Correlation Between Corneal Biomechanics and Anterior Segment Parameters in Healthy Saudi Females
View through CrossRef
Abstract
Background: The ocular response analyzer (ORA) can measure corneal hysteresis (CH), corneal resistance factor (CRF), Goldmann correlated IOP (IOPg) and corneal compensated intraocular pressure (IOPcc).1 Anterior segment parameters such as central corneal thickness (CCT), thinnest corneal thickness (TCT), apex corneal thickness (Apex CT), corneal volume (CV), anterior chamber depth (ACD), anterior chamber volume (ACV) and corneal astigmatism (CA) can be measured by Pentacam which is a Scheimpflug imaging device.2 Many studies 3-10 investigated the correlation between corneal biomechanics and anterior segment parameters in healthy eyes and demonstrated a strong correlation between CH, CRF with CCT. Hwang et al 2013 7 found that CV was positively correlated with CH, but not CRF. However, Çevik et al 2016 10 reported positive correlations between CH, CRF, and CV and negative correlations between CH, CRF and both of posterior steep and average posterior ???? values. Hwang et al 2 did not show a significant association between CA and all the biomechanical properties. Conversely, Montard et al 4 reported a negative association between CA with CH and CRF. Therefore, there is still debate regarding evaluation of this relation. Up to our knowledge, the correlation between corneal biomechanics and anterior segment parameters in healthy eyes is never investigated in Saudi Arabia.The purpose of this study was conducted to evaluate the correlation between corneal biomechanics measured with ORA and anterior segment parameters assessed with Oculus Pentacam HR in healthy Saudi females.Methods: This study was a prospective, non-randomized, cross-sectional, observational and quantitative study. The study included 129 eyes of 129 healthy Saudi females from King Saud University, Riyadh, Saudi Arabia. The mean age was 19.87 ± 1.328 (18 – 29 years). All subjects underwent a comprehensive ophthalmologic examination including refraction, visual acuity measurement, slit-lamp biomicroscopic examination, IOP measurement with an air puffer tonometer, and funduscopy. In addition, anterior segment parameters were measured with Oculus Pentacam HR. Additionally, corneal biomechanical parameters were measured with ORA (Reichert Ophthalmic Instruments). All data was analyzed using a Statistical Package for the Social Sciences (SPSS) version 22.0 software (SPSS Inc., Chicago, II, USA). Associations between corneal biomechanical parameters and anterior segment parameters were analyzed by Pearson’s Correlation coefficients. P < 0.05 was considered a statistically significant.Results: In this study, mean (±SD) spherical equivalent (SE) was -1.62 ± 2.15 diopters and mean (±SD) CCT was 552.41 ± 58.90 μm. Mean (±SD) CH and CRF were 11.61 ± 1.80 and 11.26 ± 1.99 mm Hg, respectively. Correlation between ORA parameters and the anterior segment parameters using Pearson’s Correlation Coefficient for all eyes in this study showed only highly significant positive correlation between CCT and each of CH, CRF, IOP g (r = 0.381, P < 0.0001) (r = 0.395, P < 0.0001) (r = 0.304, P < 0.0001) respectively. On the other hand, no significant association was detected between IOP cc and anterior segment parameters in this study Conclusion: This work is the first one in Saudi Arabia to evaluate the correlation between corneal biomechanics and anterior segment parameters in healthy Saudi females. This study reported a positive correlation between CCT and each of CH, CRF, IOPg. Mild myopic eyes in this study showed a positive association between ASKV and each of CH and CRF. In addition, the mild myopic eyes demonstrated a positive relation between IOP g and ACV. Future prospective studies including males, different ethnic populations, different age groups with large sample sizes, using different imaging techniques, are recommended.
Title: Correlation Between Corneal Biomechanics and Anterior Segment Parameters in Healthy Saudi Females
Description:
Abstract
Background: The ocular response analyzer (ORA) can measure corneal hysteresis (CH), corneal resistance factor (CRF), Goldmann correlated IOP (IOPg) and corneal compensated intraocular pressure (IOPcc).
1 Anterior segment parameters such as central corneal thickness (CCT), thinnest corneal thickness (TCT), apex corneal thickness (Apex CT), corneal volume (CV), anterior chamber depth (ACD), anterior chamber volume (ACV) and corneal astigmatism (CA) can be measured by Pentacam which is a Scheimpflug imaging device.
2 Many studies 3-10 investigated the correlation between corneal biomechanics and anterior segment parameters in healthy eyes and demonstrated a strong correlation between CH, CRF with CCT.
Hwang et al 2013 7 found that CV was positively correlated with CH, but not CRF.
However, Çevik et al 2016 10 reported positive correlations between CH, CRF, and CV and negative correlations between CH, CRF and both of posterior steep and average posterior ???? values.
Hwang et al 2 did not show a significant association between CA and all the biomechanical properties.
Conversely, Montard et al 4 reported a negative association between CA with CH and CRF.
Therefore, there is still debate regarding evaluation of this relation.
Up to our knowledge, the correlation between corneal biomechanics and anterior segment parameters in healthy eyes is never investigated in Saudi Arabia.
The purpose of this study was conducted to evaluate the correlation between corneal biomechanics measured with ORA and anterior segment parameters assessed with Oculus Pentacam HR in healthy Saudi females.
Methods: This study was a prospective, non-randomized, cross-sectional, observational and quantitative study.
The study included 129 eyes of 129 healthy Saudi females from King Saud University, Riyadh, Saudi Arabia.
The mean age was 19.
87 ± 1.
328 (18 – 29 years).
All subjects underwent a comprehensive ophthalmologic examination including refraction, visual acuity measurement, slit-lamp biomicroscopic examination, IOP measurement with an air puffer tonometer, and funduscopy.
In addition, anterior segment parameters were measured with Oculus Pentacam HR.
Additionally, corneal biomechanical parameters were measured with ORA (Reichert Ophthalmic Instruments).
All data was analyzed using a Statistical Package for the Social Sciences (SPSS) version 22.
0 software (SPSS Inc.
, Chicago, II, USA).
Associations between corneal biomechanical parameters and anterior segment parameters were analyzed by Pearson’s Correlation coefficients.
P < 0.
05 was considered a statistically significant.
Results: In this study, mean (±SD) spherical equivalent (SE) was -1.
62 ± 2.
15 diopters and mean (±SD) CCT was 552.
41 ± 58.
90 μm.
Mean (±SD) CH and CRF were 11.
61 ± 1.
80 and 11.
26 ± 1.
99 mm Hg, respectively.
Correlation between ORA parameters and the anterior segment parameters using Pearson’s Correlation Coefficient for all eyes in this study showed only highly significant positive correlation between CCT and each of CH, CRF, IOP g (r = 0.
381, P < 0.
0001) (r = 0.
395, P < 0.
0001) (r = 0.
304, P < 0.
0001) respectively.
On the other hand, no significant association was detected between IOP cc and anterior segment parameters in this study Conclusion: This work is the first one in Saudi Arabia to evaluate the correlation between corneal biomechanics and anterior segment parameters in healthy Saudi females.
This study reported a positive correlation between CCT and each of CH, CRF, IOPg.
Mild myopic eyes in this study showed a positive association between ASKV and each of CH and CRF.
In addition, the mild myopic eyes demonstrated a positive relation between IOP g and ACV.
Future prospective studies including males, different ethnic populations, different age groups with large sample sizes, using different imaging techniques, are recommended.
Related Results
Stability of anterior segments in patients with moderate and high myopia one year after SMILE
Stability of anterior segments in patients with moderate and high myopia one year after SMILE
Abstract
Background: SMILE is one of the most leading-edge corneal refractive surgery.In our study, we aim to investigate the stability of anterior segments in patients wit...
Evaluation of the effect of the corneal suture removal on corneal biomechanics with Corneal visualization Scheimpflug technology (Corvis ST) after penetrating keratoplasty (PKP)
Evaluation of the effect of the corneal suture removal on corneal biomechanics with Corneal visualization Scheimpflug technology (Corvis ST) after penetrating keratoplasty (PKP)
Background
The cornea is a part of the anterior segment of the eye, which plays an important role in the refraction of light rays on the retina. Cornea also maintains t...
Correlation and Comparison of Anterior and Posterior Corneal Surface Parameters in Healthy Eyes
Correlation and Comparison of Anterior and Posterior Corneal Surface Parameters in Healthy Eyes
Purpose: To correlate and compare anterior and posterior corneal surface parameters like Kflat, Ksteep, Kavg, R1, R2, Ravg and astigmatism in healthy eyes.
Study Design: Descriptiv...
A retrospective analysis of outcomes of an optical reconstructive surgery combined with corneal transplantation dedicated to manage complications of inflammatory diseases of the anterior chamber of the ey
A retrospective analysis of outcomes of an optical reconstructive surgery combined with corneal transplantation dedicated to manage complications of inflammatory diseases of the anterior chamber of the ey
Background: full thickness corneal transplantation or penetrating keratoplasty (PK) is the only surgical method to manage outcomes of
inflammatory diseases of the anterior eye cham...
Corneal densitometry changes in keratoconus patients after intrastromal corneal ring segments implantation
Corneal densitometry changes in keratoconus patients after intrastromal corneal ring segments implantation
BackgroundKeratoconus (KC) is an ectatic corneal disorder characterized by progressive corneal thinning with myopic astigmatism. Intrastromal corneal ring segments (ICRS) have been...
Surface modification treatments to enhance the performance of titanium backplate in keratoprosthesis
Surface modification treatments to enhance the performance of titanium backplate in keratoprosthesis
(English) The cornea is the outermost layer of the eye, essential for vision as it provides protection and contributes 80% of the eye’s refractive power. Its transparency and lack ...
Influence of Biometric and Corneal Tomographic Parameters on Normative Corneal Aberrations Measured by Root Mean Square
Influence of Biometric and Corneal Tomographic Parameters on Normative Corneal Aberrations Measured by Root Mean Square
Background/Objectives: To determine the impact of corneal and biometry parameters on the normative root mean square (RMS) values of corneal aberrations measured at a 6 mm diameter....
Repeatability and reproducibility of Keratograph 5M corneal topography
Repeatability and reproducibility of Keratograph 5M corneal topography
Background
Corneal topography is an important diagnostic tool and highly repeatable and reproducible topographic devices are essential in eye care practice. Placido ...