Delivery accuracy of VMAT on two beam‐matched linacs provided by accelerated go live service

View through CrossRef
AbstractIntroductionDosimetric accuracy is critical when a patient treated with volumetric modulated arc therapy (VMAT) is transferred to another beam‐matched linac. To evaluate the performance of Accelerated Go Live (AGL) service, the measured beam characteristics and patient specific quality assurance (QA) results between two AGL‐matched linacs were compared.Materials and MethodsTwo VersaHD linacs were installed using the AGL service. After the installation, the beam data such as percentage depth dose (PDD), lateral profiles and output factors for all photon beams were measured. Relative doses were also measured as a function of the multi‐leaf collimator (MLC) leaf gap width. Subsequently, VMAT plans were created for prostate, pelvis, head and neck, liver, lung cancers and multiple brain metastases. Dose distributions and point doses were measured by multi‐dimensional detectors and ionization chambers for patient specific quality assurance, and comparisons were made between the two linacs.ResultsDose differences in PDDs were all within ± 1% except the entrance region, and the averaged gamma indices of the lateral profiles were within 0.3. The differences in doses as a function of the MLC leaf gap width between the two linacs were within ±0.5%. For all the plans, gamma passing rates were all higher than 95% with criteria of 2%/2 mm. The average and the SD of dose differences on the multi‐dimensional detector between both measurements was 0.06 ± 2.12%, and the average of point dose differences was −0.03 ± 0.33%.ConclusionWe have evaluated the AGL performance in the context of beam characteristics and patient specific QA. It was demonstrated that the AGL service provides an accurate VMAT treatment reproducibility for many tumor sites with gamma pass rates greater than 95% under criteria of 2%/2 mm.
Title: Delivery accuracy of VMAT on two beam‐matched linacs provided by accelerated go live service
Description:
AbstractIntroductionDosimetric accuracy is critical when a patient treated with volumetric modulated arc therapy (VMAT) is transferred to another beam‐matched linac.
To evaluate the performance of Accelerated Go Live (AGL) service, the measured beam characteristics and patient specific quality assurance (QA) results between two AGL‐matched linacs were compared.
Materials and MethodsTwo VersaHD linacs were installed using the AGL service.
After the installation, the beam data such as percentage depth dose (PDD), lateral profiles and output factors for all photon beams were measured.
Relative doses were also measured as a function of the multi‐leaf collimator (MLC) leaf gap width.
Subsequently, VMAT plans were created for prostate, pelvis, head and neck, liver, lung cancers and multiple brain metastases.
Dose distributions and point doses were measured by multi‐dimensional detectors and ionization chambers for patient specific quality assurance, and comparisons were made between the two linacs.
ResultsDose differences in PDDs were all within ± 1% except the entrance region, and the averaged gamma indices of the lateral profiles were within 0.
3.
The differences in doses as a function of the MLC leaf gap width between the two linacs were within ±0.
5%.
For all the plans, gamma passing rates were all higher than 95% with criteria of 2%/2 mm.
The average and the SD of dose differences on the multi‐dimensional detector between both measurements was 0.
06 ± 2.
12%, and the average of point dose differences was −0.
03 ± 0.
33%.
ConclusionWe have evaluated the AGL performance in the context of beam characteristics and patient specific QA.
It was demonstrated that the AGL service provides an accurate VMAT treatment reproducibility for many tumor sites with gamma pass rates greater than 95% under criteria of 2%/2 mm.

Related Results

Modified VMAT Plans for Locally Advanced Centrally Located Non-Small Cell Lung Cancer (NSCLC)
Modified VMAT Plans for Locally Advanced Centrally Located Non-Small Cell Lung Cancer (NSCLC)
Objectives: This study aimed to find the optimal radiotherapy VMAT plans, that achieved high conformity and homogeneity to the planned target volume (PTV), and minimize the dose to...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Strategies for reducing ovarian dose in volumetric modulated arc therapy (VMAT) for postoperative uterine cervical cancer
Strategies for reducing ovarian dose in volumetric modulated arc therapy (VMAT) for postoperative uterine cervical cancer
Objective: To reduce the ovarian dose with volumetric modulated arc therapy (VMAT), an original VMAT was designed with two types of arcs to restrict angles and fi...
COMPARISON OF COPLANAR AND NON- COPLANARVMAT FOR BRAIN CANCER BY USING THE DOSIMETRICAL AND RADIOBIOLOGICAL INDICES
COMPARISON OF COPLANAR AND NON- COPLANARVMAT FOR BRAIN CANCER BY USING THE DOSIMETRICAL AND RADIOBIOLOGICAL INDICES
Background: Recent techniques of radiotherapy such as volumetric modulated arc therapy (VMAT) that delivered in Coplanar technique or non-coplanar technique allows to deliver high ...
Feasibility of 4D VMAT-CT
Feasibility of 4D VMAT-CT
Abstract Objective. Feasibility of three-dimensional (3D) tracking of volumetric modulated arc therapy (VMAT) based on VMAT–computed tomography (V...

Back to Top