Impact of 3-dimensional electroanatomic mapping systems on fluoroscopy reduction in catheter ablation of supraventricular tachycardia - then and now
View through CrossRef
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
Catheter ablation is an effective first-line treatment for symptomatic and recurrent supraventricular tachycardia (SVT). The advent of 3D electroanatomic mapping (EAM) has led to a reduction in fluoroscopy use and consequently lower radiation exposure for patients and staff during SVT ablation.
Purpose
The aims of this study are to demonstrate if the use of EAM during SVT ablation reduces fluoroscopy time (FT), and determine if further reductions in FT are observed longitudinally. We hypothesise that over time, with continued use and wider adoption of EAM, greater operator experience, and technological advancements in EAM, there will be further reductions in radiation exposure as estimated by FT.
Methods
All cases of atrioventricular nodal re-entry tachycardia (AVNRT), atrioventricular re-entry tachycardia (AVRT), and/or atrial tachycardia (AT) ablation between May 2011 to May 2022 at a tertiary centre were prospectively recruited. FT between the groups with and without EAM were compared. Within the EAM subset, the trend of FT across the years was also analysed.
Results
Over this 11-year period, there were 1758 cases of SVT ablation (565 without EAM; 1193 with EAM). Between the two groups, there were no significant differences (p > 0.05) in age, sex, history of structural heart disease and/or ischaemic heart disease, stereotaxis use, and diagnosis (AVNRT, AVRT, AT). Cases using EAM were more likely to have >1 SVT ablated during the procedure (0.04) and require intracardiac echocardiography (ICE) (p < 0.001).
The use of EAM was associated with mean reductions in FT and dose area product (DAP) by 19.6 minutes and 18 581 mGy*cm2 respectively (both p < 0.001). EAM was also associated with a longer procedure time (mean +8.8 minutes, p < 0.001). There was no difference in radiofrequency application time between the two groups (p = 0.143).
After controlling for potential confounding factors including age, sex, use of stereotaxis, use of ICE, type and number of SVT, history of prior ablation, and structural heart disease and/or ischaemic heart disease, we found over time (2011-2022), further reduction in FT was observed, with a mean reduction of 0.9 minutes year on year (p < 0.001).
EAM was gradually adopted through the years with the number of cases per year plateauing after 2017. Between 2011-2017, there was a significant reduction in FT (mean of -1.1 minute year on year, p = 0.020), which was not observed from 2017 onwards (p = 0.088). The greatest reduction in FT was observed after the first year of adoption, with a mean reduction of 6.9 minutes between 2012-2013.
Conclusion
This demonstrates that use of EAM in SVT ablation reduces FT and fluoroscopy DAP. There was no incremental benefit observed after 6 years of EAM adoption likely related to operator experience. While there is increased interest in zero fluoroscopy SVT ablation, complementary use of fluoroscopy may still be necessary in select complex cases.
Title: Impact of 3-dimensional electroanatomic mapping systems on fluoroscopy reduction in catheter ablation of supraventricular tachycardia - then and now
Description:
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
Catheter ablation is an effective first-line treatment for symptomatic and recurrent supraventricular tachycardia (SVT).
The advent of 3D electroanatomic mapping (EAM) has led to a reduction in fluoroscopy use and consequently lower radiation exposure for patients and staff during SVT ablation.
Purpose
The aims of this study are to demonstrate if the use of EAM during SVT ablation reduces fluoroscopy time (FT), and determine if further reductions in FT are observed longitudinally.
We hypothesise that over time, with continued use and wider adoption of EAM, greater operator experience, and technological advancements in EAM, there will be further reductions in radiation exposure as estimated by FT.
Methods
All cases of atrioventricular nodal re-entry tachycardia (AVNRT), atrioventricular re-entry tachycardia (AVRT), and/or atrial tachycardia (AT) ablation between May 2011 to May 2022 at a tertiary centre were prospectively recruited.
FT between the groups with and without EAM were compared.
Within the EAM subset, the trend of FT across the years was also analysed.
Results
Over this 11-year period, there were 1758 cases of SVT ablation (565 without EAM; 1193 with EAM).
Between the two groups, there were no significant differences (p > 0.
05) in age, sex, history of structural heart disease and/or ischaemic heart disease, stereotaxis use, and diagnosis (AVNRT, AVRT, AT).
Cases using EAM were more likely to have >1 SVT ablated during the procedure (0.
04) and require intracardiac echocardiography (ICE) (p < 0.
001).
The use of EAM was associated with mean reductions in FT and dose area product (DAP) by 19.
6 minutes and 18 581 mGy*cm2 respectively (both p < 0.
001).
EAM was also associated with a longer procedure time (mean +8.
8 minutes, p < 0.
001).
There was no difference in radiofrequency application time between the two groups (p = 0.
143).
After controlling for potential confounding factors including age, sex, use of stereotaxis, use of ICE, type and number of SVT, history of prior ablation, and structural heart disease and/or ischaemic heart disease, we found over time (2011-2022), further reduction in FT was observed, with a mean reduction of 0.
9 minutes year on year (p < 0.
001).
EAM was gradually adopted through the years with the number of cases per year plateauing after 2017.
Between 2011-2017, there was a significant reduction in FT (mean of -1.
1 minute year on year, p = 0.
020), which was not observed from 2017 onwards (p = 0.
088).
The greatest reduction in FT was observed after the first year of adoption, with a mean reduction of 6.
9 minutes between 2012-2013.
Conclusion
This demonstrates that use of EAM in SVT ablation reduces FT and fluoroscopy DAP.
There was no incremental benefit observed after 6 years of EAM adoption likely related to operator experience.
While there is increased interest in zero fluoroscopy SVT ablation, complementary use of fluoroscopy may still be necessary in select complex cases.
Related Results
Case report: Para Hisian Atrial Tachycardia
Case report: Para Hisian Atrial Tachycardia
Atrial tachycardia accounts for around 10% of supraventricular tachycardias (1). Atrial tachycardia can either be focal or macrorenentrant. Focal AT usually start at a focal p...
Atrial Flutter, The Commonly Misdiagnosed Arrhythmia as Supraventricular Tachycardia or Sinus Tachycardia
Atrial Flutter, The Commonly Misdiagnosed Arrhythmia as Supraventricular Tachycardia or Sinus Tachycardia
Background: Atrial flutter is a common arrhythmia in structurally normal or abnormal heart. The electrocardiographic features of it can be mistaken for sinus tachycardia or suprave...
Mapping workflow trends in pulsed-field ablation procedures: an international glimpse
Mapping workflow trends in pulsed-field ablation procedures: an international glimpse
Abstract
Background
As pulsed field ablation (PFA) is increasingly used in the EP lab, the use of mapping, fluoroscopy, and intr...
The advantage of mini electrode-equipped catheter for the radiofrequency ablation of paroxysmal supraventricular tachycardia
The advantage of mini electrode-equipped catheter for the radiofrequency ablation of paroxysmal supraventricular tachycardia
Introduction:
Novel ablation catheters equipped with
mini-electrodes (ME) offer high resolution mapping for target tissue.
This study aimed to evaluate the mappin...
Fluoroless vs. conventional catheter ablation of supraventricular tachycardia - retrospective single-centre analysis
Fluoroless vs. conventional catheter ablation of supraventricular tachycardia - retrospective single-centre analysis
Abstract
Funding Acknowledgements
Type of funding sources: None.
Introduction
...
“Sequential” Mapping Mimicking “Simultaneous” Mapping Using Magnetic Navigation During Catheter Ablation of Supraventricular Tachycardia: Results of the Single DX Study
“Sequential” Mapping Mimicking “Simultaneous” Mapping Using Magnetic Navigation During Catheter Ablation of Supraventricular Tachycardia: Results of the Single DX Study
Introduction: The magnetic navigation system (MNS) allows remote‐controlled navigation of an ablation catheter from the control room. We tested the hypothesis, whether the MNS and ...
Impact of Catheter Ablation on Ventricular Tachycardia Recurrence in Patients with Chagas Cardiomyopathy Following implantable cardioverter-defibrillator Implantation: A Competing Risks Analysis
Impact of Catheter Ablation on Ventricular Tachycardia Recurrence in Patients with Chagas Cardiomyopathy Following implantable cardioverter-defibrillator Implantation: A Competing Risks Analysis
ABSTRACT
Background
Chagas cardiomyopathy (ChC) is a highly arrhythmogenic condition with frequent implantable cardioverter-def...
ASSA13-02-26 Stroke in young patients with paroxysmal supraventricular tachycardia: A report of two cases of thromboembolic stroke
ASSA13-02-26 Stroke in young patients with paroxysmal supraventricular tachycardia: A report of two cases of thromboembolic stroke
Abstract
Paroxysmal Supraventricular Tachycardia (PSVT) is the most common type of sustained tachycardia in healthy adults. It is generally believed that thromboe...