Dosimetry of large wedged high‐energy photon beams

View through CrossRef
The dependence of the wedge factor and central axis depth dose on field size was evaluated for 6‐, 10‐, and 24‐MV wedged photon beams for field sizes up to 40×40 cm2. The wedge factor for 60°, 45°, 30°, and 15° wedges in a 24‐MV beam was found to vary by as much as 25%, 12%, 9%, and 5%, respectively, over a field size range of 5×5 to 40×40 cm2. For 10 and 6 MV wedged beams, the wedge factors varied by up to 17% and 15%, respectively, over the same field size range. The depth dose curves for the wedged beams differed significantly from the open beam profiles. At 6 MV, the wedges caused beam hardening while at 24 MV, with the exception of the 15° wedge, all wedged beams were softer than the open beams, for all field sizes. At 10 MV, wedged fields of size less than 20×20 cm2 were hardened relative to the open beam, whereas larger wedged fields had depth dose values within ±1% of the 10‐MV open‐beam depth dose data. Accurate treatment planning for large wedged fields and high‐energy photon beams thus requires the use of wedged beam depth dose curves and field size specific wedge factors. It was established that an equivalent square field for a rectangular wedged field can be determined using the standard open beam formulation. The largest difference between the wedge factor for a rectangular beam and its equivalent square beam was 2.5% and occurred for 24‐MV elongated fields. The depth dose values for a rectangular wedged field and its equivalent square beam were within ±1% at all energies studied.
Title: Dosimetry of large wedged high‐energy photon beams
Description:
The dependence of the wedge factor and central axis depth dose on field size was evaluated for 6‐, 10‐, and 24‐MV wedged photon beams for field sizes up to 40×40 cm2.
The wedge factor for 60°, 45°, 30°, and 15° wedges in a 24‐MV beam was found to vary by as much as 25%, 12%, 9%, and 5%, respectively, over a field size range of 5×5 to 40×40 cm2.
For 10 and 6 MV wedged beams, the wedge factors varied by up to 17% and 15%, respectively, over the same field size range.
The depth dose curves for the wedged beams differed significantly from the open beam profiles.
At 6 MV, the wedges caused beam hardening while at 24 MV, with the exception of the 15° wedge, all wedged beams were softer than the open beams, for all field sizes.
At 10 MV, wedged fields of size less than 20×20 cm2 were hardened relative to the open beam, whereas larger wedged fields had depth dose values within ±1% of the 10‐MV open‐beam depth dose data.
Accurate treatment planning for large wedged fields and high‐energy photon beams thus requires the use of wedged beam depth dose curves and field size specific wedge factors.
It was established that an equivalent square field for a rectangular wedged field can be determined using the standard open beam formulation.
The largest difference between the wedge factor for a rectangular beam and its equivalent square beam was 2.
5% and occurred for 24‐MV elongated fields.
The depth dose values for a rectangular wedged field and its equivalent square beam were within ±1% at all energies studied.

Related Results

Odd version Mathieu-Gaussian beam based on Green function
Odd version Mathieu-Gaussian beam based on Green function
Like the theoretical pattern of non-diffracting Bessel beams, ideal non-diffracting Mathieu beams also carry infinite energy, but cannot be generated as a physically realizable ent...
The photon blockade effect of a complete Buck-Sukumar model
The photon blockade effect of a complete Buck-Sukumar model
The Buck-Sukumar (BS) model, with a nonlinear coupling between the atom and the light field, is well defined only when its coupling strength is lower than a critical coupling. Its ...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Shear Stresses of Hollow Lightweight Concrete Beams Made with Wood Waste
Shear Stresses of Hollow Lightweight Concrete Beams Made with Wood Waste
Hollow Lightweight Concrete (HLC) beams are gaining popularity due to low cost and low weight as compared with the Solid Lightweight Concrete (SLC) beams. HLC and SLC beams decreas...
SU‐F‐T‐561: Energy Dependence of a Scintillation Detector
SU‐F‐T‐561: Energy Dependence of a Scintillation Detector
Purpose:To investigate the response and dose rate dependence of a scintillation detector over a wide energy range.Methods:The energy dependence of W1 scintillation detector was tes...
Construction of Bi-Pearcey beams and their mathematical mechanism
Construction of Bi-Pearcey beams and their mathematical mechanism
We present a theoretical expression in the form of the Pearcey function by deducing the Fresnel diffraction distribution of an elliptic line. Then, we numerically simulate and expe...
Dynamic Characteristics Analysis of Three-Layer Steel–Concrete Composite Beams
Dynamic Characteristics Analysis of Three-Layer Steel–Concrete Composite Beams
The dynamic behavior of three-layer composite beams, consisting of concrete slabs and steel beams, is influenced by the structural configuration of each layer as well as the shear ...

Back to Top