Implant choice impacts initial tibial component micromotion in total ankle replacement: A biomechanical analysis

View through CrossRef
Introduction: Despite advances in technique and implant design, total ankle replacement remains at risk of early mechanical failure. Clinical data suggest that tibial components with more robust fixation, particularly stemmed designs, have lower rates of early aseptic loosening. One proposed explanation is improved load transfer and reduced micromotion at the bone-implant interface. This study compared tibial component micromotion between stemmed and low-profile peg fixation implants using a cadaveric robotic gait simulator. We hypothesized that stemmed implants would demonstrate less micromotion during stance. Methods: Five matched cadaveric pairs underwent computed tomography (CT) scans to create patient-specific instrumentation. Each specimen received either a low-profile peg or a stemmed tibial component in a randomized order. After implantation, gait was simulated with a validated robotic system that rotates a force plate beneath a fixed tibia. A digital image correlation system tracked surface markers to measure the relative motion between the bone and the implant. Post-simulation CT scans generated segmented geometries, and rigid-body transformations were used to calculate implant micromotion. Peak micromotion was defined as the maximum nodal displacement on the implant surface during stance. Paired t tests compared matched specimens. Results: Mean peak micromotion was significantly greater in low-profile implants than in stemmed implants, measuring 382 versus 240 micrometers. In every matched pair, the low-profile design demonstrated higher micromotion, with an average increase of 59 percent. Peak motion in low-profile implants occurred posteriorly, consistent with sagittal rocking. In contrast, micromotion in stemmed implants was more evenly distributed and directed superiorly into bone.Conclusion: Stemmed tibial components showed significantly less time-zero micromotion than low-profile peg designs. Excess posterior rocking in low-profile implants may contribute to early failure and subsidence.
Title: Implant choice impacts initial tibial component micromotion in total ankle replacement: A biomechanical analysis
Description:
Introduction: Despite advances in technique and implant design, total ankle replacement remains at risk of early mechanical failure.
Clinical data suggest that tibial components with more robust fixation, particularly stemmed designs, have lower rates of early aseptic loosening.
One proposed explanation is improved load transfer and reduced micromotion at the bone-implant interface.
This study compared tibial component micromotion between stemmed and low-profile peg fixation implants using a cadaveric robotic gait simulator.
We hypothesized that stemmed implants would demonstrate less micromotion during stance.
Methods: Five matched cadaveric pairs underwent computed tomography (CT) scans to create patient-specific instrumentation.
Each specimen received either a low-profile peg or a stemmed tibial component in a randomized order.
After implantation, gait was simulated with a validated robotic system that rotates a force plate beneath a fixed tibia.
A digital image correlation system tracked surface markers to measure the relative motion between the bone and the implant.
Post-simulation CT scans generated segmented geometries, and rigid-body transformations were used to calculate implant micromotion.
Peak micromotion was defined as the maximum nodal displacement on the implant surface during stance.
Paired t tests compared matched specimens.
Results: Mean peak micromotion was significantly greater in low-profile implants than in stemmed implants, measuring 382 versus 240 micrometers.
In every matched pair, the low-profile design demonstrated higher micromotion, with an average increase of 59 percent.
Peak motion in low-profile implants occurred posteriorly, consistent with sagittal rocking.
In contrast, micromotion in stemmed implants was more evenly distributed and directed superiorly into bone.
Conclusion: Stemmed tibial components showed significantly less time-zero micromotion than low-profile peg designs.
Excess posterior rocking in low-profile implants may contribute to early failure and subsidence.

Related Results

Proximal Tibial Epiphysiodesis in a Growing Dog
Proximal Tibial Epiphysiodesis in a Growing Dog
 Background: It is believed that the inclined tibial plateau angle to be a major cause of cranial cruciate ligament (CCL) rupture, and the treatment of this disease is the tibial p...
he prevalence and clinical presentation of fibularis myofascial trigger points in the assessment and treatment of inversion ankle sprains
he prevalence and clinical presentation of fibularis myofascial trigger points in the assessment and treatment of inversion ankle sprains
Ankle sprains account for 85% of all injuries to the ankle (Garrick, 1997). Inversion sprains result from a twisting of a weight-bearing foot into a plantarflexed and inverted posi...
Total ankle replacement
Total ankle replacement
Abstract. Introduction Total ankle replacement is definitely a tough issue for both orthopedic surgeons treating patients with ankle pathology and engineers who develop optimal imp...
FRACTURES OF THE BONES IN THE ANKLE JOINT
FRACTURES OF THE BONES IN THE ANKLE JOINT
Introduction: Ankle fractures are usually frequent in emergency departments worldwide, with an incidence of 187/100,000 inhabitants per year. Especially the type B fracture accordi...
Percent Weightbearing in Foot and Ankle X-rays
Percent Weightbearing in Foot and Ankle X-rays
Category: Other; Ankle; Hindfoot; Midfoot/Forefoot Introduction/Purpose: Weightbearing x-rays remain standard of care in diagnosing and evaluating foot and ankle injuries and are c...

Back to Top