Correlation between the setting of gating window width and setup accuracy in left breast cancer radiotherapy based on deep inspiration breath hold
View through CrossRef
AbstractPersonalized precision irradiation of patients with left‐sided breast cancer is possible by examining the setup errors of 3‐ and 4‐mm gated window widths for those treated with deep inspiration breath‐hold (DIBH) treatment. An observational study was performed via a retrospective analysis of 250 cone‐beam computed tomography (CBCT) images of 60 left‐breast cancer patients who underwent whole‐breast radiotherapy with the DIBH technique between January 2021 and 2022 at our hospital. Among them, 30 patients had a gated window width of 3 mm, while the remaining 30 had a gated window width of 4 mm; both groups received radiotherapy using DIBH technology. All patients underwent CBCT scans once a week, and the setup errors in the left‐right (x‐axis), inferior‐superior (y‐axis), and anterior‐posterior (z‐axis) directions were recorded. The clinical‐to‐planning target volume (CTV‐PTV) margins of the two gating windows were calculated using established methods. The setup error in the Y direction was 1.69 ± 1.33 mm for the 3‐mm – wide gated window and 2.42 ± 3.02 mm for the 4‐mm – wide gated window. The two groups had statistically significant differences in the overall mean setup error (Dif 0.7, 95% CI 0.15–1.31, t = 2.48, p= 0.014). The Z‐direction setup error was 2.32 ± 2.12 mm for the 3‐mm – wide gated window and 3.15 ± 3.34 mm for the 4‐mm – wide gated window. The overall mean setup error was statistically significant between the two groups (Dif 0.8, 95% CI 0.13–1.53, t= 2.34, p = 0.020). There was no significant difference in the X‐direction setup error (p > 0.05). Therefore, the CTV‐PTV margin values for a 3‐mm gated window width in the X, Y, and Z directions are 5.51, 5.15, and 7.28 mm, respectively; those for a 4‐mm gated window width in the X, Y, and Z directions are 5.52, 8.16, and 10.21 mm, respectively. The setup errors of the 3‐mm – wide gating window are smaller than those of the 4‐mm – wide gating window in the three dimensions. Therefore, when the patient's respiratory gating window width is reduced, the margin values of CTV‐PTV can be reduced to increase the distance between the PTV and the organs at risk (OARs), which ensures adequate space for the dose to decrease, resulting in lower dose exposure to the OARs (heart, lungs, etc.), thus sparing the OARs from further damage. However, some patients with poor pulmonary function or unstable breathing amplitudes must be treated with a slightly larger gating window. Therefore, this study lays a theoretical basis for personalized precision radiotherapy, which can save time and reduce manpower in the delivery of clinical treatment to a certain extent. Another potential benefit of this work is to bring awareness to the potential implications of a slightly larger gating window during treatment without considering the resulting dosimetric impact.
Title: Correlation between the setting of gating window width and setup accuracy in left breast cancer radiotherapy based on deep inspiration breath hold
Description:
AbstractPersonalized precision irradiation of patients with left‐sided breast cancer is possible by examining the setup errors of 3‐ and 4‐mm gated window widths for those treated with deep inspiration breath‐hold (DIBH) treatment.
An observational study was performed via a retrospective analysis of 250 cone‐beam computed tomography (CBCT) images of 60 left‐breast cancer patients who underwent whole‐breast radiotherapy with the DIBH technique between January 2021 and 2022 at our hospital.
Among them, 30 patients had a gated window width of 3 mm, while the remaining 30 had a gated window width of 4 mm; both groups received radiotherapy using DIBH technology.
All patients underwent CBCT scans once a week, and the setup errors in the left‐right (x‐axis), inferior‐superior (y‐axis), and anterior‐posterior (z‐axis) directions were recorded.
The clinical‐to‐planning target volume (CTV‐PTV) margins of the two gating windows were calculated using established methods.
The setup error in the Y direction was 1.
69 ± 1.
33 mm for the 3‐mm – wide gated window and 2.
42 ± 3.
02 mm for the 4‐mm – wide gated window.
The two groups had statistically significant differences in the overall mean setup error (Dif 0.
7, 95% CI 0.
15–1.
31, t = 2.
48, p= 0.
014).
The Z‐direction setup error was 2.
32 ± 2.
12 mm for the 3‐mm – wide gated window and 3.
15 ± 3.
34 mm for the 4‐mm – wide gated window.
The overall mean setup error was statistically significant between the two groups (Dif 0.
8, 95% CI 0.
13–1.
53, t= 2.
34, p = 0.
020).
There was no significant difference in the X‐direction setup error (p > 0.
05).
Therefore, the CTV‐PTV margin values for a 3‐mm gated window width in the X, Y, and Z directions are 5.
51, 5.
15, and 7.
28 mm, respectively; those for a 4‐mm gated window width in the X, Y, and Z directions are 5.
52, 8.
16, and 10.
21 mm, respectively.
The setup errors of the 3‐mm – wide gating window are smaller than those of the 4‐mm – wide gating window in the three dimensions.
Therefore, when the patient's respiratory gating window width is reduced, the margin values of CTV‐PTV can be reduced to increase the distance between the PTV and the organs at risk (OARs), which ensures adequate space for the dose to decrease, resulting in lower dose exposure to the OARs (heart, lungs, etc.
), thus sparing the OARs from further damage.
However, some patients with poor pulmonary function or unstable breathing amplitudes must be treated with a slightly larger gating window.
Therefore, this study lays a theoretical basis for personalized precision radiotherapy, which can save time and reduce manpower in the delivery of clinical treatment to a certain extent.
Another potential benefit of this work is to bring awareness to the potential implications of a slightly larger gating window during treatment without considering the resulting dosimetric impact.
Related Results
Coexisting Granulomatous Mastitis and Breast Cancer: A Systematic Review
Coexisting Granulomatous Mastitis and Breast Cancer: A Systematic Review
Abstract
Introduction: Granulomatous mastitis (GM) is a rare inflammatory breast disease that mimics carcinoma. GM can coexist with breast cancer (BC), though the relationship rema...
Breast Carcinoma within Fibroadenoma: A Systematic Review
Breast Carcinoma within Fibroadenoma: A Systematic Review
Abstract
Introduction
Fibroadenoma is the most common benign breast lesion; however, it carries a potential risk of malignant transformation. This systematic review provides an ove...
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct
Introduction
Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
Desmoid-Type Fibromatosis of The Breast: A Case Series
Desmoid-Type Fibromatosis of The Breast: A Case Series
Abstract
IntroductionDesmoid-type fibromatosis (DTF), also called aggressive fibromatosis, is a rare, benign, locally aggressive condition. Mammary DTF originates from fibroblasts ...
Improving Setup Precision in DIBH Radiotherapy for Left-Sided Breast Cancer: Tattoo-less vs. Tattoo-Based Surface-Guided and Respiratory Gating Techniques
Improving Setup Precision in DIBH Radiotherapy for Left-Sided Breast Cancer: Tattoo-less vs. Tattoo-Based Surface-Guided and Respiratory Gating Techniques
Introduction: Deep inspiration breath-hold (DIBH) is a widely used technique to minimize radiation exposure to the heart during radiotherapy for left-sided breast cancer. Several c...
Abstract OI-1: OI-1 Decoding breast cancer predisposition genes
Abstract OI-1: OI-1 Decoding breast cancer predisposition genes
Abstract
Women with one or more first-degree female relatives with a history of breast cancer have a two-fold increased risk of developing breast cancer. This risk i...
Spanish Breast Cancer Research Group (GEICAM)
Spanish Breast Cancer Research Group (GEICAM)
This section provides current contact details and a summary of recent or ongoing clinical trials being coordinated by Spanish Breast Cancer Research Group (GEICAM). Clinical trials...
Complex Collision Tumors: A Systematic Review
Complex Collision Tumors: A Systematic Review
Abstract
Introduction: A collision tumor consists of two distinct neoplastic components located within the same organ, separated by stromal tissue, without histological intermixing...