Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 Cp810 Sound Processor
View through CrossRef
The Nucleus® 5 System Sound Processor (CP810, Cochlear™, Macquarie University, NSW, Australia) contains two omnidirectional microphones. They can be configured as a fixed directional microphone combination (called Zoom) or as an adaptive beamformer (called Beam), which adjusts the directivity continuously to maximally reduce the interfering noise. Initial evaluation studies with the CP810 had compared performance and usability of the new processor in comparison with the Freedom™ Sound Processor (Cochlear™) for speech in quiet and noise for a subset of the processing options. This study compares the two processing options suggested to be used in noisy environments, Zoom and Beam, for various sound field conditions using a standardized speech in noise matrix test (Oldenburg sentences test). Nine German-speaking subjects who previously had been using the Freedom speech processor and subsequently were upgraded to the CP810 device participated in this series of additional evaluation tests. The speech reception threshold (SRT for 50% speech intelligibility in noise) was determined using sentences presented via loudspeaker at 65 dB SPL in front of the listener and noise presented either via the same loudspeaker (S0N0) or at 90 degrees at either the ear with the sound processor (S0NCI+) or the opposite unaided ear (S0NCI-). The fourth noise condition consisted of three uncorrelated noise sources placed at 90, 180 and 270 degrees. The noise level was adjusted through an adaptive procedure to yield a signal to noise ratio where 50% of the words in the sentences were correctly understood. In spatially separated speech and noise conditions both Zoom and Beam could improve the SRT significantly. For single noise sources, either ipsilateral or contralateral to the cochlear implant sound processor, average improvements with Beam of 12.9 and 7.9 dB in SRT were found. The average SRT of –8 dB for Beam in the diffuse noise condition (uncorrelated noise from both sides and back) is truly remarkable and comparable to the performance of normal hearing listeners in the same test environment. The static directivity (Zoom) option in the diffuse noise condition still provides a significant benefit of 5.9 dB in comparison with the standard omnidirectional microphone setting. These results indicate that CI recipients may improve their speech recognition in noisy environments significantly using these directional microphone-processing options.
Title: Speech Intelligibility in Various Noise Conditions with the Nucleus® 5 Cp810 Sound Processor
Description:
The Nucleus® 5 System Sound Processor (CP810, Cochlear™, Macquarie University, NSW, Australia) contains two omnidirectional microphones.
They can be configured as a fixed directional microphone combination (called Zoom) or as an adaptive beamformer (called Beam), which adjusts the directivity continuously to maximally reduce the interfering noise.
Initial evaluation studies with the CP810 had compared performance and usability of the new processor in comparison with the Freedom™ Sound Processor (Cochlear™) for speech in quiet and noise for a subset of the processing options.
This study compares the two processing options suggested to be used in noisy environments, Zoom and Beam, for various sound field conditions using a standardized speech in noise matrix test (Oldenburg sentences test).
Nine German-speaking subjects who previously had been using the Freedom speech processor and subsequently were upgraded to the CP810 device participated in this series of additional evaluation tests.
The speech reception threshold (SRT for 50% speech intelligibility in noise) was determined using sentences presented via loudspeaker at 65 dB SPL in front of the listener and noise presented either via the same loudspeaker (S0N0) or at 90 degrees at either the ear with the sound processor (S0NCI+) or the opposite unaided ear (S0NCI-).
The fourth noise condition consisted of three uncorrelated noise sources placed at 90, 180 and 270 degrees.
The noise level was adjusted through an adaptive procedure to yield a signal to noise ratio where 50% of the words in the sentences were correctly understood.
In spatially separated speech and noise conditions both Zoom and Beam could improve the SRT significantly.
For single noise sources, either ipsilateral or contralateral to the cochlear implant sound processor, average improvements with Beam of 12.
9 and 7.
9 dB in SRT were found.
The average SRT of –8 dB for Beam in the diffuse noise condition (uncorrelated noise from both sides and back) is truly remarkable and comparable to the performance of normal hearing listeners in the same test environment.
The static directivity (Zoom) option in the diffuse noise condition still provides a significant benefit of 5.
9 dB in comparison with the standard omnidirectional microphone setting.
These results indicate that CI recipients may improve their speech recognition in noisy environments significantly using these directional microphone-processing options.
Related Results
Prediction of Speech Intelligibility by Means of EEG Responses to Sentences in Noise
Prediction of Speech Intelligibility by Means of EEG Responses to Sentences in Noise
ObjectiveUnderstanding speech in noisy conditions is challenging even for people with mild hearing loss, and intelligibility for an individual person is usually evaluated by using ...
Environmental History of Oceanic Noise Pollution
Environmental History of Oceanic Noise Pollution
The concept of “ocean noise” precedes the concept of “ocean noise pollution” by about half a century. Those seeking a body of scholarly literature on ocean noise as an environmenta...
Pola Komunikasi Interpersonal Terapis Wicara pada Anak Telambat Bicara (Speech Delay)
Pola Komunikasi Interpersonal Terapis Wicara pada Anak Telambat Bicara (Speech Delay)
Abstract. The tittle of this study is Interpersonal Communication Patterns of Speech Therapists in Children with Speech Delay. Children with speech delay disorders are also social ...
Comprehension of acoustically degraded speech in Alzheimer’s disease and primary progressive aphasia
Comprehension of acoustically degraded speech in Alzheimer’s disease and primary progressive aphasia
ABSTRACTSuccessful communication in daily life frequently depends on accurate decoding of speech signals that are acoustically degraded by challenging listening conditions. This pr...
Single-channel speech enhancement by PSO-GSA with harmonic regeneration noise reduction
Single-channel speech enhancement by PSO-GSA with harmonic regeneration noise reduction
Speech quality significantly affects the performance of speech dependent systems. Noise in the background lowers the clarity and intelligibility of speech. The augmentation of spee...
Speech, communication, and neuroimaging in Parkinson's disease : characterisation and intervention outcomes
Speech, communication, and neuroimaging in Parkinson's disease : characterisation and intervention outcomes
<p dir="ltr">Most individuals with Parkinson's disease (PD) experience changes in speech, voice or communication. Speech changes often manifest as hypokinetic dysarthria, a m...
Speech, communication, and neuroimaging in Parkinson's disease : characterisation and intervention outcomes
Speech, communication, and neuroimaging in Parkinson's disease : characterisation and intervention outcomes
<p dir="ltr">Most individuals with Parkinson's disease (PD) experience changes in speech, voice or communication. Speech changes often manifest as hypokinetic dysarthria, a m...
Speech, communication, and neuroimaging in Parkinson's disease : Characterisation and intervention outcomes
Speech, communication, and neuroimaging in Parkinson's disease : Characterisation and intervention outcomes
<p dir="ltr">Most individuals with Parkinson's disease (PD) experience changes in speech, voice or communication. Speech changes often manifest as hypokinetic dysarthria, a m...