The Effect of Implant Length and Diameter on Stress Distribution around Single Implant Placement in 3D Posterior Mandibular FE Model Directly Constructed Form In Vivo CT

View through CrossRef
A three-dimensional (3D) finite element (FE) model of the mandibular bone was created from 3D X-ray CT scan images of a live human subject. Simulating the clinical situation of implant therapy at the mandibular first molar, virtual extraction of the tooth was performed at the 3D FE mandibular model, and 12 different implant diameters and lengths were virtually inserted in order to carry out a mechanical analysis. (1) High stress concentration was found at the surfaces of the buccal and lingual peri-implant bone adjacent to the sides of the neck in all the implants. (2) The greatest stress value was approximately 6.0 MPa with implant diameter of 3.8 mm, approx. 4.5 MPa with implant diameter of 4.3 mm, and approx. 3.2 MPa with implant diameter of 6.0 mm. (3) The stress on the peri-implant bone was found to decrease with increasing length and mainly in diameter of the implant.
Title: The Effect of Implant Length and Diameter on Stress Distribution around Single Implant Placement in 3D Posterior Mandibular FE Model Directly Constructed Form In Vivo CT
Description:
A three-dimensional (3D) finite element (FE) model of the mandibular bone was created from 3D X-ray CT scan images of a live human subject.
Simulating the clinical situation of implant therapy at the mandibular first molar, virtual extraction of the tooth was performed at the 3D FE mandibular model, and 12 different implant diameters and lengths were virtually inserted in order to carry out a mechanical analysis.
(1) High stress concentration was found at the surfaces of the buccal and lingual peri-implant bone adjacent to the sides of the neck in all the implants.
(2) The greatest stress value was approximately 6.
0 MPa with implant diameter of 3.
8 mm, approx.
4.
5 MPa with implant diameter of 4.
3 mm, and approx.
3.
2 MPa with implant diameter of 6.
0 mm.
(3) The stress on the peri-implant bone was found to decrease with increasing length and mainly in diameter of the implant.

Related Results

Osteometric Measures of Mandible on Its Various Aspects in South Indian Dry Skull
Osteometric Measures of Mandible on Its Various Aspects in South Indian Dry Skull
Introduction: Mandible is the single bone articulated with cranium forming a temporomandibular joint and it contributes to the lower jaw containing teeth. The mandible is composed ...
Evaluation of hybrid implant in posterior edentulous maxilla with inadequate bone
Evaluation of hybrid implant in posterior edentulous maxilla with inadequate bone
Aim of this study is to investigate.Hybrid implant has got satisfactory results in our previous study in various edentulous areas with inadequate bone for endosseous implant placem...
Observation of the effect on immediate mandibular implant-supported overdenture of micro implant system of Italy
Observation of the effect on immediate mandibular implant-supported overdenture of micro implant system of Italy
The study was design to investigate the clinical effect of Micro implant system in Italy on immediate mandibular implant-supported overdenture. A total of 192 patients with mandibu...
SINGLE IMPLANT-RETAINED MANDIBULAR OVERDENTURE: A CASE REPORT
SINGLE IMPLANT-RETAINED MANDIBULAR OVERDENTURE: A CASE REPORT
Edentulism is a public health problem, impacting the quality of life of many people worldwide. Factors related to the mandibular ridge minimal support may hinder the common use of ...

Back to Top