Improvement of image quality applying iterative scatter correction for grid-less skeletal radiography in trauma room setting
View through CrossRef
Background
Iterative reconstruction is well established for CT. Plain radiography also takes advantage of iterative algorithms to reduce scatter radiation and improve image quality. First applications have been described for bedside chest X-ray. A recent experimental approach also provided proof of principle for skeletal imaging.
Purpose
To examine clinical applicability of iterative scatter correction for skeletal imaging in the trauma setting.
Material and Methods
In this retrospective single-center study, 209 grid-less radiographs were routinely acquired in the trauma room for 12 months, with imaging of the chest (n = 31), knee (n = 111), pelvis (n = 14), shoulder (n = 24), and other regions close to the trunk (n = 29). Radiographs were postprocessed with iterative scatter correction, doubling the number of images. The radiographs were then independently evaluated by three radiologists and three surgeons. A five-step rating scale and visual grading characteristics analysis were used. The area under the VGC curve (AUC
VGC
) quantified differences in image quality.
Results
Images with iterative scatter correction were generally rated significantly better (AUC
VGC
= 0.59, P < 0.01). This included both radiologists (AUC
VGC
= 0.61, P < 0.01) and surgeons (AUC
VGC
= 0.56, P < 0.01). The image-improving effect was significant for all body regions; in detail: chest (AUC
VGC
= 0.64, P < 0.01), knee (AUC
VGC
= 0.61, P < 0.01), pelvis (AUC
VGC
= 0.60, P = 0.01), shoulder (AUC
VGC
= 0.59, P = 0.02), and others close to the trunk (AUC
VGC
= 0.59, P < 0.01).
Conclusion
Iterative scatter correction improves the image quality of grid-less skeletal radiography in the clinical setting for a wide range of body regions. Therefore, iterative scatter correction may be the future method of choice for free exposure imaging when an anti-scatter grid is omitted due to high risk of tube-detector misalignment.
Title: Improvement of image quality applying iterative scatter correction for grid-less skeletal radiography in trauma room setting
Description:
Background
Iterative reconstruction is well established for CT.
Plain radiography also takes advantage of iterative algorithms to reduce scatter radiation and improve image quality.
First applications have been described for bedside chest X-ray.
A recent experimental approach also provided proof of principle for skeletal imaging.
Purpose
To examine clinical applicability of iterative scatter correction for skeletal imaging in the trauma setting.
Material and Methods
In this retrospective single-center study, 209 grid-less radiographs were routinely acquired in the trauma room for 12 months, with imaging of the chest (n = 31), knee (n = 111), pelvis (n = 14), shoulder (n = 24), and other regions close to the trunk (n = 29).
Radiographs were postprocessed with iterative scatter correction, doubling the number of images.
The radiographs were then independently evaluated by three radiologists and three surgeons.
A five-step rating scale and visual grading characteristics analysis were used.
The area under the VGC curve (AUC
VGC
) quantified differences in image quality.
Results
Images with iterative scatter correction were generally rated significantly better (AUC
VGC
= 0.
59, P < 0.
01).
This included both radiologists (AUC
VGC
= 0.
61, P < 0.
01) and surgeons (AUC
VGC
= 0.
56, P < 0.
01).
The image-improving effect was significant for all body regions; in detail: chest (AUC
VGC
= 0.
64, P < 0.
01), knee (AUC
VGC
= 0.
61, P < 0.
01), pelvis (AUC
VGC
= 0.
60, P = 0.
01), shoulder (AUC
VGC
= 0.
59, P = 0.
02), and others close to the trunk (AUC
VGC
= 0.
59, P < 0.
01).
Conclusion
Iterative scatter correction improves the image quality of grid-less skeletal radiography in the clinical setting for a wide range of body regions.
Therefore, iterative scatter correction may be the future method of choice for free exposure imaging when an anti-scatter grid is omitted due to high risk of tube-detector misalignment.
Related Results
KONTESTASI TASAWUF SUNNÎ DAN TASAWUF FALSAFÎ DI NUSANTARA
KONTESTASI TASAWUF SUNNÎ DAN TASAWUF FALSAFÎ DI NUSANTARA
<p>This article scrutinizes the history of Islamic development in Nusantara between 15th to 18th centuries, which has been colored from theological mysticism thought. Uniquel...
Korelasi Kadar Karboksihemoglobin terhadap Tekanan Darah Penduduk di Sekitar Terminal Bus Tirtonadi Surakarta
Korelasi Kadar Karboksihemoglobin terhadap Tekanan Darah Penduduk di Sekitar Terminal Bus Tirtonadi Surakarta
<table width="645" border="1" cellspacing="0" cellpadding="0"><tbody><tr><td valign="top" width="408"><p> </p><p>Carbon monoxide is a gas ...
CONJUNTURA
CONJUNTURA
<!--[if gte mso 9]><xml> <o:DocumentProperties> <o:Revision>0</o:Revision> <o:TotalTime>0</o:TotalTime> <o:Pages>1</o:Pages> &...
Single‐scan patient‐specific scatter correction in computed tomography using peripheral detection of scatter and compressed sensing scatter retrieval
Single‐scan patient‐specific scatter correction in computed tomography using peripheral detection of scatter and compressed sensing scatter retrieval
Purpose:X‐ray scatter results in a significant degradation of image quality in computed tomography (CT), representing a major limitation in cone‐beam CT (CBCT) and large field‐of‐v...
Efeitos do Pilates solo e exercício resistido sobre a obesidade central e o índice de massa corpórea em idosos
Efeitos do Pilates solo e exercício resistido sobre a obesidade central e o índice de massa corpórea em idosos
<!--[if gte mso 9]><xml> <w:WordDocument> <w:View>Normal</w:View> <w:Zoom>0</w:Zoom> <w:TrackMoves /> <w:TrackFormatting /> &l...
Tratamento de fratura mandibular por projétil de arma de fogo com uso de fixador externo: relato de caso
Tratamento de fratura mandibular por projétil de arma de fogo com uso de fixador externo: relato de caso
<!--[if gte mso 9]><xml> <o:OfficeDocumentSettings> <o:RelyOnVML /> <o:AllowPNG /> </o:OfficeDocumentSettings> </xml><![endif]--> &l...
Iterative scatter correction for grid-less skeletal radiography allows improved image quality equal to an antiscatter grid in adjunct with dose reduction: a visual grading study of 20 body donors
Iterative scatter correction for grid-less skeletal radiography allows improved image quality equal to an antiscatter grid in adjunct with dose reduction: a visual grading study of 20 body donors
Background
Iterative scatter correction (ISC) is a new technique applicable to plain radiography; comparable to iterative reconstruction for computed tomography...
Three Dimensional Simulations in Real Time for Personalized Drug Release Prosthesis Used in Lumbosacral Rehabilitation
Three Dimensional Simulations in Real Time for Personalized Drug Release Prosthesis Used in Lumbosacral Rehabilitation
This paper presents a theoretical method for simulation and three-dimensional reconstruction of the anatomical elements of the spine in order to achieve hydrogel disc prosthesis by...