Real-time Motion Management for Robotic Treatment Couch in Proton Therapy System

View through CrossRef
Abstract Background: In proton therapy, robotic couch was used for patient positioning to achieve high precision in cancer treatment. However, no studies have focused on the real-time motion management system (MMS) for the robotic couch. This study was to develop the MMS and verify its feasibility. Methods: A real-time MMS was set up with a robotic couch, an NDI Vega optical tracking system, a Special Measurement Tool (SMT) and the developed software. A Leica laser tracker was adopted to verify the accuracy of the tests in this study. The stability and accuracy of the SMT in NDI Vega system were verified to guarantee the reliability of the results. The coordinate transformation was used to unify the measurement. 83 points inside the treatment volume were applied to verify the accuracy of the MMS, with the comparison of results measured by the laser tracker. The synchronism of the SMT in the MMS and Patient Positioning System (PPS) based on the reciprocating motion along axis z was tested to obtain the maximum time delay. Trajectory tracking was performed with different payloads on the couch top. Results: The Means and Standard Deviation (SD) of the stability of the SMT in NDI tracking system on the axis x, y, z, and Rx, Ry, Rz were 0.005 ± 0.004 mm, 0.013 ± 0.005 mm, 0.014 ± 0.005 mm, 0.010 ± 0.003º, 0.013 ± 0.005º, 0.008 ± 0.007º, respectively. For the MMS, the means ± SD along axis x, y, z were 0.24 ± 0.09 mm (0.12 ± 0.03 mm With Laser Tracker (WLT)), 0.31 ± 0.12 mm (0.10 ± 0.03 mm WLT), 0.18 ± 0.06 mm (0.14 ± 0.05 mm WLT), respectively. The max time delay was about 110 ms, with a 60 s of motion tracking. Trajectories tracking showed the real-time graphical traces of the SMT in NDI Vega and Room Coordinate system, and the combined curves by using the coordinate transformation. Conclusions: A real-time MMS was developed and verified with the accuracy of submillimeter in this study. The results demonstrated the feasibility of potential applications based on real-time motion capturing and monitoring.
Title: Real-time Motion Management for Robotic Treatment Couch in Proton Therapy System
Description:
Abstract Background: In proton therapy, robotic couch was used for patient positioning to achieve high precision in cancer treatment.
However, no studies have focused on the real-time motion management system (MMS) for the robotic couch.
This study was to develop the MMS and verify its feasibility.
Methods: A real-time MMS was set up with a robotic couch, an NDI Vega optical tracking system, a Special Measurement Tool (SMT) and the developed software.
A Leica laser tracker was adopted to verify the accuracy of the tests in this study.
The stability and accuracy of the SMT in NDI Vega system were verified to guarantee the reliability of the results.
The coordinate transformation was used to unify the measurement.
83 points inside the treatment volume were applied to verify the accuracy of the MMS, with the comparison of results measured by the laser tracker.
The synchronism of the SMT in the MMS and Patient Positioning System (PPS) based on the reciprocating motion along axis z was tested to obtain the maximum time delay.
Trajectory tracking was performed with different payloads on the couch top.
Results: The Means and Standard Deviation (SD) of the stability of the SMT in NDI tracking system on the axis x, y, z, and Rx, Ry, Rz were 0.
005 ± 0.
004 mm, 0.
013 ± 0.
005 mm, 0.
014 ± 0.
005 mm, 0.
010 ± 0.
003º, 0.
013 ± 0.
005º, 0.
008 ± 0.
007º, respectively.
For the MMS, the means ± SD along axis x, y, z were 0.
24 ± 0.
09 mm (0.
12 ± 0.
03 mm With Laser Tracker (WLT)), 0.
31 ± 0.
12 mm (0.
10 ± 0.
03 mm WLT), 0.
18 ± 0.
06 mm (0.
14 ± 0.
05 mm WLT), respectively.
The max time delay was about 110 ms, with a 60 s of motion tracking.
Trajectories tracking showed the real-time graphical traces of the SMT in NDI Vega and Room Coordinate system, and the combined curves by using the coordinate transformation.
Conclusions: A real-time MMS was developed and verified with the accuracy of submillimeter in this study.
The results demonstrated the feasibility of potential applications based on real-time motion capturing and monitoring.

Related Results

MO‐SAM‐BRB‐02: Proton Physics and Technology
MO‐SAM‐BRB‐02: Proton Physics and Technology
The dose localization advantages of proton beams derive primarily from the Bragg peak in the proton stopping distribution. Therefore, the potential clinical gains expected form pro...
Woningcorporaties en Vastgoedontwikkeling
Woningcorporaties en Vastgoedontwikkeling
This summary highlights the findings of the PhD-thesis ‘Woningcorporaties en Vastgoedontwikkeling: Fit for Use’ (‘Housing associations and Real Estate Development: Fit for Use?’). ...
The Culture of Couch Surfing Amongst Students
The Culture of Couch Surfing Amongst Students
<p>This thesis explores ‘couch surfing’ as it becomes more prevalent within the tertiary student experience. In recent years, there has been a significant decline in the avai...
Evaluating the Cost for Robotic vs “Non-Robotic” Transhiatal Esophagectomy
Evaluating the Cost for Robotic vs “Non-Robotic” Transhiatal Esophagectomy
Introduction This study was undertaken to analyze and compare the cost of robotic transhiatal esophagectomy (THE) to “non-robotic” THE (ie, “open” and laparosco...
Proton aurora related to intervals of pulsations of diminishing periods
Proton aurora related to intervals of pulsations of diminishing periods
Geomagnetic pulsations in the Pc1 frequency range are believed to be an indicator of electromagnetic ion cyclotron waves arriving from the equatorial magnetosphere, where the waves...
Methodology to Define Design Motion Criteria for Performance of Floating LNG Process Facilities
Methodology to Define Design Motion Criteria for Performance of Floating LNG Process Facilities
Abstract This paper proposes a generalized methodology to determine motion criteria for required performance of process facilities using the Abadi Floating LNG (A...
SU‐E‐P‐39: Characterization of Dose Impact On Different Beam Fields Size Attenuation for Elekta IGRT Couch
SU‐E‐P‐39: Characterization of Dose Impact On Different Beam Fields Size Attenuation for Elekta IGRT Couch
Purpose:Quantification of the dosimetric impact of the Elekta IGRT treatment couch in different beam field sizes. Established the relationship of relative dose versus beam angle at...
Computer simulation of an excess proton in aqueous systems
Computer simulation of an excess proton in aqueous systems
This thesis aims at studying the microscopic physical-chemical properties of an excess proton in aqueous systems. From bulk water environments to narrow hydrophobic channels constr...

Back to Top