Search engine for discovering works of Art, research articles, and books related to Art and Culture
ShareThis
Javascript must be enabled to continue!

Dislocation force of scleral flange-fixated intraocular lens haptics

View through CrossRef
Abstract Purpose To measure the dislocation forces in relation to haptic material, flange size and needle used. Setting Hanusch Hospital, Vienna, Austria. Design Laboratory Investigation. Methods, main outcome measures 30 G (gauge) thin wall and 27 G standard needles were used for a 2 mm tangential scleral tunnel in combination with different PVDF (polyvinylidene fluoride) and PMMA (polymethylmethacrylate haptics). Flanges were created by heating 1 mm of the haptic end, non-forceps assisted in PVDF and forceps assisted in PMMA haptics. The dislocation force was measured in non-preserved cadaver sclera using a tensiometer device. Results PVDF flanges achieved were of a mushroom-like shape and PMMA flanges were of a conic shape. For 30 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 1.58 ± 0.68 N (n = 10) and 0.70 ± 0.14 N (n = 9) (p = 0.003) respectively. For 27 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 0.31 ± 0.35 N (n = 3) and 0.0 N (n = 4), respectively. The flange size correlated with the occurring dislocation force in experiments with 30 G needle tunnels (r = 0.92), when flanges were bigger than 384 micrometres. Conclusions The highest dislocation forces were found for PVDF haptic flanges and their characteristic mushroom-like shape for 30 G thin wall needle scleral tunnels. Forceps assisted flange creation in PMMA haptics did not compensate the disadvantage of PMMA haptics with their characteristic conic shape flange.
Title: Dislocation force of scleral flange-fixated intraocular lens haptics
Description:
Abstract Purpose To measure the dislocation forces in relation to haptic material, flange size and needle used.
Setting Hanusch Hospital, Vienna, Austria.
Design Laboratory Investigation.
Methods, main outcome measures 30 G (gauge) thin wall and 27 G standard needles were used for a 2 mm tangential scleral tunnel in combination with different PVDF (polyvinylidene fluoride) and PMMA (polymethylmethacrylate haptics).
Flanges were created by heating 1 mm of the haptic end, non-forceps assisted in PVDF and forceps assisted in PMMA haptics.
The dislocation force was measured in non-preserved cadaver sclera using a tensiometer device.
Results PVDF flanges achieved were of a mushroom-like shape and PMMA flanges were of a conic shape.
For 30 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 1.
58 ± 0.
68 N (n = 10) and 0.
70 ± 0.
14 N (n = 9) (p = 0.
003) respectively.
For 27 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 0.
31 ± 0.
35 N (n = 3) and 0.
0 N (n = 4), respectively.
The flange size correlated with the occurring dislocation force in experiments with 30 G needle tunnels (r = 0.
92), when flanges were bigger than 384 micrometres.
Conclusions The highest dislocation forces were found for PVDF haptic flanges and their characteristic mushroom-like shape for 30 G thin wall needle scleral tunnels.
Forceps assisted flange creation in PMMA haptics did not compensate the disadvantage of PMMA haptics with their characteristic conic shape flange.

Related Results

Visual outcome following scleral fixated intraocular lens implantation
Visual outcome following scleral fixated intraocular lens implantation
Introduction: Scleral fixated intraocular lens(SFIOL) implantation to correct aphakia offers superior visual rehabilitation in comparison to aphakic spectacles or contact lens. Thi...
Finger Flange Designs on Cast Equipment
Finger Flange Designs on Cast Equipment
Abstract Piping systems connecting critical equipment such as pumps with non-standard flanges should be designed so that all components have compatible strength. One...
Visual rehabilitation with scleral lenses after corneal transplantation
Visual rehabilitation with scleral lenses after corneal transplantation
Background: Patients undergoing corneal transplantation often present with high astigmatism and corneal irregularity, making conventional contact lens fitting difficult and limitin...
Visual Outcome After Sutureless Scleral Fixation of Intraocular Lens
Visual Outcome After Sutureless Scleral Fixation of Intraocular Lens
Purpose:  To study the results of sutureless scleral fixation of intraocular lens (IOL) in cases of complicated cataract surgery. Study Design:  Interventional case series.&#...
Enhancing Scleral Buckling Surgery Using an Illuminated Scleral Depressor
Enhancing Scleral Buckling Surgery Using an Illuminated Scleral Depressor
Purpose: To report a technique for scleral buckling surgery under the operating microscope with wide-field viewing. Methods: ...
Deep Learning for Joint Scleral and Scleral Vessel Segmentation: A Comparative Study
Deep Learning for Joint Scleral and Scleral Vessel Segmentation: A Comparative Study
Scleral segmentation and scleral vessel segmentation are increasingly recognized as critical components in medical image analysis, with broad applications in ocular disease diagnos...
Scleral Flap Dehiscence
Scleral Flap Dehiscence
Following its introduction by Cairns in 1968, trabeculectomy quickly became the procedure of choice for the management of elevated intraocular pressure (IOP) inadequately controlle...