Pullout strength of monocortical and bicortical screws in meta -physeal and diaphyseal regions of the canine humerus
View through CrossRef
Summary
Objective: Monocortical screws are commonly employed in locking plate fixation, but specific recommendations for their placement are lacking and use of short monocortical screws in metaphyseal bone may be contra indicated. Objectives of this study were to evaluate axial pullout strength of two different lengths of monocortical screws placed in various regions of the canine humerus compared to bicortical screws, and to derive cortical thickness and bone density values for those regions using quantitative computed tomography analysis (QCT).
Methods: The QCT analysis was performed on 36 cadaveric canine humeri for six regions of interest (ROI). A bicortical, short monocortical, or 50% transcortical 3.5 mm screw was implanted in each ROI and axial pullout testing was performed.
Results: Bicortical screws were stronger than monocortical screws in all ROI except the lateral epicondylar crest. Short monocortical metaphyseal screws were weaker than those placed in other regions. The 50% transcortical screws were stronger than the short monocortical screws in the condyle. A linear relationship between screw length and pull-out strength was observed.
Clinical significance: Cortical thickness and bone density measurements were obtained from multiple regions of the canine humerus using QCT. Use of short monocortical screws may contribute to failure of locking plate fixation of humeral fractures, especially when placed in the condyle. When bicortical screw placement is not possible, maximizing monocortical screw length may optimize fixation stability for distal humeral fractures.
Title: Pullout strength of monocortical and bicortical screws in meta -physeal and diaphyseal regions of the canine humerus
Description:
Summary
Objective: Monocortical screws are commonly employed in locking plate fixation, but specific recommendations for their placement are lacking and use of short monocortical screws in metaphyseal bone may be contra indicated.
Objectives of this study were to evaluate axial pullout strength of two different lengths of monocortical screws placed in various regions of the canine humerus compared to bicortical screws, and to derive cortical thickness and bone density values for those regions using quantitative computed tomography analysis (QCT).
Methods: The QCT analysis was performed on 36 cadaveric canine humeri for six regions of interest (ROI).
A bicortical, short monocortical, or 50% transcortical 3.
5 mm screw was implanted in each ROI and axial pullout testing was performed.
Results: Bicortical screws were stronger than monocortical screws in all ROI except the lateral epicondylar crest.
Short monocortical metaphyseal screws were weaker than those placed in other regions.
The 50% transcortical screws were stronger than the short monocortical screws in the condyle.
A linear relationship between screw length and pull-out strength was observed.
Clinical significance: Cortical thickness and bone density measurements were obtained from multiple regions of the canine humerus using QCT.
Use of short monocortical screws may contribute to failure of locking plate fixation of humeral fractures, especially when placed in the condyle.
When bicortical screw placement is not possible, maximizing monocortical screw length may optimize fixation stability for distal humeral fractures.
Related Results
COMPARATIVE ANALYSIS OF MONOCORTICAL AND BICORTICAL METHODS OF INSTALLING MINI-IMPLANTS
COMPARATIVE ANALYSIS OF MONOCORTICAL AND BICORTICAL METHODS OF INSTALLING MINI-IMPLANTS
The aim of the study. Carry out a comparative analysis of the support ability of human jaw bone tissue in monocortical and bicortical installation of a mini-implant of own design O...
ENHANCED FIXATION STABILITY IN MEDIAL MALLEOLAR FRACTURES: COMPARISON OF SINGLE VERSUS DOUBLE BICORTICAL SCREWS IN HEALTHY AND OSTEOPOROTIC BONE MODELS
ENHANCED FIXATION STABILITY IN MEDIAL MALLEOLAR FRACTURES: COMPARISON OF SINGLE VERSUS DOUBLE BICORTICAL SCREWS IN HEALTHY AND OSTEOPOROTIC BONE MODELS
Medial malleolar fractures are common orthopaedic injuries, particularly in osteoporotic populations, and present unique challenges due to reduced bone density and an increased ris...
PROXIMAL HUMERUS FRACTURES, ANATOMY, EPIDEMIOLOGY, MECHANISMS OF ACTION, CLASSIFICATION, CLINICAL PRESENTATION, IMAGING PRESENTATION, DIFFERENTIAL DIAGNOSIS, TREATMENT AND COMPLICATIONS
PROXIMAL HUMERUS FRACTURES, ANATOMY, EPIDEMIOLOGY, MECHANISMS OF ACTION, CLASSIFICATION, CLINICAL PRESENTATION, IMAGING PRESENTATION, DIFFERENTIAL DIAGNOSIS, TREATMENT AND COMPLICATIONS
Introduction: Proximal humerus fractures (PHF) make up 5 to 6% of all fractures presented in adults. Approximately 67 to 85% of proximal humerus fractures are managed non-surgicall...
7.3-mm cannulated-shaft screws show no difference in compression or load to failure compared to 7.0-mm variable-pitch screws for bovine capital physeal fracture repair
7.3-mm cannulated-shaft screws show no difference in compression or load to failure compared to 7.0-mm variable-pitch screws for bovine capital physeal fracture repair
Abstract
Objective
To compare compression and load to failure of 7.3-mm threaded-shaft screws (7.3 TSSs) and 7.0-mm variable-pitch (7.0 VP) fully threaded compression screws for b...
The anti-loosening characteristics of cortical bone trajectory screws and conventional pedicle screws under different bone quality conditions: a comparative study
The anti-loosening characteristics of cortical bone trajectory screws and conventional pedicle screws under different bone quality conditions: a comparative study
Abstract
Purpose: We aimed to compare the characteristics of anti-loosening and the pullout force between cortical bone trajectory (CBT) and conventional trajectory (CT) sc...
Bone Allograft Pedicle Screw Augmentation
Bone Allograft Pedicle Screw Augmentation
Study Design:
We performed a comprehensive cadaveric biomechanical study to compare the fixation strength of primary screws, screws augmented with bone allograft, and s...
Fixation stability comparison of bone screws based on thread design: buttress thread, triangle thread, and square thread
Fixation stability comparison of bone screws based on thread design: buttress thread, triangle thread, and square thread
Abstract
Background
The influence of thread profile on the fixation stability of bone screws remains unclear. This study aimed to compare the fixati...
The Epidemiology of Infant Shaft Fractures of Femur or Humerus by Incidence, Birth, Accidents, and Other Causes
The Epidemiology of Infant Shaft Fractures of Femur or Humerus by Incidence, Birth, Accidents, and Other Causes
Abstract
Background: Knowledge of femur and humerus shaft fractures during infancy is scarce. The purpose of this population-based registry study was to analyze both birth-...