (MTF) measured using IndoQCT and ImQuest software on GE CT phantom images with variations of slice thickness an...
SU‐FF‐I‐89: Effects of EPI Slice Angle, Slice Thickness and Phase Encoding Direction On FMRI Sensitivity in Areas with Susceptibility Artifact
SU‐FF‐I‐89: Effects of EPI Slice Angle, Slice Thickness and Phase Encoding Direction On FMRI Sensitivity in Areas with Susceptibility Artifact
Purpose: A key problem of using EPI for BOLD fMRI is the field inhomogeneities near air/tissue interfaces, or the susceptibility artifact. Therefore, fMRI studies in brain areas su...
The Effect of Slice Thickness Variation on The Anatomical Information of CT Scan Paranasal Sinus Coronal Section in Clinical Rhinosinusitis
The Effect of Slice Thickness Variation on The Anatomical Information of CT Scan Paranasal Sinus Coronal Section in Clinical Rhinosinusitis
CT Scan is an excellent technique in assessing anatomical structures, anatomical abnormalities, and visualising the presence or absence of pathological tissue in the four sinuses a...
ANALISIS ANALISIS PENGARUH VARIASI SLICE THICKNESS TERHADAP KUALITAS CITRA DAN INFORMASI ANATOMI PEMERIKSAAN MSCT SCAN THORAX PADA KASUS EMPHYSEMA DI RSD MANGUSADA BADUNG
ANALISIS ANALISIS PENGARUH VARIASI SLICE THICKNESS TERHADAP KUALITAS CITRA DAN INFORMASI ANATOMI PEMERIKSAAN MSCT SCAN THORAX PADA KASUS EMPHYSEMA DI RSD MANGUSADA BADUNG
One of the clinical features that can be found in Thorax is emphysema. Emphysema is one of the pathological processes of the lung parenchyma in COPD (Chronic Obstructive Pulmonary ...
Evaluation and validation of tungsten fiducial marker-based image-guided radiotherapy
Evaluation and validation of tungsten fiducial marker-based image-guided radiotherapy
Abstract
In this research work, a simple homemade cubic phantom was designed to validate the Image-Guided Radiotherapy (IGRT) set up and verified with the help of...