P799A new invasive validation method for non-invasive central blood pressure measurement using a suprasystolic sphygmomanometer

View through CrossRef
Abstract Background Central blood pressure (cBP) is a better predictor of the damage caused by hypertension in comparison with peripheral blood pressure (pBP). Although challenging to measure, numerous devices are trying to reliably estimate cBP non-invasively. Pulse wave velocity (PWV) is another important independent cardiovascular risk factor. Aim We sought to deploy a new validation method using a high-fidelity pressure wire as the invasive gold standard measurement for sphygmomanometer devices estimating cBP. Moreover, we invasively calculated the PWV to investigate its relationship to the non-invasively estimated cBP. Methods In 50 patients requiring a cardiac catheterization, we measured the blood pressure in the ascending aorta (AAo) with a fluid-filled (FF) guiding catheter (NaCl 0,9%). We compared these values with the results derived simultaneously with a novel sphygmomanometer that estimates cBP from the analysis of brachial artery suprasystolic pressure waves, based on the pressure-wave propagation of a water-hammer acoustic model. This was measured on the left arm with the BP+ device from USCOM Pty Ltd (Sydney, NSW, Australia) while the catheterization was performed via the right radial artery. On 14 of these patients so far, we placed a 0.014" high-fidelity pressure wire in the AAo to measure cBP, when it was clinically indicated to evaluate one or more coronary stenosis by Fractional Flow Reserve (FFR). Ultimately, the wire was pulled back into the humeral artery (HUM). PWV was then calculated from the length of the pullback and the time delay between AAo and HUM pulses by gating to the R-wave of the ECG for both measurements, using MatLab software. Results Bland-Altman analysis of the sphygmomanometrically estimated cBPsys and the measured one by a FFR wire (left on figure) demonstrates less scatter than with the FF catheter (right). The mean difference with the sphygmomanometrically derived cBPsys was −1.2±4.7 mmHg (CI95%: −3.8; 1.8) for the FFR wire and 6.0±9.8 mmHg (CI95%: 3.2; 8.9) for the FF catheter. Central diastolic and mean BP were both overestimated by the sphygmomanometer, with respectively −7.8±6.8 mmHg (CI95%: −11.4; −4.2) and −5.5±6.1 mmHg (CI95%: −8.6; −2.3) compared to the FFR wire and −10.3±6.7 mmHg (CI95%: −12.2; −8.3) and −5.5±6.5 (CI95% : −7.3; −3.6) for the FF catheter. The average PWV was 7.0±1.4 m/s. No significant relationship of PWV and cBP was identified (p=0.189). The PWV was 0.8 m/s lower in patients with only one cardiovascular risk factor versus more than one, but without reaching statistical significance. cBPsys measured. A. FFR; B. FluidF vs BP+ Conclusions Using a FFR wire in the AAo as a high-fidelity pressure reference, we demonstrated that cBPsys derived from this new sphygmomanometer was accurate, with a non-significant bias (<5mmHg) and high precision (standard deviation <8mmHg) as recommended, criteria not met using the FF guiding catheter measurements. PWV measurements were also easily obtained from the FFR wire method.
Title: P799A new invasive validation method for non-invasive central blood pressure measurement using a suprasystolic sphygmomanometer
Description:
Abstract Background Central blood pressure (cBP) is a better predictor of the damage caused by hypertension in comparison with peripheral blood pressure (pBP).
Although challenging to measure, numerous devices are trying to reliably estimate cBP non-invasively.
Pulse wave velocity (PWV) is another important independent cardiovascular risk factor.
Aim We sought to deploy a new validation method using a high-fidelity pressure wire as the invasive gold standard measurement for sphygmomanometer devices estimating cBP.
Moreover, we invasively calculated the PWV to investigate its relationship to the non-invasively estimated cBP.
Methods In 50 patients requiring a cardiac catheterization, we measured the blood pressure in the ascending aorta (AAo) with a fluid-filled (FF) guiding catheter (NaCl 0,9%).
We compared these values with the results derived simultaneously with a novel sphygmomanometer that estimates cBP from the analysis of brachial artery suprasystolic pressure waves, based on the pressure-wave propagation of a water-hammer acoustic model.
This was measured on the left arm with the BP+ device from USCOM Pty Ltd (Sydney, NSW, Australia) while the catheterization was performed via the right radial artery.
On 14 of these patients so far, we placed a 0.
014" high-fidelity pressure wire in the AAo to measure cBP, when it was clinically indicated to evaluate one or more coronary stenosis by Fractional Flow Reserve (FFR).
Ultimately, the wire was pulled back into the humeral artery (HUM).
PWV was then calculated from the length of the pullback and the time delay between AAo and HUM pulses by gating to the R-wave of the ECG for both measurements, using MatLab software.
Results Bland-Altman analysis of the sphygmomanometrically estimated cBPsys and the measured one by a FFR wire (left on figure) demonstrates less scatter than with the FF catheter (right).
The mean difference with the sphygmomanometrically derived cBPsys was −1.
2±4.
7 mmHg (CI95%: −3.
8; 1.
8) for the FFR wire and 6.
0±9.
8 mmHg (CI95%: 3.
2; 8.
9) for the FF catheter.
Central diastolic and mean BP were both overestimated by the sphygmomanometer, with respectively −7.
8±6.
8 mmHg (CI95%: −11.
4; −4.
2) and −5.
5±6.
1 mmHg (CI95%: −8.
6; −2.
3) compared to the FFR wire and −10.
3±6.
7 mmHg (CI95%: −12.
2; −8.
3) and −5.
5±6.
5 (CI95% : −7.
3; −3.
6) for the FF catheter.
The average PWV was 7.
0±1.
4 m/s.
No significant relationship of PWV and cBP was identified (p=0.
189).
The PWV was 0.
8 m/s lower in patients with only one cardiovascular risk factor versus more than one, but without reaching statistical significance.
cBPsys measured.
A.
FFR; B.
FluidF vs BP+ Conclusions Using a FFR wire in the AAo as a high-fidelity pressure reference, we demonstrated that cBPsys derived from this new sphygmomanometer was accurate, with a non-significant bias (<5mmHg) and high precision (standard deviation <8mmHg) as recommended, criteria not met using the FF guiding catheter measurements.
PWV measurements were also easily obtained from the FFR wire method.

Related Results

[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Aim/Purpose: The aim of this article is to make a case of the role of validation in doctoral education. The purpose is to detail findings from three studies which explore PhD stude...
Research on Blood Pressure Monitoring Based on Flexible Encapsulated Sensors
Research on Blood Pressure Monitoring Based on Flexible Encapsulated Sensors
In order to realize the accurate measurement of wearable sphygmomanometer, this paper selects the micro electro mechanical system (MEMS) pressure sensor with small size, low cost a...
Blood Cross Matching Without Anti-Human Globulin (AHG) and Bovine Serum: A New Interest for an Old Idea
Blood Cross Matching Without Anti-Human Globulin (AHG) and Bovine Serum: A New Interest for an Old Idea
Abstract  Introduction Transfusion medicine promotes the safety of blood transfusions by rigorously testing to eliminate risks of infection and hemolytic. The efficacy (to correct ...
Determinants of Cerebrovascular Reserve in Patients with Significant Carotid Stenosis
Determinants of Cerebrovascular Reserve in Patients with Significant Carotid Stenosis
AbstractIntroductionIn patients with 70% to 99% diameter carotid artery stenosis cerebral blood flow reserve may be protective of future ischemic cerebral events. Reserve cerebral ...
Analyzing Well Performance VII
Analyzing Well Performance VII
Introduction Analyzing well performance is an important step toward increasing profits by improving production techniques. The analysis is made by field tests and...

Back to Top