Dislocation force of scleral flange-fixated intraocular lens haptics
View through CrossRef
Abstract
Purpose
To measure the dislocation forces in relation to haptic material, flange size and needle used.
Setting
Hanusch Hospital, Vienna, Austria.
Design
Laboratory Investigation.
Methods, main outcome measures
30 G (gauge) thin wall and 27 G standard needles were used for a 2 mm tangential scleral tunnel in combination with different PVDF (polyvinylidene fluoride) and PMMA (polymethylmethacrylate haptics). Flanges were created by heating 1 mm of the haptic end, non-forceps assisted in PVDF and forceps assisted in PMMA haptics. The dislocation force was measured in non-preserved cadaver sclera using a tensiometer device.
Results
PVDF flanges achieved were of a mushroom-like shape and PMMA flanges were of a conic shape. For 30 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 1.58 ± 0.68 N (n = 10) and 0.70 ± 0.14 N (n = 9) (p = 0.003) respectively. For 27 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 0.31 ± 0.35 N (n = 3) and 0.0 N (n = 4), respectively. The flange size correlated with the occurring dislocation force in experiments with 30 G needle tunnels (r = 0.92), when flanges were bigger than 384 micrometres.
Conclusions
The highest dislocation forces were found for PVDF haptic flanges and their characteristic mushroom-like shape for 30 G thin wall needle scleral tunnels. Forceps assisted flange creation in PMMA haptics did not compensate the disadvantage of PMMA haptics with their characteristic conic shape flange.
Springer Science and Business Media LLC
Title: Dislocation force of scleral flange-fixated intraocular lens haptics
Description:
Abstract
Purpose
To measure the dislocation forces in relation to haptic material, flange size and needle used.
Setting
Hanusch Hospital, Vienna, Austria.
Design
Laboratory Investigation.
Methods, main outcome measures
30 G (gauge) thin wall and 27 G standard needles were used for a 2 mm tangential scleral tunnel in combination with different PVDF (polyvinylidene fluoride) and PMMA (polymethylmethacrylate haptics).
Flanges were created by heating 1 mm of the haptic end, non-forceps assisted in PVDF and forceps assisted in PMMA haptics.
The dislocation force was measured in non-preserved cadaver sclera using a tensiometer device.
Results
PVDF flanges achieved were of a mushroom-like shape and PMMA flanges were of a conic shape.
For 30 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 1.
58 ± 0.
68 N (n = 10) and 0.
70 ± 0.
14 N (n = 9) (p = 0.
003) respectively.
For 27 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 0.
31 ± 0.
35 N (n = 3) and 0.
0 N (n = 4), respectively.
The flange size correlated with the occurring dislocation force in experiments with 30 G needle tunnels (r = 0.
92), when flanges were bigger than 384 micrometres.
Conclusions
The highest dislocation forces were found for PVDF haptic flanges and their characteristic mushroom-like shape for 30 G thin wall needle scleral tunnels.
Forceps assisted flange creation in PMMA haptics did not compensate the disadvantage of PMMA haptics with their characteristic conic shape flange.
Related Results
Visual outcome following scleral fixated intraocular lens implantation
Visual outcome following scleral fixated intraocular lens implantation
Introduction: Scleral fixated intraocular lens(SFIOL) implantation to correct aphakia offers superior visual rehabilitation in comparison to aphakic spectacles or contact lens. Thi...
Finger Flange Designs on Cast Equipment
Finger Flange Designs on Cast Equipment
Abstract
Piping systems connecting critical equipment such as pumps with non-standard flanges should be designed so that all components have compatible strength. One...
Visual rehabilitation with scleral lenses after corneal transplantation
Visual rehabilitation with scleral lenses after corneal transplantation
Background: Patients undergoing corneal transplantation often present with high astigmatism and corneal irregularity, making conventional contact lens fitting difficult and limitin...
Visual Outcome After Sutureless Scleral Fixation of Intraocular Lens
Visual Outcome After Sutureless Scleral Fixation of Intraocular Lens
Purpose: To study the results of sutureless scleral fixation of intraocular lens (IOL) in cases of complicated cataract surgery.
Study Design: Interventional case series....
Enhancing Scleral Buckling Surgery Using an Illuminated Scleral Depressor
Enhancing Scleral Buckling Surgery Using an Illuminated Scleral Depressor
Purpose:
To report a technique for scleral buckling surgery under the operating microscope with wide-field viewing.
Methods:
...
Clinical, functional and anatomical-topographic results of retinal detachment treatment using method of circular scleral depression and personalized calculation of the length of a circular scleral buckle
Clinical, functional and anatomical-topographic results of retinal detachment treatment using method of circular scleral depression and personalized calculation of the length of a circular scleral buckle
Relevance. Circular scleral buckling (SB) is a surgical technique that ensures adaptation of the retina to the underlying membranes due to the created depression shaft, which reduc...
Deep Learning for Joint Scleral and Scleral Vessel Segmentation: A Comparative Study
Deep Learning for Joint Scleral and Scleral Vessel Segmentation: A Comparative Study
Scleral segmentation and scleral vessel segmentation are increasingly recognized as critical components in medical image analysis, with broad applications in ocular disease diagnos...
Scleral Flap Dehiscence
Scleral Flap Dehiscence
Following its introduction by Cairns in 1968, trabeculectomy quickly became the procedure of choice for the management of elevated intraocular pressure (IOP) inadequately controlle...