Abstract Su403: Iterative Simulation to Optimize Mechanical CPR Device Application

View through CrossRef
Mechanical chest compression devices (MCDs) are commonly used for in-hospital and emergency department (ED) cardiac arrest resuscitation. Manufacturer guidelines and educational materials demonstrate that in the prehospital setting, MCDs can be deployed with an interruption in CPR of less than 15 seconds. No similar materials exist for the in-hospital environment, where observational studies have noted median CPR interruptions of up to 40 seconds for MCD deployment. With the integration of extracorporeal CPR (eCPR) into OHCA systems of care, the need to transfer patients with ongoing CPR to MCDs in the ED is growing. To develop a hospital-specific procedure for receiving a patient in cardiac arrest and deploying the MCD with minimal no-flow time. We used an iterative participatory design process to develop a procedure for receiving patients in cardiac arrest from Emergency Medical Services (EMS). Tasks included transferring a patient, replacing defibrillator pads, and deploying the hospital MCD with minimal no-flow time. Participants including ED physicians, nurses, and respiratory therapists performed a series of simulations using a high-fidelity manikin, resuscitation equipment, and an MCD. Simulations were followed by focused debriefings, participant-directed modification of the working procedure, and repeat simulation until participants expressed satisfaction with the procedure. Participant roles were shuffled after each round. The primary outcome was the duration of CPR interruption required to deploy the MCD; change in this time over successive iterations was assessed using a Wilcoxon signed-rank test. The final procedure described a sequence of steps and responsibilities that participants deemed efficient and reproducible. Through seven simulation-debriefing cycles, CPR interruption associated with MCD deployment decreased from 19 to 10 seconds (p = 0.033). The most common failure modes that participants corrected during modifications involved needed equipment that was out of reach, obstructed provider positioning, and poorly defined roles. On subsequent days, two additional groups of on-shift ED providers deployed the MCD in less than 15 seconds following the developed procedure. Using an iterative design process, we created an MCD deployment procedure that minimized CPR interruptions and was acceptable to hospital staff. This represents an effective, low-cost methodology to design a process for transfer of care during CPR and MCD deployment.
Title: Abstract Su403: Iterative Simulation to Optimize Mechanical CPR Device Application
Description:
Mechanical chest compression devices (MCDs) are commonly used for in-hospital and emergency department (ED) cardiac arrest resuscitation.
Manufacturer guidelines and educational materials demonstrate that in the prehospital setting, MCDs can be deployed with an interruption in CPR of less than 15 seconds.
No similar materials exist for the in-hospital environment, where observational studies have noted median CPR interruptions of up to 40 seconds for MCD deployment.
With the integration of extracorporeal CPR (eCPR) into OHCA systems of care, the need to transfer patients with ongoing CPR to MCDs in the ED is growing.
To develop a hospital-specific procedure for receiving a patient in cardiac arrest and deploying the MCD with minimal no-flow time.
We used an iterative participatory design process to develop a procedure for receiving patients in cardiac arrest from Emergency Medical Services (EMS).
Tasks included transferring a patient, replacing defibrillator pads, and deploying the hospital MCD with minimal no-flow time.
Participants including ED physicians, nurses, and respiratory therapists performed a series of simulations using a high-fidelity manikin, resuscitation equipment, and an MCD.
Simulations were followed by focused debriefings, participant-directed modification of the working procedure, and repeat simulation until participants expressed satisfaction with the procedure.
Participant roles were shuffled after each round.
The primary outcome was the duration of CPR interruption required to deploy the MCD; change in this time over successive iterations was assessed using a Wilcoxon signed-rank test.
The final procedure described a sequence of steps and responsibilities that participants deemed efficient and reproducible.
Through seven simulation-debriefing cycles, CPR interruption associated with MCD deployment decreased from 19 to 10 seconds (p = 0.
033).
The most common failure modes that participants corrected during modifications involved needed equipment that was out of reach, obstructed provider positioning, and poorly defined roles.
On subsequent days, two additional groups of on-shift ED providers deployed the MCD in less than 15 seconds following the developed procedure.
Using an iterative design process, we created an MCD deployment procedure that minimized CPR interruptions and was acceptable to hospital staff.
This represents an effective, low-cost methodology to design a process for transfer of care during CPR and MCD deployment.

Related Results

CONTINUOUS COMPRESSION WITHOUT DEFIBRILLATION FAVOURED NO SHORT-TERM SURVIVAL IN PROLONGED VENTRICULAR FIBRILLATION
CONTINUOUS COMPRESSION WITHOUT DEFIBRILLATION FAVOURED NO SHORT-TERM SURVIVAL IN PROLONGED VENTRICULAR FIBRILLATION
Objectives Aims: During the 2005 American Heart Association (AHA) Consensus Conference, compression first versus defibrillation first for sudden cardiac arrest wi...
Common Pool Resources
Common Pool Resources
Common pool resources (CPRs) are characterized as resources for which the exclusion of users is difficult (referred to as excludability), and the use of such a resource by one user...
ONE-SHOCK VERSUS CONTINUOUS DEFIBRILLATION IN AN 8-MIN VENTRICULAR FIBRILLATION CANINE MODEL OF CARDIAC ARREST
ONE-SHOCK VERSUS CONTINUOUS DEFIBRILLATION IN AN 8-MIN VENTRICULAR FIBRILLATION CANINE MODEL OF CARDIAC ARREST
Objectives To investigate the resuscitation effect of a one-shock defibrillation protocol versus conventional continuous defibrillation with treatment variation i...
Extracorporeal resuscitation with carbon monoxide improves renal function by targeting inflammatory pathways in cardiac arrest in pigs
Extracorporeal resuscitation with carbon monoxide improves renal function by targeting inflammatory pathways in cardiac arrest in pigs
Deleterious consequences like acute kidney injury frequently occur upon successful resuscitation from cardiac arrest. Extracorporeal life support is increasingly used to overcome h...
Public perception and willingness towards bystander cardiopulmonary resuscitation (CPR) training and performance in Pakistan
Public perception and willingness towards bystander cardiopulmonary resuscitation (CPR) training and performance in Pakistan
Background/Aim: Bystander cardiopulmonary resuscitation (CPR) during out of-hospital cardiac arrest increases both survival rates and neurological recovery, but in Pakistan, an ala...
Nurse Knowledge About High Quality Cardiopulmonary Resuscitation (CPR)
Nurse Knowledge About High Quality Cardiopulmonary Resuscitation (CPR)
High quality CPR is administering CPR with the standarts already specified by the American Heart Association (AHA) 2015 which consists of 5 main components. Based on the data obtai...

Back to Top