Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications
View through CrossRef
Background
In many countries, disposable high-value consumables are sterilized and reused for economic reasons. We aimed to study the feasibility and ways to reduce potential risks of reusing these consumables in shoulder arthroscopy.
Methods
We reported cases with complications potentially related to reusing high-value consumables. We evaluated the outflow water temperature from new and reused radiofrequency (RF) devices of different brands and under varying surgical practices. We also conducted fatigue tests on suture-passing needles from different brands subjected to various disinfection methods.
Results
Reused RF devices from MECHAN demonstrated significantly higher peak temperature (PT) (68.3 ± 2.3°C vs 63.9 ± 3.0°C,
P
< 0.001) and peak sustaining temperature (PST) (62.6 ± 2.3°C vs 58.5 ± 3.1°C,
P
< 0.001) compared to new devices. The mean flow rate of MECHAN devices was higher than that of DePuy (63.2 ± 8.2 ml/min vs 42.0 ± 11.1 ml/min,
P
< 0.001). Ethylene oxide disinfection did not significantly affect the excitation times before fatigue fracture of Arthrex (291.6 ± 76.9 vs 305.1 ± 86.6,
P
= 0.717) and DePuy (173.3 ± 26.8 vs 174.6 ± 31.7,
P
= 0.922) suture-passing needles. High-pressure steam disinfection also did not markedly affect the fatigue time of Arthrex (291.6 ± 76.9 vs 292.0 ± 65.7,
P
= 0.990) and DePuy (173.3 ± 26.8 vs 168.8 ± 36.1,
P
= 0.755) suture-passing needles.
Conclusions
For both RF devices and suture-passing needles, being reused after ethylene oxide disinfection for appropriate times can be feasible and safe if surgeons and disinfection suppliers exercise caution.
SAGE Publications
Title: Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications
Description:
Background
In many countries, disposable high-value consumables are sterilized and reused for economic reasons.
We aimed to study the feasibility and ways to reduce potential risks of reusing these consumables in shoulder arthroscopy.
Methods
We reported cases with complications potentially related to reusing high-value consumables.
We evaluated the outflow water temperature from new and reused radiofrequency (RF) devices of different brands and under varying surgical practices.
We also conducted fatigue tests on suture-passing needles from different brands subjected to various disinfection methods.
Results
Reused RF devices from MECHAN demonstrated significantly higher peak temperature (PT) (68.
3 ± 2.
3°C vs 63.
9 ± 3.
0°C,
P
< 0.
001) and peak sustaining temperature (PST) (62.
6 ± 2.
3°C vs 58.
5 ± 3.
1°C,
P
< 0.
001) compared to new devices.
The mean flow rate of MECHAN devices was higher than that of DePuy (63.
2 ± 8.
2 ml/min vs 42.
0 ± 11.
1 ml/min,
P
< 0.
001).
Ethylene oxide disinfection did not significantly affect the excitation times before fatigue fracture of Arthrex (291.
6 ± 76.
9 vs 305.
1 ± 86.
6,
P
= 0.
717) and DePuy (173.
3 ± 26.
8 vs 174.
6 ± 31.
7,
P
= 0.
922) suture-passing needles.
High-pressure steam disinfection also did not markedly affect the fatigue time of Arthrex (291.
6 ± 76.
9 vs 292.
0 ± 65.
7,
P
= 0.
990) and DePuy (173.
3 ± 26.
8 vs 168.
8 ± 36.
1,
P
= 0.
755) suture-passing needles.
Conclusions
For both RF devices and suture-passing needles, being reused after ethylene oxide disinfection for appropriate times can be feasible and safe if surgeons and disinfection suppliers exercise caution.
Related Results
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Differential Diagnosis of Neurogenic Thoracic Outlet Syndrome: A Review
Abstract
Thoracic outlet syndrome (TOS) is a complex and often overlooked condition caused by the compression of neurovascular structures as they pass through the thoracic outlet. ...
Variability of Arthroscopy Case Volume in Orthopaedic Surgery Residency
Variability of Arthroscopy Case Volume in Orthopaedic Surgery Residency
Purpose
To examine orthopaedic surgery case logs for arthroscopy case volume during residency training and to evaluate trends in case volume and variability ove...
The effect of prior hip arthroscopy on outcomes of total hip arthroplasty: a systematic review of comparative studies
The effect of prior hip arthroscopy on outcomes of total hip arthroplasty: a systematic review of comparative studies
Abstract
Background
Hip arthroscopy is increasingly used to treat femoroacetabular impingement, labral tears, and related...
Hip Arthroscopy Surgical Volume Trends and 30‐Day Postoperative Complications
Hip Arthroscopy Surgical Volume Trends and 30‐Day Postoperative Complications
Purpose
To determine hip arthroscopy surgical volume trends from 2006 to 2013 using the National Surgical Quality Improvement Program (NSQIP) database, the inci...
Hip Arthroscopy for Femoroacetabular Impingement Syndrome Reduces the Risk for Developing Hip Osteoarthritis: A Systematic Review
Hip Arthroscopy for Femoroacetabular Impingement Syndrome Reduces the Risk for Developing Hip Osteoarthritis: A Systematic Review
Purpose
To perform a systematic review to directly compare hip arthroscopy and nonoperative treatment for femoroacetabular impingement syndrome (FAIS) with outc...
Accuracy analysis of the new artificial anatomical marker positioning method (shoulder-to-shoulder) in preventing leg length discrepancy in total hip arthroplasty
Accuracy analysis of the new artificial anatomical marker positioning method (shoulder-to-shoulder) in preventing leg length discrepancy in total hip arthroplasty
ObjectiveBy comparing the hip arthroplasty parameters planned with the AIHIP three-dimensional simulation surgery system, this study analyzes the accuracy of the new femoral-side “...
Acute and Subacute Changes in Dynamic Postural Control After Hip Arthroscopy and Postoperative Rehabilitation
Acute and Subacute Changes in Dynamic Postural Control After Hip Arthroscopy and Postoperative Rehabilitation
Context
Hip pain is associated with impairments in postural control and balance. The Star Excursion Balance Test (SEBT) is a reliable and valid method for measuri...
Effects of shoulder distraction on canine shoulder arthroscopy
Effects of shoulder distraction on canine shoulder arthroscopy
AbstractObjectiveTo evaluate the use of the Leipzig distractor during canine shoulder arthroscopy.Study designExperimental, ex vivo.Sample populationPaired shoulder joints from 15 ...