COMPARATIVE EVALUATION OF FLEXURAL STRENGTH OF TWO STAINLESS STEEL PRE-FORMED SPACE MAINTAINERS- AN IN VITRO STUDY
View through CrossRef
Background:
Premature loss of primary teeth often results in space discrepancies and malocclusion, making space maintainers an
essential component of pediatric dentistry. Preformed stainless steel space maintainers have gained popularity due to
their ease of use, durability, and reproducibility. However, the clinical performance of these appliances is highly
dependent on their flexural strength, which determines their resistance to deformation under masticatory forces.
Aim:
The present study aimed to compare the flexural strength of two commercially available preformed stainless steel
space maintainers—Kedo-S Fit and Kids E Dental—using an Instron universal testing machine.
Materials and Methods:
A total of 20 preformed stainless steel space maintainers were tested, with 10 samples allocated to each group: Group
1 – Kedo-S Fit and Group 2 – Kids E Dental. The samples were subjected to a cantilever flexural bending test on the
Instron machine at a crosshead speed of 1 mm/min, with the point of load application set at 10 mm. The maximum
force required to cause failure was recorded. Data were analyzed using an independent t-test, with the level of
significance set at p < 0.05.
Results:
The mean flexural strength of Kedo-S Fit was 48.6 ± 3.2 N, while Kids E Dental demonstrated a mean of 46.9 ± 3.5
N. Statistical analysis revealed no significant difference between the groups (t = 0.92, p = 0.37). However, Kedo-S Fit
exhibited marginally higher values.
Conclusion:
Both systems showed adequate flexural strength, confirming their clinical reliability. Kedo-S Fit demonstrated slightly
superior mechanical resistance, which may provide an additional margin of durability in functional conditions.
Dr. Yashwant Research Labs Pvt. Ltd.
Title: COMPARATIVE EVALUATION OF FLEXURAL STRENGTH OF TWO STAINLESS STEEL PRE-FORMED SPACE MAINTAINERS- AN IN VITRO STUDY
Description:
Background:
Premature loss of primary teeth often results in space discrepancies and malocclusion, making space maintainers an
essential component of pediatric dentistry.
Preformed stainless steel space maintainers have gained popularity due to
their ease of use, durability, and reproducibility.
However, the clinical performance of these appliances is highly
dependent on their flexural strength, which determines their resistance to deformation under masticatory forces.
Aim:
The present study aimed to compare the flexural strength of two commercially available preformed stainless steel
space maintainers—Kedo-S Fit and Kids E Dental—using an Instron universal testing machine.
Materials and Methods:
A total of 20 preformed stainless steel space maintainers were tested, with 10 samples allocated to each group: Group
1 – Kedo-S Fit and Group 2 – Kids E Dental.
The samples were subjected to a cantilever flexural bending test on the
Instron machine at a crosshead speed of 1 mm/min, with the point of load application set at 10 mm.
The maximum
force required to cause failure was recorded.
Data were analyzed using an independent t-test, with the level of
significance set at p < 0.
05.
Results:
The mean flexural strength of Kedo-S Fit was 48.
6 ± 3.
2 N, while Kids E Dental demonstrated a mean of 46.
9 ± 3.
5
N.
Statistical analysis revealed no significant difference between the groups (t = 0.
92, p = 0.
37).
However, Kedo-S Fit
exhibited marginally higher values.
Conclusion:
Both systems showed adequate flexural strength, confirming their clinical reliability.
Kedo-S Fit demonstrated slightly
superior mechanical resistance, which may provide an additional margin of durability in functional conditions.
Related Results
Aportaciones al estudio del comportamiento a flexión de estructuras de acero inoxidable
Aportaciones al estudio del comportamiento a flexión de estructuras de acero inoxidable
L'acer inoxidable està essent utilitzat de manera creixent en els últims anys als sectors de la indústria i de l'arquitectura gràcies a la seva resistència a la corrosió, facilitat...
On the Sustainability of Deep Learning Projects: Maintainers’ Perspective
On the Sustainability of Deep Learning Projects: Maintainers’ Perspective
Deep learning (DL) techniques have grown in leaps and bounds in
both academia and industry over the past few years. Despite the growth
of DL projects, there has been little study o...
PATIENT- AND PARENT-CENTERED EVALUATION OF SATISFACTION WITH FIBER-REINFORCED COMPOSITE VERSUS CONVENTIONAL BAND-AND-LOOP SPACE MAINTAINERS IN PEDIATRIC DENTISTRY
PATIENT- AND PARENT-CENTERED EVALUATION OF SATISFACTION WITH FIBER-REINFORCED COMPOSITE VERSUS CONVENTIONAL BAND-AND-LOOP SPACE MAINTAINERS IN PEDIATRIC DENTISTRY
Fixed space maintainers for prematurely lost primary molars aim not only to preserve arch integrity but also to ensure patient comfort and satisfaction. The most widely used Band a...
Primerjalna književnost na prelomu tisočletja
Primerjalna književnost na prelomu tisočletja
In a comprehensive and at times critical manner, this volume seeks to shed light on the development of events in Western (i.e., European and North American) comparative literature ...
Clad Steel Pipe for Corrosive Gas Transportation
Clad Steel Pipe for Corrosive Gas Transportation
ABSTRACT
This paper describes the applicability and reliability Of clad steel pipe and its welds in sour gas environments in comparison with those of 22%Cr-5.5%Ni...
A Comparative Evaluation of Frictional Resistance of Conventional, Teflon and Epoxy Coated Stainless Steel Archwires in Metal, Ceramic Brackets – An In vitro Study
A Comparative Evaluation of Frictional Resistance of Conventional, Teflon and Epoxy Coated Stainless Steel Archwires in Metal, Ceramic Brackets – An In vitro Study
Aim and Objectives: To evaluate the frictional resistance of Conventional, Teflon, and Epoxy coated stainless steel archwires in Metal, Ceramic brackets.
Materials and Method...
Effect Of Methyl Formate-Methyl Acetate Treatment On Flexural Strength Between Denture Base And Self-Cured Hard Reline Materials
Effect Of Methyl Formate-Methyl Acetate Treatment On Flexural Strength Between Denture Base And Self-Cured Hard Reline Materials
The purpose of this study was to evaluate the effect of methyl formate-methyl acetate (MF-MA) surface treatment on the flexural strength between denture base and hard reline materi...
Comparative Evaluation of Flexural Strength and Flexural Modulus of Different Periodontal Splint Materials: An In Vitro Study
Comparative Evaluation of Flexural Strength and Flexural Modulus of Different Periodontal Splint Materials: An In Vitro Study
Splinting of the mobile teeth is a critical part of periodontal management to improve the prognosis and longevity of stable results of periodontally compromised teeth with increase...