Javascript must be enabled to continue!
Improved quality tests of R11410-21 photomultiplier tubes for the XENONnT experiment
View through CrossRef
Abstract
Photomultiplier tubes (PMTs) are often used in
low-background particle physics experiments, which rely on an
excellent response to single-photon signals and stable long-term
operation. In particular, the Hamamatsu R11410 model is the light
sensor of choice for liquid xenon dark matter experiments, including
XENONnT. The same PMT model was also used for the predecessor,
XENON1T, where issues affecting its long-term operation were
observed. Here, we report on an improved PMT testing procedure
which ensures optimal performance in XENONnT. Using both new and
upgraded facilities, we tested 368 new PMTs in a cryogenic xenon
environment. We developed new tests targeted at the detection of
light emission and the degradation of the PMT vacuum through small
leaks, which can lead to spurious signals known as afterpulses, both
of which were observed in XENON1T.
We exclude the use of 26 of the 368 tested PMTs and categorise the
remainder according to their performance. Given that we have
improved the testing procedure, yet we rejected fewer PMTs, we
expect significantly better PMT performance in XENONnT.
Title: Improved quality tests of R11410-21 photomultiplier tubes for the XENONnT experiment
Description:
Abstract
Photomultiplier tubes (PMTs) are often used in
low-background particle physics experiments, which rely on an
excellent response to single-photon signals and stable long-term
operation.
In particular, the Hamamatsu R11410 model is the light
sensor of choice for liquid xenon dark matter experiments, including
XENONnT.
The same PMT model was also used for the predecessor,
XENON1T, where issues affecting its long-term operation were
observed.
Here, we report on an improved PMT testing procedure
which ensures optimal performance in XENONnT.
Using both new and
upgraded facilities, we tested 368 new PMTs in a cryogenic xenon
environment.
We developed new tests targeted at the detection of
light emission and the degradation of the PMT vacuum through small
leaks, which can lead to spurious signals known as afterpulses, both
of which were observed in XENON1T.
We exclude the use of 26 of the 368 tested PMTs and categorise the
remainder according to their performance.
Given that we have
improved the testing procedure, yet we rejected fewer PMTs, we
expect significantly better PMT performance in XENONnT.
Related Results
Hydrogel and Polyester Ventilation Tubes in an Animal Model
Hydrogel and Polyester Ventilation Tubes in an Animal Model
ProblemDetermine the resorption rate and biocompatibility characteristics of novel cross‐linked hydrogel ventilations tubes and varied formulations of polyester ventilation tubes.M...
The triggerless data acquisition system of the XENONnT experiment
The triggerless data acquisition system of the XENONnT experiment
Abstract
The XENONnT detector uses the latest and largest liquid xenon-based time projection chamber (TPC) operated by the XENON Collaboration, aimed at detecting We...
Potable Water Sources, Household Hygiene, and Sanitation Practices in Ikpoba Okha LGA, Edo State: Implications for Public Health and Sustainable Water Management
Omoregie, Andrew Edosa.1 Omoregie Abieyuwa Peace2 Okoro, Enyinnaya Okoro.3
1 College of Medi
Potable Water Sources, Household Hygiene, and Sanitation Practices in Ikpoba Okha LGA, Edo State: Implications for Public Health and Sustainable Water Management
Omoregie, Andrew Edosa.1 Omoregie Abieyuwa Peace2 Okoro, Enyinnaya Okoro.3
1 College of Medi
BACKGROUND
Access to potable drinking water and sufficient sanitation continues to be an urgent global concern, particularly in developing regions where con...
Evaluation of Additively Manufactured Heat Exchanger Tubes Using Experimental and Conjugate Heat Transfer Analysis
Evaluation of Additively Manufactured Heat Exchanger Tubes Using Experimental and Conjugate Heat Transfer Analysis
Abstract
Additive manufacturing offers many benefits for the design and production of gas turbine engine components. Direct metal laser sintering (DMLS) expands the ...
XENONnT analysis: Signal reconstruction, calibration, and event selection
XENONnT analysis: Signal reconstruction, calibration, and event selection
The XENONnT experiment, located at the INFN Laboratori Nazionali del Gran Sasso, Italy, features a 5.9 tonne liquid xenon time projection chamber surrounded by an instrumented neut...
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
EnvironmentalSurveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) This comprehensive protocol suite enables systematic environmental surveillance for avian influenza...
Enhancing the care of cancer patients with ureteral strictures managed with percutaneous nephrostomy tubes.
Enhancing the care of cancer patients with ureteral strictures managed with percutaneous nephrostomy tubes.
e264 Background: Ureteral strictures resulting from either active disease or the sequelae of cancer therapy are often managed with percutaneous nephrostomy (PCN) tubes. PCN tubes ...
Failure Analysis of a Heat Exchanger 212-E9
Failure Analysis of a Heat Exchanger 212-E9
Heat Exchanger 212-E9 is one part of the working fluid Debutanizer unit with seawater on inner tubes and H2S on the outside of the tubes. These tubes have failed in the form of lea...