Comparison of the Outcome between Lateral, Perimodilar and Midscalar Electrodes in Cochlear Implantation Surgery: A Systematic Review
View through CrossRef
Abstract
Background
Cochlear Implants are neuroprosthetic devices that are surgically implanted receiver-stimulator package and electrode array, which when used with external sound processor can produce a sensation of sound in the recipient in people who are deaf and can't benefit from acoustic hearing aids or middle ear implants or bone conduction devices.
Objective
To compare Lateral, Midscalar, and Perimodiolar electrodes, by using a systematic review as regards: trauma to the cochlea, hearing preservation (if there is a residual hearing).
Material and Methods
The search on Google Scholar, Scopus and PubMed using the following keywords (lateral wall electrode, Midscalar electrode, perimodiolar electrode, cochlear implant, hearing preservation, cochlear trauma).
The inclusion criteria: randomized control, case control, retrospective studies, human living patients, patients with profound sensory neural hearing loss (SNHL). The quality of relevant studies was assessed by NIH quality assessment tool for the type of study (“Study Quality Assessment Tools/National Heart, Lung and Blood Institute (NHLBI),”2019). 186 articles were yielded, from which 77 articles met our criteria. After duplicates removal and restricted to English language, 67 articles were screened in title/Abstract screening, while 31 articles were screened in full text screening were included.
Results
Our study showed that hearing preservation rate with lateral electrode was 53%. Also, hearing preservation with Mid-Scala electrode was 58%. Ratio of cochlear trauma with Perimodiolar electrode was 29% (7%, 55%). Hearing preservation with Lateral electrode versus Perimodiolar electrode odds ratio was 2.20 favoring lateral electrode over Perimodiolar electrode. Hearing preservation with Lateral electrode versus Mid-Scala odd ratio was 0.79 favoring no electrode type over the other. Preserving functional residual hearing in cochlear implant is higher in lateral electrode and Midscalar than in Perimodiolar electrode.
Conclusion
Preserving functional residual hearing in cochlear implant is higher in lateral electrode and Midscalar than in Perimodiolar electrode.
Title: Comparison of the Outcome between Lateral, Perimodilar and Midscalar Electrodes in Cochlear Implantation Surgery: A Systematic Review
Description:
Abstract
Background
Cochlear Implants are neuroprosthetic devices that are surgically implanted receiver-stimulator package and electrode array, which when used with external sound processor can produce a sensation of sound in the recipient in people who are deaf and can't benefit from acoustic hearing aids or middle ear implants or bone conduction devices.
Objective
To compare Lateral, Midscalar, and Perimodiolar electrodes, by using a systematic review as regards: trauma to the cochlea, hearing preservation (if there is a residual hearing).
Material and Methods
The search on Google Scholar, Scopus and PubMed using the following keywords (lateral wall electrode, Midscalar electrode, perimodiolar electrode, cochlear implant, hearing preservation, cochlear trauma).
The inclusion criteria: randomized control, case control, retrospective studies, human living patients, patients with profound sensory neural hearing loss (SNHL).
The quality of relevant studies was assessed by NIH quality assessment tool for the type of study (“Study Quality Assessment Tools/National Heart, Lung and Blood Institute (NHLBI),”2019).
186 articles were yielded, from which 77 articles met our criteria.
After duplicates removal and restricted to English language, 67 articles were screened in title/Abstract screening, while 31 articles were screened in full text screening were included.
Results
Our study showed that hearing preservation rate with lateral electrode was 53%.
Also, hearing preservation with Mid-Scala electrode was 58%.
Ratio of cochlear trauma with Perimodiolar electrode was 29% (7%, 55%).
Hearing preservation with Lateral electrode versus Perimodiolar electrode odds ratio was 2.
20 favoring lateral electrode over Perimodiolar electrode.
Hearing preservation with Lateral electrode versus Mid-Scala odd ratio was 0.
79 favoring no electrode type over the other.
Preserving functional residual hearing in cochlear implant is higher in lateral electrode and Midscalar than in Perimodiolar electrode.
Conclusion
Preserving functional residual hearing in cochlear implant is higher in lateral electrode and Midscalar than in Perimodiolar electrode.
Related Results
Evaluating the Science to Inform the Physical Activity Guidelines for Americans Midcourse Report
Evaluating the Science to Inform the Physical Activity Guidelines for Americans Midcourse Report
Abstract
The Physical Activity Guidelines for Americans (Guidelines) advises older adults to be as active as possible. Yet, despite the well documented benefits of physical activi...
Hydatid Cyst of The Orbit: A Systematic Review with Meta-Data
Hydatid Cyst of The Orbit: A Systematic Review with Meta-Data
Abstarct
Introduction
Orbital hydatid cysts (HCs) constitute less than 1% of all cases of hydatidosis, yet their occurrence is often linked to severe visual complications. This stu...
Objective Measures at Different Stages of Cochlear Implantation: A Data Analysis
Objective Measures at Different Stages of Cochlear Implantation: A Data Analysis
Objectives: The aim of this literature review was to summarize the results of scientific publications on the use of objective electrophysiological methods at different stages of co...
The Therapeutic Dilemma of Cochlear Nerve Deficiency: Cochlear or Brainstem Implantation?
The Therapeutic Dilemma of Cochlear Nerve Deficiency: Cochlear or Brainstem Implantation?
ObjectiveTo compare the outcomes between 2 age‐matched cohorts of children with cochlear nerve deficiency: those receiving auditory brainstem implants (group A) or cochlear implant...
Structured 3D Printed Dry ECG Electrodes Using Copper Based Filament
Structured 3D Printed Dry ECG Electrodes Using Copper Based Filament
Commercial wet Silver and Silver Chloride electrodes are used to monitor electrocardiogram (ECG) signals in numerous bioimpedance applications. These electrodes are frequently sing...
The Effect of Cochlear Coverage on Auditory and Speech Performance in Cochlear Implant Patients
The Effect of Cochlear Coverage on Auditory and Speech Performance in Cochlear Implant Patients
Objective:
To determine the effect of cochlear coverage on audiological and speech parameters in patients with cochlear implants. Previous work has investigated the eff...
Chronic Cochlear Implantation with and without Electric Stimulation in a Mouse Model Induces Robust Cochlear Influx of CX3CR1
+/GFP
Macrophages
Chronic Cochlear Implantation with and without Electric Stimulation in a Mouse Model Induces Robust Cochlear Influx of CX3CR1
+/GFP
Macrophages
Abstract
Background
Cochlear implantation is an effective auditory rehabilitation strategy for those with profound hearing loss...
A novel scoring system for vestibular schwannomas to identify candidacy for cochlear implantation
A novel scoring system for vestibular schwannomas to identify candidacy for cochlear implantation
Objectives Here we present the audiometric outcomes of patients
undergoing vestibular schwannoma resection and cochlear implantation. We
additionally reviewed preoperative audiomet...