Two-Year Outcomes of Novel Dual-Mobility Implant in Primary Total Hip Arthroplasty
View through CrossRef
Introduction. Dual-mobility (DM) implants for total hip arthroplasty (THA) have gained popularity due to their potential to reduce hip instability and dislocation events that may lead to revision surgery. These implants consist of a femoral head articulated within a polyethylene liner, which articulates within an outer acetabular shell, creating a dual-bearing surface. Our study aimed to report our observations on the survivorship of a novel DM implant for primary total hip arthroplasty at two years. Methods. We conducted a retrospective, multicenter study to assess the clinical outcomes of patients undergoing a THA with a novel DM implant (OR3O acetabular system™, Smith & Nephew, Inc., Memphis, TN) from January 2020 to September 2021. Patient demographics, surgical information, and survivorship data were collected from medical records for patients with a minimum of two years of follow-up. Primary outcomes included overall implant survivorship at two years as well as aseptic survivorship, revision rates of the DM acetabular shell, and average time to revision. Patient-reported outcomes were collected in the form of HOOS JR. Results. A total of 250 hips in 245 patients had a minimum two-year follow-up. Primary osteoarthritis (80%) was the most common indication for index THA. The average aseptic survivorship of the DM acetabular components at two years for the cohort was 98.4% and survivorship of the acetabular implants overall was 97.6%. There were a total of four (1.6%) aseptic revisions of the DM acetabular component. Reasons for aseptic acetabular revision included one case of instability, one intraprosthetic dislocation, one periprosthetic acetabular fracture, and one malpositioned acetabular cup resulting in impingement. The mean time of follow-up was 893.9 days. Eighty-seven patients had preoperative and two-year HOOS JR available. HOOS JR improved by an average of 38.5 points. Conclusion. This novel DM acetabular implant demonstrates excellent survivorship at two years follow-up with low rates of instability and intraprosthetic dislocation and no episodes of metal-on-metal corrosion. Use of the DM implant demonstrated clinically relevant improvements in patient-reported outcomes at two years.
Title: Two-Year Outcomes of Novel Dual-Mobility Implant in Primary Total Hip Arthroplasty
Description:
Introduction.
Dual-mobility (DM) implants for total hip arthroplasty (THA) have gained popularity due to their potential to reduce hip instability and dislocation events that may lead to revision surgery.
These implants consist of a femoral head articulated within a polyethylene liner, which articulates within an outer acetabular shell, creating a dual-bearing surface.
Our study aimed to report our observations on the survivorship of a novel DM implant for primary total hip arthroplasty at two years.
Methods.
We conducted a retrospective, multicenter study to assess the clinical outcomes of patients undergoing a THA with a novel DM implant (OR3O acetabular system™, Smith & Nephew, Inc.
, Memphis, TN) from January 2020 to September 2021.
Patient demographics, surgical information, and survivorship data were collected from medical records for patients with a minimum of two years of follow-up.
Primary outcomes included overall implant survivorship at two years as well as aseptic survivorship, revision rates of the DM acetabular shell, and average time to revision.
Patient-reported outcomes were collected in the form of HOOS JR.
Results.
A total of 250 hips in 245 patients had a minimum two-year follow-up.
Primary osteoarthritis (80%) was the most common indication for index THA.
The average aseptic survivorship of the DM acetabular components at two years for the cohort was 98.
4% and survivorship of the acetabular implants overall was 97.
6%.
There were a total of four (1.
6%) aseptic revisions of the DM acetabular component.
Reasons for aseptic acetabular revision included one case of instability, one intraprosthetic dislocation, one periprosthetic acetabular fracture, and one malpositioned acetabular cup resulting in impingement.
The mean time of follow-up was 893.
9 days.
Eighty-seven patients had preoperative and two-year HOOS JR available.
HOOS JR improved by an average of 38.
5 points.
Conclusion.
This novel DM acetabular implant demonstrates excellent survivorship at two years follow-up with low rates of instability and intraprosthetic dislocation and no episodes of metal-on-metal corrosion.
Use of the DM implant demonstrated clinically relevant improvements in patient-reported outcomes at two years.
Related Results
Physiological and biomechanical factors contributing to the hip adduction angle in female runners
Physiological and biomechanical factors contributing to the hip adduction angle in female runners
Running is a popular form of exercise that is accompanied by many health benefits. However, running also comes with a risk of overuse injuries. Women have a higher risk for overuse...
Hip Stability Parameters with Dual Mobility, Modular Dual Mobility and Fixed Bearing in Total Hip Arthroplasty: an Analytical Evaluation.
Hip Stability Parameters with Dual Mobility, Modular Dual Mobility and Fixed Bearing in Total Hip Arthroplasty: an Analytical Evaluation.
Abstract
Background. Use of dual mobility in total hip arthroplasty has gained popularity due to the ability to reduce dislocation through increased jumping distance and im...
Comparison of functional outcome between Bipolar Hemiarthroplasty and total Hip Arthroplasty for displaced femoral neck fractures in elderly patients
Comparison of functional outcome between Bipolar Hemiarthroplasty and total Hip Arthroplasty for displaced femoral neck fractures in elderly patients
Objectives: The intention of this study to compare functional outcomes between Total Hip Arthroplasty and Bipolar Hemi Arthroplasty for displaced femoral neck fracture after 3 mont...
Revise and remember : managing hip resurfacing implant recall
Revise and remember : managing hip resurfacing implant recall
<h4>Background</h4><p dir="ltr">Total hip arthroplasty (THA) with metal-on-polyethylene articulation provides excellent clinical outcomes but remains associated w...
Revise and remember : managing hip resurfacing implant recall
Revise and remember : managing hip resurfacing implant recall
<h4>Background</h4><p dir="ltr">Total hip arthroplasty (THA) with metal-on-polyethylene articulation provides excellent clinical outcomes but remains associated w...
Mix En Meng It Op: Emile YX?'s Alternative Race and Language Politics in South African Hip-Hop
Mix En Meng It Op: Emile YX?'s Alternative Race and Language Politics in South African Hip-Hop
This paper explores South African hip-hop activist Emile YX?'s work to suggest that he presents an alternative take on mainstream US and South African hip-hop. While it is arguable...
Vitamin D and periprosthetic infection: Review of literature (Preprint)
Vitamin D and periprosthetic infection: Review of literature (Preprint)
BACKGROUND
Infections are rare events in arthroplasty surgeries, occurring in only 0.5% to 2% of procedures. It significantly impacts all involved parts (pa...
Association of Hip Radiograph Findings With Pain and Function in Patients Presenting With Low Back Pain
Association of Hip Radiograph Findings With Pain and Function in Patients Presenting With Low Back Pain
AbstractBackgroundRelationships between low back pain (LBP) and the hip in patient cohorts have been described primarily in patients with moderate to severe hip osteoarthritis (OA)...