Semi-automatic puncture robotic system based on real-time multi-modal image fusion: preclinical evaluation
View through CrossRef
Abstract
Purpose
Traditional surgical robot system relying on computed tomography (CT) navigation suffers from two drawbacks during abdominal organ puncture surgeries. Firstly, the puncture target is displaced under the influence of respiration, thereby reducing the puncture accuracy. Secondly, the puncture process lacks real-time visualization, which may potentially give rise to medical accidents. This paper presents a semi-automatic surgical robot system based on the fusion guidance of real-time ultrasound images and CT, along with the monitoring of the patient’s respiratory state, to address these issues.
Method
This system utilizes a six-axis force sensor in contact with the human body, and a respiratory model can be constructed through data got from force sensor to monitor the patient’s respiratory phase and recommend the optimal puncture phase. The issue of non-real-time puncture guidance is addressed through the real-time registration and fusion of ultrasound (US) images with preoperative CT images.
Results
Phantom experiments and animal experiments were carried out based on this design. The test results indicate that in these two experiments, the average fusion error between US and CT of the main tissues in the liver is within 3 mm. For puncture accuracy, in the phantom experiment, the average puncture error was 1.0 mm, with a minimum of 0 mm and a maximum of 2.1 mm. In the animal experiment, the average puncture error was 2.5 mm, ranging from a minimum of 1.6 mm to a maximum of 3.0 mm.
Conclusion
The results of two experiments show both the image fusion accuracy and puncture accuracy of this system are within 3mm, which can meet the requirement of 5 mm puncture accuracy in clinical practice. Approximately 70% of the operation are automatically accomplished by robot system, greatly reducing the reliance on the doctor’s experience.
Springer Science and Business Media LLC
Title: Semi-automatic puncture robotic system based on real-time multi-modal image fusion: preclinical evaluation
Description:
Abstract
Purpose
Traditional surgical robot system relying on computed tomography (CT) navigation suffers from two drawbacks during abdominal organ puncture surgeries.
Firstly, the puncture target is displaced under the influence of respiration, thereby reducing the puncture accuracy.
Secondly, the puncture process lacks real-time visualization, which may potentially give rise to medical accidents.
This paper presents a semi-automatic surgical robot system based on the fusion guidance of real-time ultrasound images and CT, along with the monitoring of the patient’s respiratory state, to address these issues.
Method
This system utilizes a six-axis force sensor in contact with the human body, and a respiratory model can be constructed through data got from force sensor to monitor the patient’s respiratory phase and recommend the optimal puncture phase.
The issue of non-real-time puncture guidance is addressed through the real-time registration and fusion of ultrasound (US) images with preoperative CT images.
Results
Phantom experiments and animal experiments were carried out based on this design.
The test results indicate that in these two experiments, the average fusion error between US and CT of the main tissues in the liver is within 3 mm.
For puncture accuracy, in the phantom experiment, the average puncture error was 1.
0 mm, with a minimum of 0 mm and a maximum of 2.
1 mm.
In the animal experiment, the average puncture error was 2.
5 mm, ranging from a minimum of 1.
6 mm to a maximum of 3.
0 mm.
Conclusion
The results of two experiments show both the image fusion accuracy and puncture accuracy of this system are within 3mm, which can meet the requirement of 5 mm puncture accuracy in clinical practice.
Approximately 70% of the operation are automatically accomplished by robot system, greatly reducing the reliance on the doctor’s experience.
Related Results
The Nuclear Fusion Award
The Nuclear Fusion Award
The Nuclear Fusion Award ceremony for 2009 and 2010 award winners was held during the 23rd IAEA Fusion Energy Conference in Daejeon. This time, both 2009 and 2010 award winners w...
Modified Ablation Catheter Guided Transseptal Puncture under ICE Guidance
Modified Ablation Catheter Guided Transseptal Puncture under ICE Guidance
Abstract
Background
Zero-fluoroscopy procedures have become increasingly popular in electrophysiological interventional surgery. As the key technology to achieve zero-fluo...
A 3D Printing Personalized Percutaneous Nephroscope Puncture Guide Plate in Percutaneous Nephrolithotomy: A Pilot Study
A 3D Printing Personalized Percutaneous Nephroscope Puncture Guide Plate in Percutaneous Nephrolithotomy: A Pilot Study
Abstract
Objective Using CTU, combined with 3D printing technology, digital design, the development of individualized PCNL) puncture guides, preliminary discussion of its f...
Incidence and risk factors of Postdural Puncture Headache: Prospective cohort study design
Incidence and risk factors of Postdural Puncture Headache: Prospective cohort study design
Abstract
Background:Postdural puncture headache is one of the complications following spinal anaesthesia and accidental dural puncture. Several modifiable risk factors cont...
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct
Introduction
Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
Nonproliferation and fusion power plants
Nonproliferation and fusion power plants
Abstract
The world now appears to be on the brink of realizing commercial fusion. As fusion energy progresses towards near-term commercial deployment, the question arises a...
Incidence and risk factors of Postdural Puncture Headache: Prospective cohort study design
Incidence and risk factors of Postdural Puncture Headache: Prospective cohort study design
Abstract
Background: Post-dural puncture headache is one of the complications following spinal anaesthesia and accidental dural puncture. Several modifiable risk factors co...
Multi-perspective, Multi-modal Image Registration and Fusion
Multi-perspective, Multi-modal Image Registration and Fusion
Multi-modal image fusion is an active research area with many civilian and military applications. Fusion is defined as strategic combination of information collected by various sen...