Evaluation of noninvasive electrophysiological imaging accuracy for focal atrial arrhythmias

View through CrossRef
Abstract Background High accuracy of noninvasive electrocardiographic imaging (ECGI) has recently been shown for topical diagnostics of ventricular arrhythmias. However, the precision of diagnostics of atrial focal arrhythmias requires clarification. To estimate the accuracy of ECGI for premature atrial contraction (PAC) we performed atrial pacing in patients with CRT system and compared early activation zone (EAZ) with pacemaker's tip location. Purpose To determine the accuracy of ECGI for focal atrial arrhythmias using atrial pacing. Methods Twenty-six patients (m/f – 18/9), age (min–max) 52 (26–78) with CRT system and pacemaker's tip location in the right atrium (RA) appendage underwent ECGI (“Amycard 01C”) in combination with CT or MR imaging. Thirty-four atrial pacing (mono- and bipolar) was performed in all patients using standard amplitude 1.5–3.8 mV. Epi-/endocardial polygonal heart models were created and isopotential maps were calculated. The distance between EAZ and the pacemaker's tip were measured for ECG recordings without using the isoline filter on endocardial surface (Fig. 1) as well as for epicardial surface. The time between epicardial and endocardial EAZ breakthrough was calculated also. Results On endocardial surface the EAZ was located in RA appendage, the base of superior cava vena or superior lateral RA wall. The distance (mm) (Me (min; max)) between EAZ and the pacemacer's tip was 28 (6; 68). For epicardial surface in most cases the EAZ was also located in RA appendage, the base of superior cava vena or superior lateral RA wall. In two cases the EAZ was located in inferior septal RA wall, in one case - in superior septal RA wall and in five cases the EAZ was undetectable. The distance between EAZ and the pacemacer's tip was 22 (6; 48). The time (ms) (Mean; Me (min; max)) between EAZ of the endocardial and epicardial surfaces was 16; 7 (0; 68). Conclusion ECGI allows to assess the location of focal atrial arrhythmias on endocardial surface and sometimes on epicardial surface also within the three segments. The results of this study revealed that accuracy of ECGI for atrial arrhythmias is worse than for ventricular arrhythmias. However, it is better on epicardial surface of atrium when EAZ can be determined. Funding Acknowledgement Type of funding source: None
Title: Evaluation of noninvasive electrophysiological imaging accuracy for focal atrial arrhythmias
Description:
Abstract Background High accuracy of noninvasive electrocardiographic imaging (ECGI) has recently been shown for topical diagnostics of ventricular arrhythmias.
However, the precision of diagnostics of atrial focal arrhythmias requires clarification.
To estimate the accuracy of ECGI for premature atrial contraction (PAC) we performed atrial pacing in patients with CRT system and compared early activation zone (EAZ) with pacemaker's tip location.
Purpose To determine the accuracy of ECGI for focal atrial arrhythmias using atrial pacing.
Methods Twenty-six patients (m/f – 18/9), age (min–max) 52 (26–78) with CRT system and pacemaker's tip location in the right atrium (RA) appendage underwent ECGI (“Amycard 01C”) in combination with CT or MR imaging.
Thirty-four atrial pacing (mono- and bipolar) was performed in all patients using standard amplitude 1.
5–3.
8 mV.
Epi-/endocardial polygonal heart models were created and isopotential maps were calculated.
The distance between EAZ and the pacemaker's tip were measured for ECG recordings without using the isoline filter on endocardial surface (Fig.
1) as well as for epicardial surface.
The time between epicardial and endocardial EAZ breakthrough was calculated also.
Results On endocardial surface the EAZ was located in RA appendage, the base of superior cava vena or superior lateral RA wall.
The distance (mm) (Me (min; max)) between EAZ and the pacemacer's tip was 28 (6; 68).
For epicardial surface in most cases the EAZ was also located in RA appendage, the base of superior cava vena or superior lateral RA wall.
In two cases the EAZ was located in inferior septal RA wall, in one case - in superior septal RA wall and in five cases the EAZ was undetectable.
The distance between EAZ and the pacemacer's tip was 22 (6; 48).
The time (ms) (Mean; Me (min; max)) between EAZ of the endocardial and epicardial surfaces was 16; 7 (0; 68).
Conclusion ECGI allows to assess the location of focal atrial arrhythmias on endocardial surface and sometimes on epicardial surface also within the three segments.
The results of this study revealed that accuracy of ECGI for atrial arrhythmias is worse than for ventricular arrhythmias.
However, it is better on epicardial surface of atrium when EAZ can be determined.
Funding Acknowledgement Type of funding source: None.

Related Results

Case report: Para Hisian Atrial Tachycardia
Case report: Para Hisian Atrial Tachycardia
Atrial tachycardia accounts for around 10% of supraventricular tachycardias (1). Atrial      tachycardia can either be focal or macrorenentrant. Focal AT usually start at a focal p...
P920Understanding arrhythmia mechanisms in patients with atrial septal defects
P920Understanding arrhythmia mechanisms in patients with atrial septal defects
Abstract Background Atrial arrhythmias represent a major cause of morbidity and hospitalization in patients with atrial septal d...
Inter‐Relationships Between Atrial Flutter and Atrial Fibrillation
Inter‐Relationships Between Atrial Flutter and Atrial Fibrillation
It has been appreciated for a long time that atrial flutter and atrial fibrillation have a clinical relationship. Now, with the technological advances that permit more sophisticate...
Diabetes mellitus and cardiac arrhythmias
Diabetes mellitus and cardiac arrhythmias
Background. The relationship between diabetes mellitus, the development of arrhythmias and cardiac conduction disorders has not been sufficiently studied. Rhythm and conduction dis...
Atrial arrhythmias with mediastinal lymphadenopathy presentation of Isolated atrial myocarditis.
Atrial arrhythmias with mediastinal lymphadenopathy presentation of Isolated atrial myocarditis.
Objectives: We present a case series of patients with granulomatous myocarditis presenting as atrial arrhythmias accompanied by lymphadenopathy. Background: Atrial myocarditis (AM)...
Is There an Association of Atrial Septal Aneurysm with Arrhythmias?
Is There an Association of Atrial Septal Aneurysm with Arrhythmias?
An association of atrial septal aneurysm (ASA) with cardiac arrhythmias has been described, and it has been suggested that undulating movements of the aneurysm initiate these arrhy...
Left atrial appendage anatomy and function: short term response to sustained atrial fibrillation
Left atrial appendage anatomy and function: short term response to sustained atrial fibrillation
OBJECTIVE To determine whether there is significant atrial or atrial appendage enlargement or functional remodelling as a result of one to two months of sustained...
Heart rhythm disorder in patients with obstructive sleep apnoea syndrome
Heart rhythm disorder in patients with obstructive sleep apnoea syndrome
Objective To investigate the risk factors which lead to arrhythmias in patients with obstructive sleep apnoea syndrome (OSAS), to explore incidence and types of a...

Back to Top