Evaluating a virtual reality visual fields analyzer in an urban, underserved glaucoma & glaucoma suspect patient populations to identify disparities
View through CrossRef
Abstract
Purpose
To evaluate the utility of wearable visual field perimetry in an urban, underserved patient population and identify disparities in its utility as a screening tool.
Methods
175 eyes from 105 participants (46 non-glaucomatous eyes from 34 participants and 113 glaucomatous eyes from 74 participants; 16 eyes failed inclusion criteria) presenting at University Hospital in Newark, New Jersey for glaucoma evaluation underwent testing by both the Humphrey Visual Field Analyzer™ (HFA) and PalmScan VF2000 G2™. Glaucoma severity was classified as per the Hoddap criteria. Mean deviation (MD), pattern standard deviation (PSD), visual field index (VFI), mean sensitivity (MS), & area under the receiver operating curve (AUC) on analysis adjusted for inter-eye correlation.
Results
The VF2000 and HFA significantly differed in VFI as the VF2000 consistently underestimated VFI (
p
= 0.003) but did not significantly differ in MD (
p
= 0.664) or PSD (
p
= 0.584). The VF2000 had significantly fewer false positives (
p
< 0.001) and fixation losses (
p
= 0.001) but was a significantly longer exam (
p
= 0.018). On a multivariate logistic regression model adjusting for both inter-eye correlation and demographic variables, the VF2000 had an AUC of 0.7007, indicating fair agreement when identifying severe glaucoma. Language, age, and sex did not independently impact odds of agreement between the two devices; however, differences based on the interaction of age and language were observed.
Conclusion
Our analysis of the Humphrey Visual Field against the virtual reality PalmScan VF2000 G2™ in an urban, diverse population found subtle disparities in predictive staging of glaucoma. Future studies may need to account for these disparities by evaluating the combinations of demographic interactions rather than evaluating them as independent, unrelated factors.
Key messages
What is known
Portable perimetry and virtual reality headsets have been used with moderate efficacy in screenings for glaucoma, but gaps exist in the quality of results as compared to the Humphrey Visual Fields Analyzer.
What is new
The PalmScan VF2000 G2, a portable perimetry headset, may be suitable as a screening device, but it is not advanced enough to differentiate glaucoma stage as effectively as HFA analysis.
Disparities along social determinants of health do exist in VF2000 detection of glaucoma, though these manifestations may be subtle and tied to the interaction of many complex factors.
Future studies may benefit from examining the interaction between demographic factors as variables predictive of outcome.
Title: Evaluating a virtual reality visual fields analyzer in an urban, underserved glaucoma & glaucoma suspect patient populations to identify disparities
Description:
Abstract
Purpose
To evaluate the utility of wearable visual field perimetry in an urban, underserved patient population and identify disparities in its utility as a screening tool.
Methods
175 eyes from 105 participants (46 non-glaucomatous eyes from 34 participants and 113 glaucomatous eyes from 74 participants; 16 eyes failed inclusion criteria) presenting at University Hospital in Newark, New Jersey for glaucoma evaluation underwent testing by both the Humphrey Visual Field Analyzer™ (HFA) and PalmScan VF2000 G2™.
Glaucoma severity was classified as per the Hoddap criteria.
Mean deviation (MD), pattern standard deviation (PSD), visual field index (VFI), mean sensitivity (MS), & area under the receiver operating curve (AUC) on analysis adjusted for inter-eye correlation.
Results
The VF2000 and HFA significantly differed in VFI as the VF2000 consistently underestimated VFI (
p
= 0.
003) but did not significantly differ in MD (
p
= 0.
664) or PSD (
p
= 0.
584).
The VF2000 had significantly fewer false positives (
p
< 0.
001) and fixation losses (
p
= 0.
001) but was a significantly longer exam (
p
= 0.
018).
On a multivariate logistic regression model adjusting for both inter-eye correlation and demographic variables, the VF2000 had an AUC of 0.
7007, indicating fair agreement when identifying severe glaucoma.
Language, age, and sex did not independently impact odds of agreement between the two devices; however, differences based on the interaction of age and language were observed.
Conclusion
Our analysis of the Humphrey Visual Field against the virtual reality PalmScan VF2000 G2™ in an urban, diverse population found subtle disparities in predictive staging of glaucoma.
Future studies may need to account for these disparities by evaluating the combinations of demographic interactions rather than evaluating them as independent, unrelated factors.
Key messages
What is known
Portable perimetry and virtual reality headsets have been used with moderate efficacy in screenings for glaucoma, but gaps exist in the quality of results as compared to the Humphrey Visual Fields Analyzer.
What is new
The PalmScan VF2000 G2, a portable perimetry headset, may be suitable as a screening device, but it is not advanced enough to differentiate glaucoma stage as effectively as HFA analysis.
Disparities along social determinants of health do exist in VF2000 detection of glaucoma, though these manifestations may be subtle and tied to the interaction of many complex factors.
Future studies may benefit from examining the interaction between demographic factors as variables predictive of outcome.
Related Results
Focus on slow rotators - first results from stellar occultations campaign on long-period asteroids
Focus on slow rotators - first results from stellar occultations campaign on long-period asteroids
<p><strong>Most asteroids are slow rotators &#160;</strong> &#160;&#160;&#160;&#160;&#160;...
Cometary Physics Laboratory: spectrophotometric experiments
Cometary Physics Laboratory: spectrophotometric experiments
<p><strong><span dir="ltr" role="presentation">1. Introduction</span></strong&...
North Syrian Mortaria and Other Late Roman Personal and Utility Objects Bearing Inscriptions of Good Luck
North Syrian Mortaria and Other Late Roman Personal and Utility Objects Bearing Inscriptions of Good Luck
<span style="font-size: 11pt; color: black; font-family: 'Times New Roman','serif'">ΠΗΛΙΝΑ ΙΓ&Delta...
Morphometry of an hexagonal pit crater in Pavonis Mons, Mars
Morphometry of an hexagonal pit crater in Pavonis Mons, Mars
<p><strong>Introduction:</strong></p>
<p>Pit craters are peculiar depressions found in almost every terrestria...
Un manoscritto equivocato del copista santo Theophilos († 1548)
Un manoscritto equivocato del copista santo Theophilos († 1548)
<p><font size="3"><span class="A1"><span style="font-family: 'Times New Roman','serif'">ΕΝΑ ΛΑΝ&...
A Touch of Space Weather - Outreach project for visually impaired students
A Touch of Space Weather - Outreach project for visually impaired students
<p><em><span data-preserver-spaces="true">'A Touch of Space Weather' is a project that brings space weather science into...
Ballistic landslides on comet 67P/Churyumov–Gerasimenko
Ballistic landslides on comet 67P/Churyumov–Gerasimenko
<p><strong>Introduction:</strong></p><p>The slow ejecta (i.e., with velocity lower than escape velocity) and l...
Stress transfer process in doublet events studied by numerical TREMOL simulations: Study case Ometepec 1982 Doublet.
Stress transfer process in doublet events studied by numerical TREMOL simulations: Study case Ometepec 1982 Doublet.
<pre class="western"><span><span lang="en-US">Earthquake doublets are a characteristic rupture <...