Rod Lengthening With the Magnetically Controlled Growing Rod
View through CrossRef
Study Design.
Prospective study.
Objective.
To identify the factors that are associated with rod slippage and to study the pattern of achieved length gain with a standard distraction methodology.
Summary of Background Data.
Ability to achieve successful magnetically controlled growing rod (MCGR) distraction is crucial for gradual spine lengthening. Rod slippage has been described as a failure of internal magnet rotation leading to a slippage and an inability to distract the rod. However, its onset, significance, and risk factors are currently unknown. In addition, how this phenomenon pertains to actual distracted lengths is also unknown.
Methods.
A total of 22 patients with MCGR and at least six distraction episodes were prospectively studied. Patients with rod slippage occurring less than six distraction episodes were considered early rod slippage whereas those with more than six episodes or have yet to slip were grouped as late rod slippage. The association of parameters including body habitus, maturity status, age of implantation, total number of distractions, months of distraction from initial implantation, initial and postoperative Cobb angle, T1-T12, T1-S1, T5-T12 kyphosis, curve flexibility, instrumented length, and distance between magnets in dual rods and between the magnets and apex of the curve with early or late onset of rod slippage were studied. Differences between expected and achieved distraction lengths were assessed with reference to rod slippage episodes and rod exchanges to determine any patterns of diminishing returns.
Results.
Patients had mean age of 7.1 years at diagnosis with mean follow-up of 49.8 months. A mean 32.4 distractions were performed per patient. Early rod slippage occurred in 14 patients and late rod slippage occurred in eight patients. Increased height, weight, body mass index, older age, increased T1–12 and T1-S1 lengths, and less distance between magnets were significantly associated with early rod slippage. Expected distraction lengths did not translate to achieve distraction lengths and reduced gains were only observed after achieving one-third of the allowable distracted length in the MCGR. Length gains return to baseline after rod exchange.
Conclusion.
This is the first study to specifically analyze the impact of rod slippage on distraction lengths and the risk factors associated with its onset and frequency. Increased body habitus and reduced distance between internal magnets significantly influenced rod slippage events. Diminishing returns in distracted length gains were only observed after a period of usage.
Level of Evidence: 3
Ovid Technologies (Wolters Kluwer Health)
Title: Rod Lengthening With the Magnetically Controlled Growing Rod
Description:
Study Design.
Prospective study.
Objective.
To identify the factors that are associated with rod slippage and to study the pattern of achieved length gain with a standard distraction methodology.
Summary of Background Data.
Ability to achieve successful magnetically controlled growing rod (MCGR) distraction is crucial for gradual spine lengthening.
Rod slippage has been described as a failure of internal magnet rotation leading to a slippage and an inability to distract the rod.
However, its onset, significance, and risk factors are currently unknown.
In addition, how this phenomenon pertains to actual distracted lengths is also unknown.
Methods.
A total of 22 patients with MCGR and at least six distraction episodes were prospectively studied.
Patients with rod slippage occurring less than six distraction episodes were considered early rod slippage whereas those with more than six episodes or have yet to slip were grouped as late rod slippage.
The association of parameters including body habitus, maturity status, age of implantation, total number of distractions, months of distraction from initial implantation, initial and postoperative Cobb angle, T1-T12, T1-S1, T5-T12 kyphosis, curve flexibility, instrumented length, and distance between magnets in dual rods and between the magnets and apex of the curve with early or late onset of rod slippage were studied.
Differences between expected and achieved distraction lengths were assessed with reference to rod slippage episodes and rod exchanges to determine any patterns of diminishing returns.
Results.
Patients had mean age of 7.
1 years at diagnosis with mean follow-up of 49.
8 months.
A mean 32.
4 distractions were performed per patient.
Early rod slippage occurred in 14 patients and late rod slippage occurred in eight patients.
Increased height, weight, body mass index, older age, increased T1–12 and T1-S1 lengths, and less distance between magnets were significantly associated with early rod slippage.
Expected distraction lengths did not translate to achieve distraction lengths and reduced gains were only observed after achieving one-third of the allowable distracted length in the MCGR.
Length gains return to baseline after rod exchange.
Conclusion.
This is the first study to specifically analyze the impact of rod slippage on distraction lengths and the risk factors associated with its onset and frequency.
Increased body habitus and reduced distance between internal magnets significantly influenced rod slippage events.
Diminishing returns in distracted length gains were only observed after a period of usage.
Level of Evidence: 3.
Related Results
Ribbon Rod - Improvement in Sucker Rod Technology Shows Need to Re-Evaluate Current Artificial Lift Installations
Ribbon Rod - Improvement in Sucker Rod Technology Shows Need to Re-Evaluate Current Artificial Lift Installations
Abstract
This study describes ribbon rod as a continuous graphite composite material that can be currently used to partially replace a conventional sucker rod str...
Rod bipolar cells in the macaque monkey retina: immunoreactivity and connectivity
Rod bipolar cells in the macaque monkey retina: immunoreactivity and connectivity
Rod bipolar cells in the macaque monkey retina were labeled by three antibodies: an antibody against the alpha- and beta-subspecies of protein kinase C (PKC), a polyclonal antiseru...
Effects of Criterion Values on Estimation of the Radius of Drainage and Stabilization Time
Effects of Criterion Values on Estimation of the Radius of Drainage and Stabilization Time
Abstract
After a well starts flowing, a larger portion of the reservoir contributes to production. At any given time, the radius of the portion of the reservoir t...
Effects of Criterion Values on Estimation of the Radius of Drainage and Stabilization Time
Effects of Criterion Values on Estimation of the Radius of Drainage and Stabilization Time
Abstract
After a well starts flowing, a larger and larger portion of the reservoir contributes to production. At any given time, the radius of the portion of the ...
In VitroBiomechanical Comparison of 3.5 mm LC‐DCP/Intramedullary Rod and 5 mm Clamp‐Rod Internal Fixator (CRIF)/Intramedullary Rod Fixation in a Canine Femoral Gap Model
In VitroBiomechanical Comparison of 3.5 mm LC‐DCP/Intramedullary Rod and 5 mm Clamp‐Rod Internal Fixator (CRIF)/Intramedullary Rod Fixation in a Canine Femoral Gap Model
ObjectiveTo compare the biomechanical properties of clamp rod internal fixation (CRIF)/rod and LC‐DCP/rod constructs in a canine femoral gap model.Study DesignCadaveric biomechanic...
Bent Telescopic Rods in Patients With Osteogenesis Imperfecta
Bent Telescopic Rods in Patients With Osteogenesis Imperfecta
Background:
Telescopic rods require alignment of 2 rods to enable lengthening. A telescopic rod converts functionally into a solid rod if either rod bends, preventing p...
Peak power of muscles injured by lengthening contractions
Peak power of muscles injured by lengthening contractions
AbstractExcessive or extreme lengthening contractions have a well‐characterized depressive effect on skeletal muscle isometric force. In addition to producing force, active muscles...
Magnetically Controlled Capsule Endoscopy Versus Conventional Gastroscopy
Magnetically Controlled Capsule Endoscopy Versus Conventional Gastroscopy
Background:
The introduction of magnetically controlled capsule endoscopy overcame the restriction of passive capsule endoscopy movement, thus allowing an improved visu...