Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase

View through CrossRef
On average, arterial oxygen saturation measured by pulse oximetry (SpO2) is higher in hypoxemia than the true oxygen saturation measured invasively (SaO2), thereby increasing the risk of occult hypoxemia. In the current article, measurements of SpO2 on 17 cyanotic newborns were performed by means of a Nellcor pulse oximeter (POx), based on light with two wavelengths in the red and infrared regions (660 and 900 nm), and by means of a novel POx, based on two wavelengths in the infrared region (761 and 820 nm). The SpO2 readings from the two POxs showed higher values than the invasive SaO2 readings, and the disparity increased with decreasing SaO2. SpO2 measured using the two infrared wavelengths showed better correlation with SaO2 than SpO2 measured using the red and infrared wavelengths. After appropriate calibration, the standard deviation of the individual SpO2−SaO2 differences for the two-infrared POx was smaller (3.6%) than that for the red and infrared POx (6.5%, p < 0.05). The overestimation of SpO2 readings in hypoxemia was explained by the increase in hypoxemia of the optical pathlengths-ratio between the two wavelengths. The two-infrared POx can reduce the overestimation of SpO2 measurement in hypoxemia and the consequent risk of occult hypoxemia, owing to its smaller increase in pathlengths-ratio in hypoxemia.
Title: Overestimation of Oxygen Saturation Measured by Pulse Oximetry in Hypoxemia. Part 1: Effect of Optical Pathlengths-Ratio Increase
Description:
On average, arterial oxygen saturation measured by pulse oximetry (SpO2) is higher in hypoxemia than the true oxygen saturation measured invasively (SaO2), thereby increasing the risk of occult hypoxemia.
In the current article, measurements of SpO2 on 17 cyanotic newborns were performed by means of a Nellcor pulse oximeter (POx), based on light with two wavelengths in the red and infrared regions (660 and 900 nm), and by means of a novel POx, based on two wavelengths in the infrared region (761 and 820 nm).
The SpO2 readings from the two POxs showed higher values than the invasive SaO2 readings, and the disparity increased with decreasing SaO2.
SpO2 measured using the two infrared wavelengths showed better correlation with SaO2 than SpO2 measured using the red and infrared wavelengths.
After appropriate calibration, the standard deviation of the individual SpO2−SaO2 differences for the two-infrared POx was smaller (3.
6%) than that for the red and infrared POx (6.
5%, p < 0.
05).
The overestimation of SpO2 readings in hypoxemia was explained by the increase in hypoxemia of the optical pathlengths-ratio between the two wavelengths.
The two-infrared POx can reduce the overestimation of SpO2 measurement in hypoxemia and the consequent risk of occult hypoxemia, owing to its smaller increase in pathlengths-ratio in hypoxemia.

Related Results

High Concentration Oxygen and Hypercapnia in Respiratory Disease
High Concentration Oxygen and Hypercapnia in Respiratory Disease
<p>Oxygen-induced elevations in arterial carbon dioxide tension have been demonstrated in patients with chronic obstructive pulmonary disease (COPD), asthma, pneumonia, obesi...
Closed-Loop Oxygen Control
Closed-Loop Oxygen Control
<p>Guidelines recommend that oxygen should be titrated to achieve a target oxygen saturation (SpO2 ) range in acutely unwell patients, a concept colloquially known as “swimmi...
A Randomized Controlled Trial of Naloxone for Optimization of Hypoxemia in Lung Donors After Brain Death
A Randomized Controlled Trial of Naloxone for Optimization of Hypoxemia in Lung Donors After Brain Death
Background. Persistent hypoxemia is the principal reason lungs from otherwise eligible brain dead (BD) organ donors are not transplanted. Experimental models and retros...
Understanding pulse oximetry in hematology patients: Hemoglobinopathies, racial differences, and beyond
Understanding pulse oximetry in hematology patients: Hemoglobinopathies, racial differences, and beyond
AbstractPulse oximetry (SpO2) is a widely used, non‐invasive method of estimating arterial oxygen saturation. Measurement of SpO2 relies on comparing the relative absorption of lig...
Deep spectral learning for label-free optical imaging oximetry with uncertainty quantification
Deep spectral learning for label-free optical imaging oximetry with uncertainty quantification
AbstractMeasurement of blood oxygen saturation (sO2) by optical imaging oximetry provides invaluable insight into local tissue functions and metabolism. Despite different embodimen...
Step oximetry test: a validation study
Step oximetry test: a validation study
Introduction Step climbing is a potentially useful modality for testing exercise capacity. However, there are significant variations between test protocols and la...

Back to Top