Comparison of the PlusoptiX A16 and vision screener V100

View through CrossRef
Clinical relevanceThis study compares a novel photoscreening device with a previously validated one in a school-age population. It highlights a tendency of the new device to underestimate myopic spherical equivalent and overestimate hyperopic cases.PurposeTo compare the PlusoptiX A16 and Vision Screener V100 photoscreeners in a study population of school-age children.MethodsOne hundred and thirty-three children, with a mean age of 6.4 ± 0.5 years, were evaluated using both the PlusoptiX A16 and Vision Screener V100 photoscreeners. The measurements were taken in random order in a room with diminished ambient lighting.ResultsThe mean refractive error values for the M component were 0.27 ± 0.67D (PlusoptiX A16) and 0.21 ± 0.58D (Vision Screener V100). For the J0 component, means were 0.16 ± 0.38D (PlusoptiX A16) and 0.06 ± 0.33D (Vision Screener V100) and for theJ45 component the means were 0.03 ± 0.17D (PlusoptiX A16) and 0.06 ± 0.22D (Vision Screener V100). When compared both instruments, statistically significant differences were observed for the M (p=0.017) and J0 (p=0.004) components. The agreement rates between PlusoptiX A16 and Vision Screener V100 across different refractive components were 80.5% for sphere, 82.0% for cylinder, and 40.6% for axis when considering a range of ±0.75 D for sphere and cylinder and ±25.0 degrees for cylinder axis. Simultaneously considering all three conditions, the overall agreement was 73.7%.ConclusionThe Vision Screener V100, while generally aligning well with PlusoptiX A16, tends to underestimate myopic spherical equivalent, overestimate hyperopic cases, and underestimate J0 astigmatism.
Title: Comparison of the PlusoptiX A16 and vision screener V100
Description:
Clinical relevanceThis study compares a novel photoscreening device with a previously validated one in a school-age population.
It highlights a tendency of the new device to underestimate myopic spherical equivalent and overestimate hyperopic cases.
PurposeTo compare the PlusoptiX A16 and Vision Screener V100 photoscreeners in a study population of school-age children.
MethodsOne hundred and thirty-three children, with a mean age of 6.
4 ± 0.
5 years, were evaluated using both the PlusoptiX A16 and Vision Screener V100 photoscreeners.
The measurements were taken in random order in a room with diminished ambient lighting.
ResultsThe mean refractive error values for the M component were 0.
27 ± 0.
67D (PlusoptiX A16) and 0.
21 ± 0.
58D (Vision Screener V100).
For the J0 component, means were 0.
16 ± 0.
38D (PlusoptiX A16) and 0.
06 ± 0.
33D (Vision Screener V100) and for theJ45 component the means were 0.
03 ± 0.
17D (PlusoptiX A16) and 0.
06 ± 0.
22D (Vision Screener V100).
When compared both instruments, statistically significant differences were observed for the M (p=0.
017) and J0 (p=0.
004) components.
The agreement rates between PlusoptiX A16 and Vision Screener V100 across different refractive components were 80.
5% for sphere, 82.
0% for cylinder, and 40.
6% for axis when considering a range of ±0.
75 D for sphere and cylinder and ±25.
0 degrees for cylinder axis.
Simultaneously considering all three conditions, the overall agreement was 73.
7%.
ConclusionThe Vision Screener V100, while generally aligning well with PlusoptiX A16, tends to underestimate myopic spherical equivalent, overestimate hyperopic cases, and underestimate J0 astigmatism.

Related Results

Genomic epidemiology of coxsackievirus A16 in mainland of China, 2000–18
Genomic epidemiology of coxsackievirus A16 in mainland of China, 2000–18
Abstract Hand, foot, and mouth disease (HFMD), which is a frequently reported and concerning disease worldwide, is a severe burden on societies globally, especial...
Virtual flavonoids screening identifies potent compounds against enterovirus A71 and coxsackievirus A16
Virtual flavonoids screening identifies potent compounds against enterovirus A71 and coxsackievirus A16
Abstract Worldwide outbreaks of hand, foot, and mouth disease (HFMD) are caused by enterovirus A71 (EV-A71) and coxsackievirus A16 (CV-A16). Since no anti-HFMD drugs are cu...
Plusoptix photoscreener use for paediatric vision screening in Flanders and Iran
Plusoptix photoscreener use for paediatric vision screening in Flanders and Iran
AbstractPurposePhotoscreening assesses risk factors for amblyopia, as an alternative to measurement of visual acuity (VA) to detect amblyopia, on the premise that its early correct...
Validity of the Titmus Vision Screener: A Comparison with the Snellen Chart
Validity of the Titmus Vision Screener: A Comparison with the Snellen Chart
Given limited knowledge regarding validity of the Titmus vision screener, we sought to compare visual acuity measurements obtained from the Titmus with that from the Snellen chart ...
Validity of The Mediterranean Diet Screener
Validity of The Mediterranean Diet Screener
The Mediterranean Diet Intervention for Neurodegenerative Delay (MIND) screener, a 16‐item tool, has been developed to assess accordance to a Mediterranean‐style food pattern for A...
Can a Short Screening Tool Discriminate Between Overeating and Binge Eating in Treatment‐Seeking Individuals with Obesity?
Can a Short Screening Tool Discriminate Between Overeating and Binge Eating in Treatment‐Seeking Individuals with Obesity?
ObjectiveExisting screening tools are inadequate in differentiating binge eating from normative overeating in treatment‐seeking individuals with overweight or obesity, as these ind...
Lists, Spatial Practice and Assistive Technologies for the Blind
Lists, Spatial Practice and Assistive Technologies for the Blind
IntroductionSupermarkets are functionally challenging environments for people with vision impairments. A supermarket is likely to house an average of 45,000 products in a median fl...
Dosimetry study of three-dimensional print template for 125I implantation therapy
Dosimetry study of three-dimensional print template for 125I implantation therapy
Abstract Background 125I seed implantation has been found to show good therapeutic effects on tumors. Recent studies showed that three-dimensional (...

Back to Top